Neisseria Unit, Statens Serum Institut,1 Copenhagen Hospital Corporation, Copenhagen University Hospital, Copenhagen, Denmark2
Received 2 November 2001/ Returned for modification 15 February 2002/ Accepted 11 April 2002
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
The laboratories performing the MIF test use a variety of in-house assays; assays based on antigens from Washington Research Foundation (WRF) (now Washington University), Seattle, Wash.; or one of the commercially available assays (13). Compared to the test originally developed by Wang and Grayston (21), the different assays each have introduced minor variations in the materials used or in procedures to be followed, e.g., different strains of C. pneumoniae as the antigen, different incubation times of sera with the antigen, and the use of fluorescein isothiocyanate-labeled anti-human immunoglobulin (Ig) antibodies from different manufacturers. One study examined the interlaboratory variation in MIF assay results obtained in 13 laboratories analyzing 22 sera from 10 patients (13). The agreement between a reference standard value and the IgM antibody results obtained by the 13 laboratories using 11 different methods was 50 to 95%; four laboratories failed to discern false-positive IgM titers, possibly because of the presence of rheumatoid factor. For IgG antibody analyses, the agreement was 68 to 87%. In another study, one laboratory compared two MIF assays and found a significant difference in IgG and IgA antibody titer levels obtained (5).
The objective of the present study was to assess the performance of two commercially available C. pneumoniae MIF assays from MRL Diagnostics (MRL), Cypress, Calif., and Labsystems (LAB), Helsinki, Finland, using a MIF assay based on C. pneumoniae antigen from WRF as a reference method. The WRF assay was chosen as the reference method because it had been available for research purposes for decades and because previous Danish studies of the prevalence of C. pneumoniae antibodies had been conducted with the WRF assay (9, 11a). The two commercial assays (LAB and MRL) were under evaluation for use in our routine laboratory. Performance was assessed by testing for C. pneumoniae antibodies in sera from patients with acute respiratory tract infections of known etiology obtained at various intervals after the onset of the disease. Sera from persons enrolled in the Copenhagen City Heart Study (1) were included due to the present interest in the detection of C. pneumoniae antibodies in patients with cardiovascular diseases (9, 16).
| MATERIALS AND METHODS |
|---|
|
|
|---|
Group I consisted of 83 sera from 28 patients enrolled in the Nordic Atypical Pneumonia (NAP) Study (12). The sera were kindly provided by J. S. Jensen, Mycoplasma Laboratory, Statens Serum Institut, with the permission of P. Saikku, University of Oulu, Finland. The 28 patients were selected because they had serological evidence of infection with C. pneumoniae (13 patients), Mycoplasma pneumoniae (11 patients), or both agents (4 patients). The original MIF analyses for C. pneumoniae antibodies had been performed using antigens from WRF (12).
Group II consisted of 37 sera from 16 patients with acute C. pneumoniae (8 patients) or Chlamydia psittaci (8 patients) respiratory tract infections confirmed by PCR (15 patients) or culture (1 patient) in our routine laboratory. The majority of the 16 patients had developed antibodies detectable in the Chlamydia complement fixation (CF) test; moreover, clinical data were available for all patients with C. pneumoniae infection and for 5 of 8 patients with C. psittaci infection. For the last group of patients, epidemiological data were also available (M. Faber, J. S. Jensen, and I. Lind, Proceedings of the 3rd Nordic-Baltic Congress on Infectious Diseases, abstr. 4, p. 12, 1998). Culture was performed in Hep2 and McCoy cells; subsequent identification was obtained with genus- and species-specific staining methods (4). PCR was performed by an in-house method (4, 14).
Group III consisted of 100 sera from 100 persons enrolled in the Copenhagen City Heart Study, which is a prospective, ongoing population study: 20 persons who developed acute myocardial infarction within 3 years after the blood samples were drawn and 80 controls matched according to gender and age at entry without subsequent myocardial infarction (1).
MIF assays. The three MIF assays are described in brief below. Prior to IgM and IgA antibody testing, all sera were treated with GullSORB (Gull Laboratories, Salt Lake City, Utah). All analyses were performed by the same person (L.B.) using a Zeiss Axioskop 20 microscope with a Plan-Neofluor 40x/0.75 objective. The procedures followed the recommendations of each manufacturer.
