Previous Article | Next Article ![]()
Clinical and Diagnostic Laboratory Immunology, May 2001, p. 588-592, Vol. 8, No. 3
Division of Bacterial and Mycotic Diseases,
National Center for Infectious Diseases, Centers for Disease
Control and Prevention, Atlanta, Georgia 30333
Received 6 November 2000/Returned for modification 6 February
2001/Accepted 27 February 2001
We compared the MRL and the Labsystems Chlamydia
pneumoniae microimmunofluorescence (MIF) immunoglobulin G (IgG)
kits and the Labsystems enzyme immunoassay (EIA) kit in a blinded study of 83 serum samples in which we evaluated titers, cross-reactivity to
other species, and reproducibility. There was no statistically significant difference between the MRL and the Labsystems MIF kits in
the endpoint titers of IgG antibody to C. pneumoniae. The correlation between the results obtained with these two MIF kits
was excellent (r = 0.95; P = 0.001).
The cross-reactivity of the C. pneumoniae-positive
sera with C. trachomatis- and C. psittaci-positive sera was assessed for each MIF kit. For
C. pneumoniae-positive sera with titers of While there are no wholly
satisfactory serologic methods for the diagnosis of Chlamydia
pneumoniae infections, the microimmunofluorescence (MIF) test,
when it is properly performed and when its results are properly read,
provides the most sensitive and species-specific method for laboratory
diagnosis of acute infection (3, 8). This test was
originally developed by Wang and Grayston (10) in 1970 for
detection of C. trachomatis antibodies. The MIF test is
an indirect fluorescent-antibody test that measures specific antibodies
to epitopes present in the cell walls of the elementary body (EB)
particles. Sensitivity and specificity can be improved by using
purified EBs of all three species of Chlamydia rather than
reticulate bodies, which predominantly express genus-specific epitopes.
The MIF test is the only antibody test available that measures the
titers of specific antibodies to all species simultaneously. The
disadvantages of the MIF test are that the endpoint fluorescence, or
the titer, is determined subjectively (8, 11), the test has low throughput, and the test requires proficiency and experience for correct reading of the endpoint titers.
Commercially available kits for detection of antibodies to
C. pneumoniae are available. MRL and Labsystems both
manufacture MIF kits. Cross-reactivity between species is reduced in
these kits by treating the EBs to remove genus-specific
lipopolysaccharide (LPS). Labsystems also offers an immunoglobulin G
(IgG) enzyme immunoassay (EIA) kit (also available in IgA and IgM
formats) for detection of antibodies to C. pneumoniae.
It could offer a high-throughput alternative to the MIF test if it
proves to be specific and sensitive. Reading of its endpoint is
objective, and it has the capacity to assay at least 200 serum samples
per day. It uses a non-LPS antigen that is reported to be specific for
C. pneumoniae. Its disadvantage is that the results are
reported as positive or negative enzymeimmuno units (EIU) on the basis of the results for a single dilution of a serum sample and not as an
endpoint titer; in addition, when low volumes are to be tested, the EIA
plates may in fact become costly due to waste.
Commercial kits offer advantages over in-house assays: they save time
and they provide quality control reagents for better reproducibility
within and between laboratories. Use of commercially available kits for
the MIF test obviates the need for a laboratory to grow and purify
chlamydia, a labor-intensive task in itself. Most importantly, as a
starting point they offer standardized quality control materials for
intra- and interlaboratory comparisons of MIF titers.
We tested and compared these commercial kits because of interest
generated in determination of C. pneumoniae antibody
titers in populations with chronic disease such as those with coronary artery disease and stroke and the need for standardized approaches and
methods for interlaboratory comparisons and interpretation of results.
