Previous Article | Next Article 
Clinical and Diagnostic Laboratory Immunology, November 2001, p. 1070-1075, Vol. 8, No. 6
1071-412X/01/$04.00+0 DOI: 10.1128/CDLI.8.6.1070-1075.2001
Comparison of Dissociation-Enhanced Lanthanide Fluorescent
Immunoassays to Enzyme-Linked Immunosorbent Assays for Detection of
Staphylococcal Enterotoxin B, Yersinia pestis-Specific
F1 Antigen, and Venezuelan Equine Encephalitis Virus
Darci R.
Smith,
Cynthia A.
Rossi,*
Todd M.
Kijek,
Erik A.
Henchal, and
George V.
Ludwig
Diagnostic Systems Division, U.S. Army Medical Research
Institute of Infectious Diseases, Fort Detrick, Maryland 21702
Received 9 March 2001/Returned for modification 11 May
2001/Accepted 2 August 2001
 |
ABSTRACT |
The dissociation-enhanced lanthanide fluorescent immunoassays
(DELFIA) were developed for the detection of staphylococcal enterotoxin
B, Yersinia pestis-specific F1 antigen, and Venezuelan equine encephalitis virus. These assays were compared to previously developed enzyme-linked immunosorbent assays (ELISAs) by determining the sensitivity or limit of detection (LOD), the dynamic range, and the
reproducibility of each assay in a number of different sample matrices.
The sensitivity and specificity of each assay were then determined by
using a small panel of blinded spiked and nonspiked samples. All three
DELFIAs demonstrated at least 1 log greater sensitivity than
corresponding ELISAs utilizing the same reagents and showed an increase
in dynamic range of at least 2 log10 concentrations. This
increased LOD resulted in higher sensitivity rates for the DELFIA. The
specificity of all of the assays evaluated was 100%, and no sample
matrix effects were observed in either format. However, the
reproducibility of the DELFIA was poor due to randomly distributed
wells exhibiting excessive background signal (hot wells), which
occurred throughout the evaluation. As this technology matures, the
reproducibility of these assays should improve, as will the ability to
identify hot wells. Despite its sensitivity, the logistical burden
associated with the DELFIA and the technical expertise required to
complete assays and interpret the data limit the application of this
technology to reference or large clinical laboratories.
 |
INTRODUCTION |
Rapid diagnosis is essential for the
surveillance and control of epidemic diseases. Early diagnosis gives
health-care providers the information necessary for effective treatment
and control of disease outbreaks. Rapid identification of infectious
agents has also taken on greater importance in recent years with the increased threat of biological warfare and terrorism. Microorganisms such as Yersinia pestis and Venezuelan equine encephalitis
(VEE) virus, as well as toxins produced by bacteria such as
Staphylococcus aureus (staphyloccal enterotoxin B [SEB]),
can cause natural disease in humans, as well as be used as biological
weapons (5). Many of these agents can also be effectively
transmitted through contaminated food and water (3).
Because of the diverse symptomology and progression of disease
associated with different pathogens, detection of the causative agent
in individuals suspected to have been exposed to an agent requires
highly sensitive and specific means of detection. Aerosol infectious
doses for these agents are low. As little as 30 ng of SEB can
incapacitate a person, while as little 1.7 µg/person is lethal.
One-hundred Y. pestis organisms are enough to infect a
human, while even fewer VEE virus particles are required
(8). In addition to this challenge, a wide variety of
clinical specimens (nasal swabs, blood, serum, urine, or tissues) and
environmental samples (food, water, plant material, or soil) containing
even lower concentrations of agent may need to be tested to identify suspected pathogens. SEB may be found in urine as early as 12 h
after an aerosol exposure at concentrations of as low as 625 pg/ml by
our enzyme-linked immunosorbent assay (ELISA) and as early as
6 h by a newly developed electrochemiluminescent assay (100 pg/ml)
(C. A. Rossi, unpublished data). VEE virus was recovered from
serum as early as 72 h after aerosol exposure at concentrations detectable only by plaque assay (10 to 100 PFU) (B. Walker, unpublished data).