(i) WRF assay. For antigens from WRF, purified C. pneumoniae elementary bodies (strain AR39) were fixed on the slides. The conjugates used were fluorescein-labeled rabbit anti-human IgM, IgG, and IgA (Dako, Glostrup, Denmark). The incubation time was 30 min for both sera and conjugates.
(ii) MRL assay. For MIF assays (IgM, IgG, and IgA) from MRL, the slides were purchased with prefixed antigen dots for C. pneumoniae (strain TW183), C. trachomatis (eight serotypes [D to K]), and C. psittaci (strains 6BC and DD34). For the detection of IgM antibodies, the incubation period with serum was 90 min; for the detection of IgG and IgA antibodies, it was 30 min. The subsequent incubation time with the conjugate (fluorescein-labeled goat anti-human Igs [MRL]) was 30 min.
(iii) LAB assay. The slides for MIF assays (IgM and IgG) from LAB were purchased with prefixed antigen dots for C. pneumoniae (strain Kajaani 6), C. trachomatis (strain LGV/L2), and C. psittaci (strains OF 6bC and EAE). For the detection of IgM antibodies, the incubation period with serum was 180 min; for the detection of IgG antibodies, it was 30 min. The subsequent incubation time with the conjugate (fluorescein-labeled goat anti-human Igs [LAB]) was 30 min.
All sera were tested in serial twofold dilutions from the following initial dilutions until an endpoint was reached: group I, 1:16 (IgM), 1:64 (IgG), and 1:16 (IgA); group II, 1:16 (IgM) and 1:64 (IgG); group III, 1:16 (IgG) and 1:16 (IgA). Table 1 presents an overview of the experiments performed.
|
Statistics.
To determine the intraday variation of endpoint titers for each serum for each day, a two-way analysis of variance was performed. The interday variation was estimated with an F test. The variation of reproducibility was estimated for each Ig class and for the three classes together. A comparison of detection rates obtained by the three assays was performed with a likelihood ratio test; for comparison of the antibody titers, Friedman's nonparametric two-way analysis of variance was used. These analyses were performed with SAS version 8 (SAS Institute Inc.). The 95% confidence intervals (CI) for detection rates were calculated as the proportion (p)
(normal approximation). A P value of <0.05 was considered significant.
| RESULTS |
|---|
|
|
|---|
1 dilution step.
Detection rates of C. pneumoniae antibodies.
The presence of C. pneumoniae IgM and IgG antibodies in 120 sera from 65 patients with acute respiratory tract infection (groups I and II) is illustrated in Tables 2, 3, 4, and 5. By the WRF, MRL, and LAB assays, the detection rates of IgM antibody titers of
16 were 27, 24, and 28%, respectively (P = 0.16), and the detection rates of IgG antibody titers of
64 were 45, 53, and 48% (P = 0.43). Concordant results (both tests positive or both tests negative) were demonstrated by the MRL assay compared to the WRF assay in 92% of the sera (111 of 120) for IgM C. pneumoniae antibodies, in 76% of the sera (91 of 120) for IgG antibodies, and in 80% of the sera (66 of 83) (data not shown) for IgA antibodies. The corresponding figures for the results by the LAB assay compared to those by the WRF assay were 97% of the sera (116 of 120) for IgM antibodies and 79% of the sera (95 of 120) for IgG antibodies. When a day-to-day variation of ±1 dilution step was taken into consideration, the results for the MRL assay versus the WRF assay were concordant in 98, 88, and 88% of the sera for IgM, IgG, and IgA antibodies, and for the LAB assay versus the WRF assay, the results were concordant in 98 and 87% of the sera for IgM and IgG antibodies.
|
The LAB and WRF assay results were discrepant in 5 of 120 (4%) sera tested for IgM antibodies (Table 3). The five IgM results were all found in the first or the last of three consecutive sera, the two correspondent sera of which were positive for C. pneumoniae IgM antibodies by both assays. For these five sera, the results by the WRF assay were concordant with those by the MRL assay.
|
|
|
|
|
16, IgG antibody titers of
64, and IgA antibody titers of
32. The geometric mean of IgM antibody titers obtained by the LAB assay was about three times as high as those obtained by the WRF assay and the MRL assay. The overall difference in IgM antibody titers obtained by the three assays was significant (P = 0.018). This difference was due to the diverging results by the LAB assay. Comparing the IgG antibody titers obtained by the three assays for sera from groups I and II, the overall difference demonstrated (P = 0.032) was mainly due to a slightly lower IgG antibody level by the WRF assay than by the other two assays. With the WRF assay, 2 of 120 sera (1.7%) had C. pneumoniae IgG antibody titers of
512 as opposed to 9 of 120 sera (7.5%) with the MRL assay and 7 of 120 (5.8%) with the LAB assay (Tables 4 and 5). However, in only five sera were IgG antibody titers of
512 demonstrated by both the MRL assay and the LAB assay.
|
In conclusion, the determinations of endpoint titers were reproducible with <1 dilution step difference for all three methods, except that the mean IgM antibody titer found by the LAB assay was almost 2 dilution steps higher than those found by the other two methods.