Others have compared the Labsystems MIF and EIA kits (3a;
and K. Persson et al., personal communication), but this is the first
comparison of titers of IgG antibody to C. pneumoniae with the commercially available MRL and Labsystems MIF kits and the
Labsystems EIA kit. We determined the reproducibility of each of the
three kits, and because the MRL and Labsystems MIF kits simultaneously
measure antibody to all three species of Chlamydia, we
measured the cross-reactivity to other Chlamydia species
with each MIF kit. We found all three kits to be satisfactory for
qualitative detection of IgG antibodies to C. pneumoniae. We did not validate these kits in this study because
of a lack of appropriate clinical specimens, and we were unable to
determine the specificity of the EIA kit due to a lack of appropriate
sera from individuals with C. trachomatis and
C. psittaci infections.
MIF and EIA kits.
MRL and Labsystems provided all of the
commercial kits for this study. We tested serial twofold dilutions of
serum with the MRL and the Labsystems MIF kits until the endpoint titer
was determined. The Labsystems EIA kit package insert recommended use
of a single dilution of 1:101 for testing. If the values were off
scale, further dilutions were made and tested as indicated in the EIA
package insert.
Serum specimens.
Serum samples of twenty-four healthy donors
were obtained through blood services at the Centers for Disease Control
and Prevention (CDC), and testing of serum was covered under an
approved CDC Institutional Review Board protocol. These adult donors
self-reported no symptoms of respiratory infection for at least 4 weeks
before the first serum specimen was obtained. To simulate retrieval of specimens at the convalescent phase, the donors were asked to provide
for a second serum sample 4 weeks after retrieval of the first one; all
donors complied.
1071-412X/01/$04.00+0 DOI: 10.1128/CDLI.8.3.588-592.2001
Comparison of Two Commercial
Microimmunofluorescence Kits and an Enzyme Immunoassay Kit for
Detection of Serum Immunoglobulin G Antibodies to Chlamydia
pneumoniae
![]()
ABSTRACT
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
32, the
Labsystems MIF kit exhibited more cross-reactivity to C. psittaci than the MRL kit did. The values obtained with the
Labsystems EIA kit represented single dilutions of serum specimens
expressed as enzymeimmuno units on a continuous scale. The results
obtained with the Labsystems EIA kit correlated moderately well with
those obtained with each MIF kit when they were compared for their
abilities to detect IgG antibodies to C. pneumoniae
(for the MRL MIF kit, r = 0.79 [P = 0.001]; for the Labsystems MIF kit, r = 0.78 [P = 0.001]). The results obtained with the
commercial MRL and Labsystems MIF kits and the Labsystems EIA kit
tested were reproducible; and the kits were standardized, had quality
control reagents, and are suitable for detection of C. pneumoniae antibodies in serum and for use in interlaboratory studies. Validation of the use of these kits for clinical diagnosis still needs further evaluation.
![]()
INTRODUCTION
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
![]()
MATERIALS AND METHODS
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
MIF assays. Serial dilutions of the serum specimens were made in phosphate-buffered saline (pH 7.2), the diluted specimens were placed on slides of both the MRL and the Labsystems MIF kits, and then the slides were washed in parallel to reduce day-to-day variations. The specimens were handled by sterile technique and kept at 4°C for 2 months, the duration of the testing. When serum specimens are tested on different days, one can expect a level of variability of plus or minus 1 dilution with experienced readers.
All steps were carried out according to the manufacturers' instructions. MIF slides were incubated in humidified chambers at 37°C for 30 min with the diluted sera and again for 30 min with the conjugate. Slides were read on a Nikon Optiphot 2 fluorescent microscope with a ×40 objective. The endpoint titer was the last serum dilution that gave a definite uniform apple green fluorescence of 1+.EIA. The EIA, based on an indirect solid-phase EIA with horseradish peroxidase enzyme conjugate, was performed in accordance with the manufacturer's instructions. The appropriately diluted patient sera were applied to the C. pneumoniae antigen attached to the polystyrene surface of the Microstrip wells. Residual patient sample was removed by washing, and horseradish peroxidase-conjugated anti-human IgG was added. Unbound conjugate was washed off, and a colorless enzyme substrate containing the chromogen tetramethylbenzidine was added. The enzyme reaction with the chromogen resulted in a colored end product. A Labsystems Multiskan Ascent enzyme-linked immunosorbent assay reader was used to measure the color intensity at 450 nm. The optical density was proportional to the C. pneumoniae antibody level in the serum samples.