Several immunological methodologies have been applied to the antigenic
detection of microorganisms. The earliest methods utilized radioactive
isotopes. Although sensitive, the expense and hazards associated with
handling radioactive materials severely restricted the wide spread
application of these techniques. ELISAs were developed as an
alternative to radioactive assays. Typically, an ELISA can detect as
little as 100 to 500 pg of a low-molecular-weight analyte per ml or
106 to 107 organisms per ml and does not
require specialized handling procedures. Another alternative to the use
of radioactivity is the use of time-resolved fluorometry (TRF).
TRF has been used for a variety of applications, such as the detection
of antibodies, microorganisms, drugs, and other therapeutic agents
(2, 6, 10, 11, 22). Theoretically, TRF offers an increase
in sensitivity and a wider dynamic range than other assays such as the
ELISA. The basis of the TRF technology is the use of lanthanide chelate
labels that have unique fluorescence properties. The labels have an
intense long-lived fluorescence and a large stokes shift (~10 times
greater than that of fluorescein), all of which contribute to an
increased signal-to-noise ratio (9). These properties also
minimize the effect of any background fluorescence inherent in the
sample matrix. Four different chelates are available, each with its
unique narrow emission spectra, making multiplexing in a single well
possible. One format, dissociation-enhanced lanthanide fluorescent
immunoassay (DELFIA; Wallac Oy, Turku, Finland), has been applied to
the detection of analytes in various samples (2, 6, 10, 11, 15,
22). This method has the added advantage of increased
sensitivity due to efficient dissociation of the bound chelate (within
a few minutes) and the formation of long-lasting fluorescent micelles.
This dissociation occurs with the addition of a low-pH enhancement
solution (Perkin-Elmer-Wallac). Assay formats using DELFIA technology
include competitive and noncompetitive designs and are applicable for
protein-protein binding assays, ligand receptor binding studies, and
assays measuring the inhibition of enzyme activity.
The major emphasis of our laboratory is the development of rapid,
sensitive immunoassays which can be conducted in portable field medical
laboratories by personnel with limited training. The mobility and
degree of operator training dictates that equipment be simple to
operate and have a small footprint. Expendable supplies must be
minimized. Thus, a single immunoassay platform and assay format are
essential. Currently, we utilize ELISAs in our mobile laboratories.
These assays are very reproducible, easy to perform and interpret, and
use a single assay format to detect a large number of microorganisms.
However, due to the lack of sensitivity of these assays for at least
some microorganisms, an alternative detection technology is required.
Because of the great diversity of potential biological warfare agents
and endemic diseases that can be encountered, each potential platform
is evaluated for its ability to detect a broad range of microorganisms.
To eliminate potential differences between reagents, whenever possible,
an assay format using reagents identical to that of our existing ELISA
is used.
This study evaluated the ability of the DELFIA technology to detect SEB
toxin, Y. pestis-specific F1 antigen, and VEE virus in a
number of different sample matrices (phosphate-based buffer, serum, and
urine) compared to our previously developed, in-house horseradish
peroxidase (HRP)-based ELISA and to determine if this platform is
suited for use in portable field medical laboratories. Since this was a
preliminary evaluation, no attempt was made to achieve statistical
significance or to determine clinical sensitivity and specificity.
 |
MATERIALS AND METHODS |
Reagents.
Purified SEB toxin was provided by Toxinology
Division, U.S. Medical Research Institute of Infectious Disease
(USAMRIID), Fort Detrick, Md., and was stored at
20°C at 2 mg/ml
until used. An SEB/SED-specific monoclonal antibody (2B; Igen, Inc.,
Rockville, Md.) was used as the capture antibody, and polyclonal rabbit
serum (Toxin Technology, Inc., Madison, Wis.) was used for the
detection of SEB toxin.
Recombinant F1 antigen was provided by the Bacteriology Division,
USAMRIID, and was stored at
20°C at 0.7 mg/ml until used (1). Hyperimmune polyclonal rabbit serum was used as the
capture antibody and an F1-specific monoclonal antibody (6H3) was used for the detection of Y. pestis F1 antigen (both provided by
John Ezzell, USAMRIID).