Patients with C. pneumoniae infection. A total of 66 sera from 25 patients with either serological evidence of C. pneumoniae infection (group I; n = 17) or microbiological evidence of C. pneumoniae infection (group II; n = 8) were included.
A total of 51 sera originated from the 17 patients in group I. For IgM antibodies, the accordance in detection rates were 98% of the sera (50 of 51) for the LAB assay versus the WRF assay and 88% of the sera (45 of 51) for the MRL assay versus the WRF assay, while accordance in endpoint titer levels, defined as the WRF assay titer ± 1 dilution step, were 78 (40 of 51) and 73% (37 of 51), respectively. For the IgG antibody detection rate, the accordance was 90% (46 of 51) for both the LAB assay versus the WRF assay and the MRL assay versus the WRF assay, while accordance in endpoint titer levels was found in 84% (43 of 51) of sera by both assays. For IgA antibodies, the accordance in the detection rates for the MRL assay versus the WRF assay was 86% (44 of 51), and the accordance in titer levels was 75% (38 of 51).
By the MRL assay, C. psittaci IgM antibodies were detected in 3 of 17 group I patients (7 of 51 sera) with C. pneumoniae IgM antibodies; one of these patients also had C. trachomatis IgM antibodies. In four of the seven sera, there were <4-fold differences between C. pneumoniae and C. psittaci IgM antibody titer levels.
By the LAB assay, C. psittaci IgM antibodies were detected in 5 of 17 group I patients (13 of 51 sera) with C. pneumoniae IgM antibodies; 2 of the 5 patients also had C. trachomatis IgM antibodies. In 9 of the 13 sera, there were <4-fold differences between C. pneumoniae and C. psittaci IgM antibody titer levels.
For group II, Table 9 shows the serological results by all three methods for each patient with confirmed C. pneumoniae infection. The accordance in IgM antibody detection rates for the 15 sera from eight patients was 87% (13 of 15) for the LAB assay versus the WRF assay and 100% (15 of 15) for the MRL assay versus the WRF assay, while the accordance in endpoint titer levels, defined as the WRF assay titer ± 1 dilution step, was 47 (7 of 15) and 67% (10 of 15). For IgG antibodies, the accordance in detection rates was 53% (8 of 15) for the LAB assay versus the WRF assay and 47% (7 of 15) for the MRL assay versus the WRF assay, while accordance in endpoint titer levels was found in 40% (6 of 15) of sera by the LAB assay and 47% (7 of 15) by the MRL assay.
|
In conclusion, the accordance in detection rates of IgM antibodies was between 87 and 100% by the two assays compared to the WRF assay, and the accordance in detection rates of IgG antibodies was 90% in sera from group I and about 50% in sera from group II. The accordance in endpoint titer levels between both assays and the WRF assay was somewhat lower than the accordance in detection rates. Cross-reacting IgM antibodies were found in 9 of 25 patients by the LAB assay and in 7 of 25 patients by the MRL assay.
Patients with M. pneumoniae infection.
The accordance in C. pneumoniae IgG antibody test results was 91% by the MRL assay versus the WRF assay and 87% by the LAB assay versus the WRF assay in the 32 sera from 11 patients included in group I. By the MRL assay and the WRF assay, the test results for C. pneumoniae IgA antibodies were in accordance in 29 of 32 sera. In none of the sera were C. pneumoniae IgM or IgG antibody titers of
512 detected by any of the assays. IgG antibody titers of 64 to 256 were detected in 46 to 57% of sera by the three assays.
Patients with C. psittaci infection. A total of 22 sera from eight patients with C. psittaci infection were analyzed (data not shown). Overall, four of eight patients had C. psittaci IgM antibodies and/or a fourfold rise of C. psittaci IgG titers by the LAB assay. By the MRL assay, one of the eight patients had IgM antibodies toward all three species, and further, one had a fourfold rise of C. psittaci IgG antibody titers. C. pneumoniae IgM antibodies were detected in four sera from two patients by the MRL assay but in none of the sera by the LAB assay or the WRF assay. C. pneumoniae IgG antibody titers of 64 to 256 were detected in sera from three, four, and six patients by the WRF assay, the LAB assay, and the MRL assay, respectively.