Statistical analyses.
The investigators were blinded as to
the dilutions used for all of the specimens, and the specimens were
tested in triplicate on different days with all three kits. In order to
run the statistical analyses for comparison of the discontinuous data
obtained with the MRL and the Labsystems MIF kits with the continuous
data obtained with the Labsystems EIA kit, we divided the MIF endpoint
titers into two groups:
32 or <32. An EIU value of 30, the cutoff
value of the assay, corresponds roughly to an MIF titer of 32 (Maritta Timonen, Labsystems, personal communication), which is also the limit
of sensitivity of the Labsystems MIF IgG kit.
| |
RESULTS |
|---|
|
|
|---|
Figure 1 is a scatter plot of the
median C. pneumoniae IgG titers from triplicate MIF
assays of the 83 individual serum samples tested with the MRL and the
Labsystems MIF kits. The distribution of endpoint titers with the
Labsystems MIF kit versus that with the MRL MIF kit showed that 3 (3.6%) samples had results that were 3 dilutions higher with the
Labsystems kit, 6 (7%) samples had results that were 2 dilutions
higher with the Labsystems kit, 35 (42%) samples had results that were
1 dilution higher with the Labsystems kit, 33 (40%) samples had
results that were identical with both the MRL and Labsystems MIF kits,
3 (3.6%) samples had results that were 4 dilutions higher with the MRL
kit, 2 (2.4%) samples had results that were 3 dilutions higher with
the MRL kit, and 1 (1%) sample had a result that was 2 dilutions
higher with the MRL kit. The correlation of the median IgG antibody
titers for the MRL and the Labsystems MIF kits with the Spearman
correlation coefficients was 0.95 (P = 0.0001),
demonstrating that the results between the MRL and the Labsystems MIF
kits are strongly correlated. Among the 83 individual serum samples
tested with each MIF kit, 33 (40%) had identical endpoint titers with
both kits. Sixty-two serum samples with titers of
32 and 16 serum
samples with titers of <32 were detected by both kits. Overall, the
endpoint titers obtained with the Labsystems kit tended to be
approximately 1 dilution higher than those obtained with the MRL kit,
but this difference was not statistically different (P = 0.36).
|
Table 1 compares the three kits for their
abilities to detect antibodies to C. pneumoniae. Table
1 is constructed such that one can identify how many of the same serum
samples gave similar results with any two kits. We grouped the endpoint
MIF titers into two groups (
32 and <32) for statistical comparisons
of the MIF test and the EIA, as described in Materials and Methods. We made no clinical inference from this cutoff. There was a moderately good correlation between the results obtained with each MIF kit and
those obtained with the EIA kit. The correlation of the Labsystems MIF
kit and the Labsystems EIA kit was 0.78 (P = 0.0001).
The correlation between the MRL MIF kit and the Labsystems EIA kit was
0.79 (P = 0.001). Among the 83 serum samples tested, 62 had titers of
32 and 16 had titers of <32 with both MIF kits
(results for 78 of 83 [94%] samples were in agreement), 59 had
titers of
32 and 11 had titers of <32 with both the EIA kit and the
Labsystems MIF kit (results for 70 of 83 [84%] samples were in
agreement), and 61 had titers of
32 and 10 had titers of <32 with
both the EIA kit and the MRL MIF kit (results for 71 of 83 [86%]
samples were in agreement).
|
The results showed high C. pneumoniae-specific IgG
titers in serum from healthy donors who reported no recent respiratory disease. Among the sera from these 24 donors, we found titers of
512
in 5 (21%) using the MRL kit and titers of >512 in 11 (46%) using
the Labsystems kit.
We compared the sera for reproducibility of endpoint titers in triplicate tests on different days with each MIF kit and with the EIA kit. The results obtained with both MIF kits were 100% reproducible because all triplicate serum endpoint titers with each MIF kit were within plus or minus 1 dilution, as expected. The mean coefficient of variation for the EIA kit was 10.2%, and the median coefficient of variation for the EIA kit was 8.6%, demonstrating good reproducibility.