VEE, Trinidad donkey strain, viral antigen was prepared from infected
hamster kidney (BHK) cells, inactivated by cobalt irradiation and
stored at
70°C until used (17). The antigen had a
titer of 1010 PFU/ml as determined by plaque assay
(18). Ascites fluid from an alphavirus-specific monoclonal
antibody (SLK-42; Alan Schmaljohn, USAMRIID) was used as the capture
antibody, and hyperimmune polyclonal rabbit serum prepared in our
laboratory was used for the detection of VEE antigen. Polyclonal rabbit
anti-VEE virus hyperimmune serum was prepared by purification of VEE on
sucrose gradients and immunization of rabbits with viable virus
harvested from these gradients combined with complete, and then
incomplete, Freund's adjuvant. All research was conducted in
compliance with the Animal Welfare Act and other Federal statutes and
regulations relating to animals and experiments involving animals and
adheres to principles stated in the Guide for the Care and Use of
Laboratory Animals (NIH publication 86-23, 1985 edition).
All antibodies were protein G purified as previously described
(14) and then lyophilized before being labeled with
europium for use in the DELFIA.
ELISA.
ELISAs previously developed in our laboratory for the
detection of these organisms were used in this study. The format was a
double-antibody sandwich identical to that previously described (16). This identical format is the basis for more than 29 antigen detection assays previously developed in our laboratory
(7, 16, 21). Optimal concentrations of capture and
detector antibodies specific for the assays described in our current
efforts were determined by serial cross-titrations of reagents. Capture
antibody was diluted in 0.01 M phosphate-buffered saline (pH 7.4; PBS), and 100 µl was adsorbed onto Maxisorb U-bottom 16-well microtiter strips (Nunc, Naperville, Ill.). Optimal coating dilutions were 1:1,000
for the SEB, 1:2,500 for the F1, and 1:500 for the VEE assays. Control
wells were similarly coated with negative nonspecific antibodies
produced in the same species as the specific antibodies. After an
overnight incubation at 4°C, the plates were washed three times (300 µl per well) with PBS supplemented with 0.1% Tween (PBS-T; Sigma,
St. Louis, Mo.). All subsequent antigen and antibody reagents added to
the plates were diluted in PBS-T containing 5% skim milk (Difco,
Detroit, Mich.). After the addition of each reagent, the plates were
incubated at 37°C for 1 h and then washed three times in PBS-T.
Antigen-spiked samples were added to both specific and nonspecific
antibody-coated wells. Bound antigen was detected by incubating
microtiter plates with detector antibody, followed by a HRP-labeled
species-specific antibody: goat anti-rabbit immunoglobulin G (IgG),
heavy and light chain specific for the SEB and VEE assays (catalog
number 074-1506; Kirkegaard and Perry, Gaithersburg, Md.), and goat
anti-mouse IgG, heavy and light chain specific for the plague F1 assay
(catalog number 605-250; Boehringer Mannheim, Indianapolis, Ind.). The
substrate used was 2,2'-azino-di(3-ethylbenzthiazoline sulfonate)
(Kirkegaard and Perry). Plates were read spectrophometrically at 410 nm
by using a 490-nm reference filter. Results were expressed as optical
density (OD) values.
DELFIA.
All equipment, platform-specific reagents, and
protocols were those specified by Perkin-Elmer-Wallac in order to
achieve optimal platform system performance. The DELFIA was performed
as the ELISA described above with the following exceptions. DELFIA
buffer (Perkin-Elmer-Wallac, Gaithersburg, Md.) was used to dilute the
antigen-spiked samples and detector antibody. Plates were vigorously
washed by using an OEM M12/2R Columbus plus washer (Tecan) between each
step. A wash step consisted of four cycles at 700 µl per well
programmed on overflow mode. These settings produced a vortex within
the well and resulted in thorough washing between steps and lower background signals. After the plates were coated, all subsequent incubations were done for 1 h at room temperature at low speed on a
model 1296-004 DELFIA Plateshaker (Perkin-Elmer-Wallac). Detector
antibodies were directly labeled with europium (Eu) either by
Perkin-Elmer-Wallac or by using a DELFIA Eu-labeling kit
(Perkin-Elmer-Wallac). Detector antibodies were reconstituted in
carbonate buffer at pH 9 and labeled with a 40-fold excess of europium
chelate according to the manufacturer's procedures. Unbound europium
was removed by gel filtration using a Sephadex G-25 column (Pharmacia,
Uppsala, Sweden). The protein concentrations were determined by using
MicroBCA assay (Pierce, Rockford, Ill.) according to the
manufacturer's instructions with bovine serum albumin as the standard.