Comparison of results for group I with those from the NAP Study.
The results for group I tested with antigens from WRF in the present study were compared to the results originally obtained in the NAP Study (12), with the same cutoff applied to both sets of results. The detection rate of C. pneumoniae IgM antibody titers of
16 was 24% (CI, 19 to 29%) versus 31% (CI, 26 to 36%) in the NAP Study, and that of IgG antibody titers of
64 was 45% (CI, 40 to 50%) versus 69% (CI, 64 to 74%), but the detection rate of IgA antibody titers of
16 was 64% (CI, 59 to 69%) versus 40% (CI, 45 to 55%). The endpoint titers obtained in the present study were 2.0 and 1.5 dilution steps lower for IgM and IgG antibodies and 0.3 dilution step higher for IgA antibodies than in the NAP Study.
| DISCUSSION |
|---|
|
|
|---|
The day-to-day variation in MIF test results implied that a difference of one dilution step is insignificant. This result is in accordance with results from other laboratories experienced in C. pneumoniae MIF analyses (11).
The reproducibility of C. pneumoniae IgG antibody endpoint titers was high, though IgG antibody titers by the WRF assay were on average 0.4 dilution step lower than those by the other two assays. However, this difference is within the day-to-day variation of one dilution step. Similarly, another recent study comparing IgG antibody titers obtained by the WRF assay and the LAB assay found an insignificant difference of one-dilution-step-higher titer levels by the latter assay (11). The IgM antibody endpoint titers found by the LAB assay were almost two dilution steps higher than those found by the WRF assay and the MRL assay. Labsystem recommends that when testing for IgM antibodies, the incubation time for sera with the antigen should be 180 min, that is, six times as long as originally recommended by Wang (19, 20). The effect of changing the incubation period from 30 to 90 min was assessed by analyzing 22 IgM antibody-positive sera with the WRF assay and the MRL assay (unpublished data). As expected, the results showed that a prolonged incubation time can increase the IgM antibody titer levels; therefore, differences in IgM titer levels obtained by different assays could be due in part to the differences in incubation times.
For the MRL assay, the average endpoint titers were 0.4 dilution step lower than those by the WRF assay for IgM antibodies and 0.4 and 0.3 dilution step higher for IgG and IgA antibodies. Freidank et al. found the average titers for IgG antibodies and IgA antibodies by the MRL assay to be 2.5 and 3.0 dilution steps higher than those obtained by the WRF assay (5). Freidank et al. used anti-human Igs from Medac in the WRF assay, whereas we used anti-human Igs from Dako. Though only minor differences among IgG antibody levels detected by the three assays were found, a difference among the proportions of sera with antibody titers of
512 was seen. By the WRF assay in 1.6% and by the MRL assay in 7.5% of sera, IgG antibody titers of
512 were demonstrated. Among sera studied by Freidank, 2% had titers of
512 with the WRF assay compared to 30% with the MRL test (5).
It has been demonstrated that the use of a C. pneumoniae antigen produced from a local isolate can result in higher detection rates of antibodies in sera and higher titers in the MIF assay than when using a standard antigen (2). The three assays studied each use a different C. pneumoniae strain as the antigen. However, the detection rates and endpoint titers by these three assays were similar when the same panel of sera were tested, with IgM antibody endpoint titers found by the LAB assay as the only exception.
Comparing the results by the MRL assay versus the WRF assay for sera from patients in groups I and II with those from group III, there were no major differences. If anything, the results obtained by the two assays were more concordant for group III sera than for sera from patients with acute respiratory tract infection (groups I and II) (Tables 2 to 7).
In the interlaboratory study by Peeling et al. (13), the agreement among the participating laboratories for IgM titers was 50 to 95%. In the present study, the agreement among the three assays in detecting IgM antibodies in sera from the eight persons with microbiologically verified C. pneumoniae infection was 87 to 100%. Considering the technical complexity of the MIF test and the subjective nature of reading the titers, it is hardly surprising that the interlaboratory variation of the study by Peeling et al. is higher than the interassay variation of the present study.