Table 2 shows the cross-reactivity of the
62 C. pneumoniae-positive serum samples with titers of
32 common to both the MRL and the Labsystems MIF kits with
C. trachomatis- and C. psittaci EBs.
None of the serum samples from donors to blood services had titers of
32 for antibodies to all of the three species with the MRL MIF kit,
nor did any of the samples have titers of
32 for antibody to
C. psittaci. The sera with the highest cross-reactivity were from the clinical donor sites. There was no statistically significant difference between the two kits in terms of
cross-reactivity of the sera with C. trachomatis
(P
0.05). However, the difference in the
cross-reactivity with C. psittaci was statistically
significant: 61% of the serum samples cross-reacted when they were
tested with the Labsystems kit, and 15% of the serum samples
cross-reacted when they were tested with the MRL kit (P < 0.001 by both the Wilcoxon signed rank statistic and McNemar's
test). Of the 83 serum samples tested, 62 of the same serum samples had
titers of
32 for antibody to C. pneumoniae with both
kits and 16 of the same serum samples had titers of <32 with both
kits, giving a 94% agreement. Of the 62 C. pneumoniae-positive serum samples with titers of
32 with both
kits, the results for 90% of the C. trachomatis-positive sera were in agreement (24 C. trachomatis-positive serum samples with titers of
32 and 32 C. trachomatis-positive serum samples with titers of
<32), and for C. psittaci the agreement was 69% (9 C. psittaci-positive serum samples with titers of
32 and 34 C. psittaci-positive serum samples with titers
of <32).
|
We experienced some quality control problems with the dots on the Labsystems slides. All slides from both kits were washed together in the same carrier, but in one lot of Labsystems slides the dots did not stay in place and either curled up or lost EBs from the dot. The dots on the MRL slides were always stable.
| |
DISCUSSION |
|---|
|
|
|---|
This study compared the results obtained with two commercially available MIF kits and one commercial EIA kit for detection of IgG antibodies to C. pneumoniae. We compared IgG endpoint titers and measured the extent of cross-reactivity to other Chlamydia species for the C. pneumoniae-positive sera that had titers of >32 when they were tested with the MRL and the Labsystems MIF kits. We made no attempt to validate any of the assays because we lacked appropriate sera for that purpose, and none of the kits has received Food and Drug Administration approval for diagnostic purposes.
The three kits that we used were similar in their ability to detect IgG antibodies to C. pneumoniae when the individual titers in 83 serum samples were measured. The results obtained with the two MIF kits had an excellent correlation, and all three kits showed good reproducibility. We could not use specificity and sensitivity to compare the kits because sera from the laboratory from which PCR-positive samples were obtained had high titers in the both the acute phase and the convalescent phase.
An EIA kit could provide a promising alternative to MIF kits. The one that we tested compared moderately well to each of the MIF kits. We did not have enough properly timed serum samples from individuals with culture-confirmed infections or appropriate samples to validate the assay, nor did we have clinical C. trachomatis- and C. psittaci-positive sera to test for cross-reactivity. Another laboratory found considerable cross-reaction of C. trachomatis-positive sera with the C. pneumoniae IgG EIA kit (3a).