Europium labeling efficiency was determined as recommended by
Perkin-Elmer-Wallac. Optimal concentrations of Eu-labeled detector
antibody were determined by serial checkerboard titrations. DELFIA
enhancement solution (100 µl/well) (Perkin-Elmer-Wallac) was used to
dissociate Eu3+ from solid-phase bound Eu-labeled detector
antibody. Plates were allowed to incubate for an additional 5 min at
room temperature on the shaker on low speed. Fluorescence was read
using the 1420 Victor2 multilabel counter
(Perkin-Elmer-Wallac). Emission peaks were captured as TRF counts at
613 nm from the top of the wells after excitation of the sample at 340 nm. Results were expressed as TRF counts.
Performance parameters.
Performance for each assay was
measured by determining the sensitivity or limit of detection (LOD) and
the reproducibility of the assay within a plate (intraplate variation)
and from day to day (interplate variation) for a number of different
sample matrices. The matrices evaluated included human serum (except for SEB assays), human urine, and PBS supplemented with 0.3% Tween 20 (PBS-0.3T). Serum used in these analyses did not contain any agent-specific antibodies as determined by antibody ELISA (data not
shown). Sample matrices were spiked with antigen and added at a 1:2
dilution in triplicate to the microtiter plate and diluted fourfold
down the microtiter plate to produce matrix-specific standard curves
and to determine intraassay variation. The linear range of the assay
was defined as the area of the standard curve that exhibited a
regression coefficient greater than 0.969. Wells in which unusually
high or low TRF counts or OD values occurred were examined for the
presence of an outlier value by applying the Dixon test
(4). These randomly distributed wells exhibiting excessive
background signal are commonly referred to as hot wells. Outliers or
hot wells successfully identified by this test were removed from
subsequent calculations. The Dixon test was only used to exclude sample
replicates from analysis. It was not used to exclude a sample from
analysis. The DELFIA was also run on three consecutive days to
determine interassay variation. In addition, a small panel of blinded
spiked and nonspiked samples representing the various sample matrices
were evaluated, in triplicate, to determine the relative assay
sensitivity and specificity. Sample sets were prepared by spiking each
set with SEB or F1 between 100 and 0.1 ng/ml in 10-fold increments or
concentrations of VEE virus of between 1 × 109 and
5 × 106 PFU/ml in half-log increments. Sample
concentrations were chosen based on what levels might be expected in
biological specimens taken from patients and the relative sensitivity
of the ELISA.
 |
RESULTS |
Europium labeling of antibodies.
Europium labeling of
antibodies resulted in molar incorporation ratios of 6.8 (F1), 9.4 (SEB), and 4.8 (VEE). There was no discernible loss of ELISA reactivity
for Eu-labeled antibodies compared to unlabeled antibodies (data not
shown). Optimal concentrations of Eu-labeled antibodies were determined
by serial checkerboard titrations and found to be 1 µg/ml for the SEB
toxin and VEE virus assays and 0.25 µg/ml for the F1 assay. In
comparison, optimal concentrations of ELISA detector antibodies were 2 µg/ml for the SEB toxin assay, 10 µg/ml for the F1 assay, and 20 µg/ml for the VEE virus assay.
ELISA.
As is typical for ELISA, adjusted OD values were
determined for each sample by subtracting negative capture antibody
values from positive capture antibody values. A positive ELISA value was determined by calculating a cutoff value that was equal to three
times the standard deviation of the matrix-specific negative controls
plus the average OD reading for the same controls, rounded up to the
nearest tenth. In the ELISA, cutoff values from samples prepared in all
three matrices were similar. No outliers were identified using the
Dixon test.
The SEB detection ELISA was linear between 10 and 1 ng/ml with an LOD
of 625 pg/ml (Fig. 1 and Table
1). This linear range is identical to
that observed in over 150 replicates of this assay conducted over 11 years (data not shown). Intraplate variation was <11% for values
within the linear portion of the curve (n = 3).
Interplate variation was not measured; however, coefficients of
variation for most ELISAs are typically between 15 and 25% (data not
shown). Variation between matrices was <20% (n = 6). Sensitivity of the assay was 75% (6 of 8) using blinded samples spiked
with SEB; however, it was 100% (6 of 6) for concentrations within the
LOD of the assay. The specificity was 100% (9 of 9).