The agreement between the serological and the microbiological diagnoses for the patients with C. pneumoniae infection was high, in contrast to that for the patients with C. psittaci infection. For most of the patients with psittacosis, sera were obtained early in the course of the disease. Furthermore, there is a greater diversity between C. psittaci strains than between C. pneumoniae strains, and it is possible that the C. psittaci strains used in the LAB assay and the MRL assay did not react with C. psittaci antibodies from Danish patients with infections caused by local strains.
Labsystems has reduced the lipopolysaccharide activity in the C. pneumoniae and C. trachomatis antigens but not in the C. psittaci antigen; therefore, it might be expected that the LAB assay would detect C. psittaci antibodies in sera with positive test results by the Chlamydia CF test. However, in group II, 15 of 16 patients had a positive CF test result, but C. psittaci IgM antibodies were detected in only 4 patients.
The detection rates of C. pneumoniae IgM and IgG antibodies, as well as the antibody endpoint titer levels, were low among sera from group I of the present study compared to the results in the NAP Study; the detection rate of C. pneumoniae IgA antibody was higher in the present study. Although the same antigen for the MIF was used in the two studies, other differences in the test procedures occurred, e.g., the use of different conjugates and interpersonal differences in the reading of slides might explain some of the variations in the results obtained. Furthermore, in the present study the sera had been absorbed before the IgA antibody analyses; removal of IgG has been shown to raise IgA antibody titers in sera with high levels of IgG antibodies (9). Finally, the patients in the NAP Study were recruited between 1990 and 1993 and the sera were analyzed shortly thereafter, whereas in our study the sera were tested after storage at -20°C for up to 8 years.
The prevalence of circulating rheumatoid factor is strongly correlated with age, and the presence of rheumatoid factor in sera containing C. pneumoniae IgG antibodies might give false-positive C. pneumoniae IgM antibody test results (13, 17). In one study, 41 of 286 patients had C. pneumoniae IgM antibody titers of
16. Although only 78% of the patients had detectable circulating rheumatoid factor, none of the 41 sera were shown to have C. pneumoniae IgM antibodies after absorption of IgG antibodies (17). Since circulating rheumatoid factor at a level below the threshold of detection by routine screening can cause false-positive IgM results, it is recommended that IgG be removed prior to IgM antibody testing (3). Since the presence of IgM antibodies is considered a sign of acute infection, it is possible that some studies have reported prevalences of acute C. pneumoniae infection that have been inflated due to the inclusion of patients with false-positive IgM antibody results.
In conclusion, the three MIF assays investigated showed about the same ability to detect C. pneumoniae IgM, IgG, and IgA antibodies, but some variation was found in the antibody levels demonstrated. In the light of the established criteria for the serological diagnosis of acute C. pneumoniae infection (6), the difference in average endpoint titer levels of IgM antibodies is of no major concern, as the emphasis of the criteria rests on the presence of IgM antibodies, not on the level. In contrast, the difference in IgG antibody endpoint titers by different assays presents a problem as long as a high level of IgG titers (
512) is used as an isolated criterion for the serodiagnosis of acute C. pneumoniae infection.
| FOOTNOTES |
|---|
The Infection, Atherosclerosis and Macrolide Antibiotics Group includes the following persons: Christian Gluud, Copenhagen Trial Unit, Copenhagen University Hospital, Copenhagen, Denmark; Jørgen Fischer Hansen, Bispebjerg Hospital, Copenhagen, Denmark; Stig Haunsø, Rigshospitalet, Copenhagen, Denmark; Per Hildebrandt, Frederiksberg Hospital, Copenhagen, Denmark; Gorm Jensen, Hvidovre Hospital, Copenhagen, Denmark; Christian Jespersen, Bispebjerg Hospital, Copenhagen, Denmark; Jens Kastrup, Rigshospitalet, Copenhagen, Denmark; Erik Kjøller, KAS Herlev, Copenhagen, Denmark; Hans Jørn Kolmos, Odense Universitets Hospital, Denmark; Beata B. Malmquist, Hvidovre Hospital, Copenhagen, Denmark; Henrik Nielsen, Amager Hospital, Copenhagen, Denmark; Rolf Steffensen, Hillerød Hospital, Hillerød, Denmark. Coauthor Inga Lind is also a member of the group. ![]()
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Antimicrob. Agents Chemother. | Clin. Microbiol. Rev. | Infect. Immun. |
|---|---|---|
| J. Clin. Microbiol. | J. Virol. | ALL ASM JOURNALS |