There is sufficient genus-reactive antigen exposed on the surfaces of the EBs to allow some cross-reactions in the MIF test. In addition to the major genus-specific LPS, the Chlamydia major outer membrane protein and other cell surface components contain both species- and genus-specific antigens, and serologic cross-reactions may be seen in both acute- and convalescent-phase samples. Species-specific antibodies also may be detected because of past exposure to other chlamydial species. The degree to which these cross-reactions are interpreted depends in part on the expertise of the microscopist (8) and on the antigen preparation. The cross-reactivity of one species with another in the MIF test was reported previously (1, 2, 4, 6, 9, 11). In one study of patients attending sexually transmitted disease clinics, antibodies reacting with two chlamydial antigens were found in 19 to 33% of patients, and 33 to 40% of serum samples reacted with antigens of all three species (1). In the same study in which sera from blood donors served as the control group, 76% of controls had antibodies to C. pneumoniae, but only 45% had antibodies to this species alone. Wong et al. (11) attempted to make the MIF test objective by using a time-resolved fluoroscopic immunoassay and eliminating the subjectivity of the microscopist. While the results of that test correlated well with those of the MIF test, the investigators still observed significant cross-reactions among the three chlamydial species, which may be attributed to a lack of removal of the LPS. The investigators estimate that one-third of the twofold dilution that makes up the final C. pneumoniae-specific antibody titer may be due to cross-reacting genus-specific epitopes.
The cross-reactivity between C. pneumoniae-positive
sera with titers of
32 and C. psittaci-positive sera
was greater with the Labsystems MIF kit than with the MRL MIF kit. This
difference may be due to the effectiveness with which the LPS was
removed from the EBs. The Labsystems LPS control is combined with the C. psittaci-positive dot because LPS is not removed
from these EBs. However, the MRL MIF kit has one dot in each well that
controls for LPS reactions, and this dot did not fluoresce for sera in which the C. psittaci antibodies were detected. EBs of
all three species should be included in every MIF test to exclude
endpoint titers that are due to cross-reactions.
We believe that paired serum samples collected at appropriate intervals should be used for the diagnosis of acute respiratory Chlamydia infections. Serodiagnosis based on the results obtained for a single serum sample is not recommended. The presence of a reaction in a single serum specimen may indicate previous exposure rather than a current infection (8). The half-life of specific IgG antibodies is reported to be as long as 3 years in some individuals (7). A significant percentage of donors with high specific antibody titers might indicate a population characterized not by chronic infection needing treatment but a population characterized by frequent past exposure (12), and high IgG titers can be present in sera from elderly individuals in the absence of clinically apparent disease (5). Even when the patient exhibits appropriate clinical symptoms, diagnosis of C. pneumoniae infection based on the results for a single serum specimen should be interpreted with caution. One should also be aware that serologic testing may be negative for a Chlamydia-infected individual.
Although serology provides a retrospective diagnosis because it requires both acute- and convalescent-phase serum specimens to show a fourfold rise in titer, the MIF test remains the best current method for serologic diagnosis of acute C. pneumoniae infections. Reproducible results were obtained with both the MRL and the Labsystems MIF kits. A fourfold rise in titer between paired serum specimens is diagnostic of acute infection, but all three species should be tested simultaneously because of occasional high levels of cross-reactivity. When cross-reactions are observed, the specific reaction will have an endpoint titer twofold or greater than the titers observed with the other two Chlamydia antigen spots (MRL kit package insert).
We found both the MRL and the Labsystems MIF kits used in the present study to be suitable for detection of endpoint titers of IgG antibody to C. pneumoniae, and there was a moderate correlation of the results obtained with the Labsystems EIA kit with those obtained with each MIF kit. While the MRL and the Labsystems MIF kits reduce the labor required for assay of specimens and increase the reproducibility and uniformity of the assay, accurate reading of the slides still requires a microscopist with experience and expertise. The advantage of the Labsystems EIA kit over the MRL and the Labsystems MIF kits is the high throughput when there are many serum samples to be analyzed. For the present, however, the MIF test is the method of choice for the detection of antibodies to C. pneumoniae.
| |
ACKNOWLEDGMENTS |
|---|
We thank Wayne Hogrefe of MRL and Rich Douglas of Labsystems for donating the MIF and EIA kits used in this study. We also thank Brian Plikaytis for help and advice with the statistics.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: 1600 Clifton Rd., MS G 05, Atlanta, GA 30333. Phone: (404) 639-1369. Fax: (404) 639-3123. E-mail: TMessmer{at}cdc.gov.
| |
REFERENCES |
|---|
|
|
|---|
| 1. | Bergstrom, K., M. Domeika, D. Vaitkiene, K. Persson, and P.-A. Mardh. 1996. Prevalence of Chlamydia trachomatis, Chlamydia psittaci, and Chlamydia pneumoniae antibodies in blood donors and attendees of STD clinics. Clin. Microbiol. Infect. 1:253-260[Medline]. |
| 2. |
Black, C. M.,
J. E. Johnson,
C. E. Farshy,
T. M. Brown, and B. P. Berdal.
1991.