View larger version (16K):
[in this window]
[in a new window]
|
FIG. 1.
SEB DELFIA and ELISA standard curves prepared in
different sample matrices. Urine and PBS-0.3T matrices were
spiked with 10,000 pg of SEB toxin per ml and diluted fourfold. Three
replicates of each concentration were used for each determination.
DELFIA results are expressed as TRF counts, whereas ELISA results are
expressed as OD values. Assay cutoff values were equal to three times
the standard deviation of negative control values (all matrices noted
above) plus the average value for the same controls. The ELISA cutoff
was then rounded up to the nearest tenth.
|
|
The linearity of the F1 detection ELISA is between 500 and 50 ng/ml and
was established in over 38 replicates of this assay conducted over 9 years (data not shown). The LOD of the assay was 25 ng/ml (Fig.
2 and Table 1). Intraplate variation
within the linear range was <13% (n = 3) and between
matrices was <9% (n = 9). The sensitivity of the
assay was 25% (3 of 12) using blinded samples spiked with F1 and 100%
(3 of 3) for concentrations within the LOD of the assay. The
specificity was 100% (9 of 9).

View larger version (18K):
[in this window]
[in a new window]
|
FIG. 2.
Y. pestis F1 antigen DELFIA and ELISA
standard curves prepared in different sample matrices. F1 antibody
negative serum, urine, and PBS-0.3T matrices were spiked with 25,000 pg
of F1 antigen per ml and diluted fourfold. Three replicates of each
concentration were used for each determination. DELFIA results are
expressed as counts, whereas ELISA results are expressed as OD values.
Assay cutoff values were equal to three times the standard deviation of
negative control values (all matrices noted above) plus the average
value for the same controls. The ELISA cutoff was then rounded up to
the nearest tenth.
|
|
The linearity of the VEE virus detection ELISA is between 5 × 108 and 5 × 107 PFU/ml and was
established in more than 40 replicates of this assay conducted over 12 years (data not shown). The LOD of the assay was 1.25 × 107 PFU/ml (Fig. 3 and Table
1). Intraplate variation was <20% (n = 3) and between
matrices was <9% (n = 9). The sensitivity of the
assay was 50% (6 of 12) using blinded samples spiked with VEE virus
and 100% (6 of 6) for concentrations within the LOD of the assay. The
specificity was 100% (9 of 9).

View larger version (17K):
[in this window]
[in a new window]
|
FIG. 3.
VEE virus DELFIA and ELISA standard curves prepared in
different sample matrices. VEE antibody-negative serum, urine, and
PBS-0.3T matrices were spiked with 5.00E + 07 PFU of VEE
virus per ml and diluted fourfold. Three replicates of each
concentration were used for each determination. DELFIA results are
expressed as TRF counts, whereas ELISA results are expressed as OD
values. Assay cutoff values were equal to three times the standard
deviation of negative control values (all matrices noted above) plus
the average value for the same controls. The ELISA cutoff was then
rounded up to the nearest tenth.
|
|
DELFIA.
As with the ELISA, adjusted TRF values were determined
for each sample, and cutoff values were calculated. In the DELFIA, cutoff values from samples prepared in all three matrices were similar.
A total of four wells were identified as outliers by the Dixon test,
and these were removed from subsequent calculations (two for F1
analysis and two for SEB analysis).
The SEB detection by DELFIA was linear between 10 and .039 ng/ml, with
an LOD of 39 pg/ml (Fig. 1 and Table 1). Assay linearity may have
covered a larger range, but higher concentrations of analyte were not
tested. Intraplate variation was <37% for values within the linear
portion of the curve (n = 3). Interplate variation was
<52% (n = 9). Variation between matrices was <33%
(n = 6). The sensitivity of the assay was 100% (8 of
8) using blinded samples spiked with SEB. All concentrations tested
were within the LOD of the assay. The specificity was 100% (9 of 9).
The F1 detection by DELFIA was linear between 25 and .024 ng/ml, with
an LOD of 24 pg/ml (Fig. 2 and Table 1). The assay linearity may have
covered a larger range, but higher concentrations of analyte were not
tested. Intraplate variation was <52% for values within the linear
portion of the curve (n = 3). Interplate variation was
<68% (n = 9). Variation between matrices was <26% (n = 9). The sensitivity of the assay was 91.7% (11 of
12) using blinded samples spiked with F1. All concentrations tested
were within the LOD of the assay. The specificity was 100% (9 of 9).
The VEE virus detection DELFIA was linear at between 5 × 107 and 3.13 × 106 PFU/ml, with an LOD of
3.13 × 106 PFU/ml (Fig. 3 and Table 1). The assay
linearity may have covered a larger range, but higher concentrations of
analyte were not tested. Intraplate variation was <18% for values
within the linear portion of the curve (n = 3). Interplate
variation was <30% (n = 9). Variation between
matrices was <25% (n = 9). The sensitivity of the
assay was 83.3% (10 of 12) using blinded samples spiked with VEE
virus. All concentrations tested were within the LOD of the assay. The
specificity was 100% (9 of 9).
 |
DISCUSSION |
DELFIA and ELISA assays were developed that can detect SEB toxin,
F1 antigen of Y. pestis, and VEE virus. Since a large
variety of sample matrices may be presented to the clinical laboratory for agent identification or patient diagnosis during a disease outbreak
or possible biowarfare scenario, a number of different sample matrices
(phosphate-based buffer, serum, and urine) were examined. Due to the
high prevalence of SEB antibodies in serum (
70% [12, 13, 19,
20]) and the rapid clearance of the toxin from the bloodstream
(8), serum is an inappropriate sample from which to
identify SEB toxin. Therefore, this matrix was not tested in our SEB experiments.
The ELISA format used was one that is an industry standard and has
proven to be simple and easy to perform. For simplicity and ease of
transition, we chose to develop a noncompetitive DELFIA format that
mimicked the ELISA. This strategy also enabled us to evaluate the
DELFIA platform by eliminating differences between reagents. While both
formats required an overnight incubation, the DELFIA required 80 min
less time to perform due to the use of a directly labeled detector
antibody and faster substrate development. However, the transition of
our ELISA to the DELFIA format was not straightforward. This assay
requires the use of a specialized washer, a plate shaker, special assay
buffers, and rigorous decontamination procedures. Experiments conducted
with standard ELISA equipment resulted in unacceptably high background
readings (data not shown). The assay requires a liter of wash buffer
per plate and dedicated pipetters to dispense enhancement solution. The
labeled detector antibody must be filtered through a 0.2-µm
(pore-size) filter just before addition to the plate. Despite these
precautions, a number of hot wells still occurred over the course of
this study. The majority of these hot wells were not identified as
outliers by the Dixon test. This resulted in the high DELFIA intra- and interplate variation that was observed and directly affects the reproducibility of the assay.
Eu labeling of antibodies was easy and simple using standard
chemistries, but unbound Eu tag must be totally removed to minimize assay background. The manufacturer suggests that labeling should be
conducted at a site not used for conducting the assay and should use
dedicated equipment and columns for this procedure. Labeling was
reproducible, as evidenced by the results from the labeling of two
different lots of SEB rabbit antibody (data not shown).
ELISA standard curves were linear over 1 log10
concentration, while DELFIA standard curves were linear over at least 3 log10 concentrations. This wide dynamic range eliminates
the need to run a large number of serial dilutions, thus allowing a
greater number of unknowns to be tested in a single assay.
The DELFIA assays were at least 1 log more sensitive than ELISA
utilizing the same reagents. This increased LOD resulted in higher
sensitivity rates for the DELFIA. The specificity of the assays were
100% when readings were adjusted to account for the reactivity of the
sample with a nonspecific capture antibody. However, in experiments
conducted without subtracting TRF counts from nonspecific capture
antibody wells and using our standard method for determining assay
cutoff values, false-positive results were seen (79%, 19 of
24) (data not shown). This finding indicates that major
modifications would be required to the way in which the cutoff is
calculated, which in turn would adversely affect the LOD of the DELFIA.
Use of these nonadjusted readings also resulted in substantial
differences between matrices (i.e., matrix effects [data not shown]).