Antigenic variation among strains of Chlamydia pneumoniae.
J. Clin. Microbiol.
29:1312-1316 |
| 3. | Ekman, M. R., M. Leinonen, H. Syrjala, E. Linnanmaki, P. Kujala, and P. Saikku. 1993. Evaluation of serological methods in the diagnosis of Chlamydia pneumoniae pneumonia during an epidemic in Finland. Eur. J. Clin. Microbiol. Infect. Dis. 12:756-760[CrossRef][Medline]. |
| 3a. | Gnarpe, J., J. Naas, and A. Lundback. 2000. Comparison of a new commercial EIA kit and the microimmunofluorescence technique for the determination of IgG and IgA antibodies to Chlamydia pneumoniae. APMIS 108:819-824[Medline]. |
| 4. |
Kern, D.,
M. A. Neill, and J. Schachter.
1993.
A seroepidemiologic study of Chlamydia pneumoniae in Rhode Island.
Chest
104:208-213 |
| 5. | Orr, P. H., R. W. Peeling, M. Fast, J. Brunka, H. Duckworth, G. K. Harding, and L. E. Nicolle. 1996. Serological study of responses to selected pathogens causing respiratory tract infection in the institutionalized elderly. Clin. Infect. Dis. 23:1240-1245[Medline]. |
| 6. | Ozanne, G., and J. Lefebvre. 1992. Specificity of the microimmunofluorescence assay for the serodiagnosis of Chlamydia pneumoniae infections. Can. J. Microbiol. 38:1185-1189[Medline]. |
| 7. | Patnode, D., S.-P. Wang, and J. T. Grayston. 1990. Persistence of Chlamydia pneumoniae, strain TWAR, micro-immunofluorescent antibody, p. 406-409. In W. R. Bowie, H. D. Caldwell, R. P. Jones, et al. (ed.), Chlamydial infections. Cambridge University Press, Cambridge, England. |
| 8. | Schachter, J. 1992. Chlamydiae, p. 661-666. In N. R. Rose, E. C. DeMarcario, J. L. Fahey, H. Friedman, and G. M. Pen (ed.), Manual of clinical laboratory immunology, 4th ed. American Society for Microbiology, Washington, D.C. |
| 9. | Wagenvoort, J. H. T., D. Koumans, and M. van de Cruijs. 1999. How useful is the Chlamydia micro-immunofluorescence (MIF) test for the gynaecologist? Eur. J. Obstet. Gynecol. Reprod. Biol. 84:13-15[CrossRef][Medline]. |
| 10. | Wang, S.-P., and J. T. Grayston. 1970. Immunologic relationship between genital TRIC, lymphogranuloma venereum, and related organisms in a new microtiter indirect immunofluorescence test. Am. J. Ophthal. 70:367-374[Medline]. |
| 11. | Wong, Y.-K., J. M. Sueur, C. H. D. Fall, J. Orfila, and M. E. Ward. 1999. The species specificity of the microimmunofluorescence antibody test and comparisons with a time resolved fluoroscopic immunoassay for measuring IgG antibodies against Chlamydia pneumoniae. J. Clin. Pathol. 52:99-102[Abstract]. |
| 12. |
Wong, Y.-K.,
P. J. Gallagher, and M. E. Ward.
1999.
Chlamydia pneumoniae and atherosclerosis.
Heart
81:232-238 |
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Antimicrob. Agents Chemother. | Clin. Microbiol. Rev. | Infect. Immun. |
|---|---|---|
| J. Clin. Microbiol. | J. Virol. | ALL ASM JOURNALS |