In conclusion, the DELFIA was a rapid and highly sensitive method of
detection, which shows great promise in clinical diagnostic applications. Like the ELISA, it can detect analytes in a number of
different sample matrices, though at much lower concentrations. Currently, the many special requirements (equipment and technical) and
frequent hot wells make these assays cumbersome to perform and hard to
interpret, especially for junior laboratory personnel. These findings
are similar to those previously reported (10). Future
improvements in this technology should simplify this format and
potentially eliminate these problems. A more rigorous evaluation, including large numbers of spiked and clinically relevant samples, would be warranted at that time.
 |
ACKNOWLEDGMENTS |
We thank Tamara Lewis and Michael Robich for technical help and
Brian Walker and Jeff Teska for reviewing the manuscript.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: 1425 Porter St.,
Diagnostic Systems Division, U.S. Army Medical Research Institute of
Infectious Diseases, Fort Detrick, MD 21702. Phone: (301) 619-4776. Fax: (301) 619-2492. E-mail:
cindy.rossi{at}det.amedd.army.mil.
 |
REFERENCES |
| 1.
|
Andrews, G. P.,
D. G. Heath,
G. W. Anderson, Jr.,
S. L. Welkos, and A. M. Friedlander.
1996.
Fraction 1 capsular antigen (F1) purification from Yersinia pestis C092 and an Escherichia coli recombinant strain and efficacy against lethal plague challenge.
Infect. Immun.
64:2180-2187[Abstract].
|
| 2.
|
Barnard, G.,
B. Helmick,
S. Madden,
C. Gilbourne, and R. Patel.
2000.
The measurement of prion protein in bovine brain tissue using differential extraction and DELFIA® as a diagnostic test for BSE.
Luminescence
15:357-362[CrossRef][Medline].
|
| 3.
|
Benenson, A. (ed.).
1995.
Control of communicable diseases manual, 16th ed.
American Public Health Association, Washington, D.C.
|
| 4.
|
Bliss, C. I.
1967.
Statistics in biology, p. 153-155.
McGraw-Hill Book Company, New York, N.Y.
|
| 5.
|
Christopher, G. W.,
T. J. Cieslak,
J. W. Pavin, and E. M. Eitzen.
1997.
Biological warfare: a historical perspective.
JAMA
278:412-417[Abstract].
|
| 6.
|
Crooks, S. R.,
P. Ross,
C. S. Thompson,
S. A. Haggan, and C. T. Elliott.
2000.
Detection of unwanted residues of ivermectin in bovine milk by dissociation-enhanced lanthanide fluoroimmunoassay.
Luminescence
15:371-376[CrossRef][Medline].
|
| 7.
|
Dohm, D. J.,
T. M. Logan,
K. J. Linthicum,
C. A. Rossi, and M. J. Turell.
1996.
Transmission of Crimean-Congo hemorrhagic fever virus by Hyalomma impeltatum (Acari: Ixodidae) after experimental infection.
J. Med. Entomol.
33:848-851[Medline].
|
| 8.
|
Franz, D. R.,
P. B. Jahrling,
A. M. Friedlander,
D. J. McClain,
D. L. Hoover,
W. R. Bryne,
J. A. Pavin,
G. W. Christopher, and E. M. Eitzen.
1997.
Clinical recognition and management of patients exposed to biological warfare agents.
JAMA
278:399-411[Abstract].
|
| 9.
|
Hemmila, I. A.,
S. Dakubu,
V.-M. Mukkala,
H. Siitara, and T. Lovgren.
1984.
Europium as a label in time-resolved immunofluorometric assays.
Anal. Biochem.
137:335-343[CrossRef][Medline].
|
| 10.
|
Hierholzer, J. C.,
K. H. Johanasson,
L. J. Anderson,
C. J. Tsou, and P. E. Halonen.
1987.
Comparison of monoclonal time-resolved fluoroimmunoassay with monoclonal capture-biotinylated detector enzyme immunoassay for adenovirus antigen detection.
J. Clin. Microbiol.
25:1662-1667[Abstract/Free Full Text].
|
| 11.
|
Hierholzer, J. C.,
P. G. Bingham,
R. A. Coombs,
K. H. Johansson,
L. J. Anderson, and P. E. Halonen.
1989.
Comparison of monoclonal antibody time-resolved fluoroimmunoassay with monoclonal antibody capture-biotinylated detector enzyme immunoassay for respiratory syncytial virus and parainfluenza virus antigen detection.
J. Clin. Microbiol.
27:1243-1249[Abstract/Free Full Text].
|
| 12.
|
Jozefczyk, Z.
1974.
Specific human antibodies to enterotoxin A, B, and C1 of Staphylococcus: their increased synthesis in staphylococcal infection.
J. Infect. Dis.
130:1-7[Medline].
|
| 13.
|
Jozefczyk, Z.,
R. N. Robbins,
J. M. Spitz, and M. S. Bergdoll.
1980.
Antibodies to staphylococcal enterotoxin in laboratory personnel.
J. Clin. Microbiol.
11:438-439[Abstract/Free Full Text].
|
| 14.
|
Kijek, T. M.,
C. A. Rossi,
D. Moss,
R. W. Parker, and E. A. Henchal.
2000.
Rapid and sensitive immunomagnetic-electrochemiluminescent detection of staphyloccocal enterotoxin B.
J. Immunol. Methods
236:9-17[CrossRef][Medline].
|
| 15.
|
Knipping, G.,
B. Gogg-Fassolter,
B. Frohnwiesser,
F. Krempler,
G. M. Kostner, and E. Malle.
1997.
Quantification of apolipoprotein D by an immunoassay with time-resolved fluorescence spectroscopy.
J. Immunol. Methods
202:85-95[CrossRef][Medline].
|
| 16.
|
Ksiazek, T. G.,
P. E. Rollin,
P. B. Jahrling,
E. Johnson,
D. W. Dalgard, and C. J. Peters.
1992.
Enzyme immunosorbent assay for Ebola virus antigens in tissues of infected primates.
J. Clin. Microbiol.
30:947-950[Abstract/Free Full Text].
|
| 17.
|
Meegan, J. M.,
R. J. Yedlounschnig, and B. A. Peleg.
1981.
Enzyme-linked immunosorbent assay for detection of antibodies to Rift Valley fever virus in ovine and bovine sera.
Am. J. Vet. Res.
48:1139-1143.
|
| 18.
|
McClain, D. J.,
P. R. Pittman,
H. H. Ramsburg,
G. O. Nelson,
C. A. Rossi,
J. A. Mangiafico,
A. L. Schmaljohn, and F. J. Malinoski.
1998.
Immunological interference from sequential administration of live-attenuated alphavirus vaccine.
J. Infect. Dis.
177:634-641[Medline].
|
| 19.
|
McGann, V. G.,
J. B. Rollins, and D. W. Mason.
1971.
Evaluation of resistance to staphylococcal enterotoxin B: acquired antibodies of man and monkey.
J. Infect. Dis.
124:206-213[Medline].
|
| 20.
|
Notermans, S.,
W. J. Van Leeuwin,
J. Dufrenne, and M. S. Bergdoll.
1983.
Serum antibodies to enterotoxins produced by Staphylococcus aureus with specific reference to enterotoxin F and toxin shock syndrome.
J. Clin. Microbiol.
18:1055-1060[Abstract/Free Full Text].
|
| 21.
|
Turell, M. J.,
C. A. Rossi,
R. F. Tammariello, and C. L. Bailey.
1986.
Reduced recovery of Rift Valley Fever virus associated with assay of mosquito (Diptera: Culcidae) larval pools.
J. Med. Entomol.
23:416-422[Medline].
|
| 22.
|
Willson, V. J.,
W. J. Lockley,
A. Mather,
J. Singh,
C. M. Gilbert,
M. A. Bayliss, and D. Wilkinson.
2000.
A time-resolved fluorescence immunoassay for the determination of a novel respiratory therapeutic agent, AR-C68397XX (Vrozan) in human plasma.
J. Pharm. Biomed. Anal.
23:947-954[CrossRef][Medline].
|
Clinical and Diagnostic Laboratory Immunology, November 2001, p. 1070-1075, Vol. 8, No. 6
1071-412X/01/$04.00+0 DOI: 10.1128/CDLI.8.6.1070-1075.2001
This article has been cited by other articles:
-
Peruski, A. H., Peruski, L. F. Jr.
(2003). Immunological Methods for Detection and Identification of Infectious Disease and Biological Warfare Agents. CVI
10: 506-513
[Full Text]