Previous Article | Next Article 
Clinical and Diagnostic Laboratory Immunology, November 2000, p. 882-884, Vol. 7, No. 6
1071-412X/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Use of a Dry-Plasma Collection Device to Overcome
Problems with Storage and Transportation of Blood Samples for
Epidemiology Studies in Developing Countries
Zhannat Z.
Nurgalieva,1
R.
Almuchambetova,2
A.
Machmudova,2
D.
Kapsultanova,2
Michael S.
Osato,1
Jeffrey
Peacock,3
Richard P.
Zoltek,4
Patrice A.
Marchildon,3
David Y.
Graham,1,* and
Abai
Zhangabylov2
Veterans Affairs Medical Center and Departments of Medicine and
Molecular Virology and Microbiology, Baylor College of Medicine,
Houston, Texas1; and EPI, Stony Brook, New
York3; Chematics, Inc., North Webster,
Indiana4; and Department of Faculty Therapy, Almaty
State Medical University, Almaty, Kazakhstan2
Received 3 April 2000/Returned for modification 14 June
2000/Accepted 25 July 2000
 |
ABSTRACT |
Studies are difficult in areas lacking modern facilities due to the
inability to reliably collect, store, and ship samples. Thus, we sought
to evaluate the use of a dry plasma collection device for
seroepidemiology studies. Plasma was obtained by fingerstick using a
commercial dry plasma collection device (Chemcard Plasma Collection
Device) and serum (venipuncture) from individuals in Kazakhstan. Plasma
samples were air dried for 15 min and then stored desiccated in foil
zip-lock pouches at 4 to 6°C and subsequently shipped to the United
States by air at ambient temperature. Serum samples remained frozen at
20°C until assayed. Helicobacter pylori status was
determined by enzyme-linked immunosorbent assay (HM-CAP EIA) for the
dry plasma and the serum samples. The results were concordant in 250 of
the 289 cases (86.5%). In 25 cases (8.6%), the dry plasma samples
gave indeterminate results and could not be retested because only one
sample was collected. Five serum samples were positive, and the
corresponding dry plasma samples were negative; one serum sample was
negative, and the corresponding plasma sample was positive. The
relative sensitivity and specificity of the Chemcard samples to serum
were 97.6 and 97.9%, respectively, excluding those with indeterminate
results. Repeated freeze-thawing had no adverse effect on the accuracy
of the test. We found the dry plasma collection device to provide an
accurate and practical alternative to serum when venipuncture may be
difficult or inconvenient and sample storage and handling present
difficulties, especially for seroepidemiologic studies in rural areas
or developing countries and where freeze-thawing may be unavoidable.
 |
INTRODUCTION |
Helicobacter pylori is a
major human pathogen that is etiologically linked to gastritis, peptic
ulcer disease, gastric adenocarcinoma, and primary gastric B-cell
lymphoma (1, 3). There is continued interest in the
epidemiology of H. pylori infection in order to better
characterize the prevalence of infection, as well as the natural
history and mode of transmission of the infection (1, 3).
One problem with obtaining specimens from developing countries, where
the infection is most common, has been the ability to reliably collect,
store, and assay serum or plasma samples (N. Broutet, G. Duperrex, B. Bergery, and F. Megraud, Letter, Lancet 354:1529-1530, 1999). The problems relating to storage and transport of these samples
have often been the limiting factor in determining which questions can
be addressed in seroepidemiologic studies, especially in areas lacking
modern medical facilities. Shipment of frozen serum samples is
expensive, requiring the use of dry ice and expedited shipping
schedules, and requires compliance with national and international
regulations governing the shipment of biohazardous materials. Delays
are not uncommon and can result in compromised samples. A method that
eliminated the need to store and ship frozen serum would therefore be
welcome. We evaluated the feasibility of using a simple device that
allows collection of dry plasma from a fingerstick that does not
require freezing for storage or shipment and does not require
biohazardous clearance for shipping or handling for obtaining blood
samples in Kazakhstan.
 |
MATERIALS AND METHODS |
We evaluated the utility of the dried plasma device using the
commercially available HM-CAP EIA (Enteric Products, Inc., Stony Brook,
N.Y.). We compared results obtained when the dry plasma sample was
rehydrated and assayed via the HM-CAP EIA to results obtained using a
conventional serum sample collected from the same patient in order to
determine if use of the dry plasma device could be an accurate,
reproducible alternative to the use of serum.
Blood sample collection.
Serum and plasma samples were
collected by local physicians in Kazakhstan using a patented dry plasma
collection device (Chemcard; Chematics, Inc., North Webster, Ind.)
which consists of a laminate of a semipermeable membrane, through which
blood cellular and particulate matter cannot pass. This membrane is
over a second membrane designed to absorb a measured amount of plasma.
A hanging drop of blood obtained via fingerstick was touched to the
test area of the commercial dry plasma collection device. The correct amount of blood applied was signified by a change in the color from
white to red of the integrated control, indicating when an adequate
volume of blood had been applied to the card. The top filter was
removed after 3 min, and the card was air dried for 15 min. The
resultant dried plasma sample was then stored in a desiccated zip-lock
pouch at between 4 and 6°C for up to 2 months before being shipped to
the United States by air at ambient temperature.
A venous blood sample was also obtained from each individual at the
same time that the plasma sample was obtained. The blood was allowed to
clot, and the serum was separated by centrifugation. Sera were stored
at
20°C and then shipped frozen to the United States for analysis.
ELISA test procedure.
The HM-CAP EIA for immunoglobulin G
antibodies for H. pylori was performed according to the
manufacturer's instructions. The plasma collection pad was removed
from the card and placed in the bottom of a test tube. A portion (170 µl) of EIA wash buffer was added to each tube so that the collection
pad was totally immersed in the wash buffer, and this was hydrated for
10 min. The collection pad and wash buffer were vortexed three times
for 10 s each time. The wash buffer used for the extraction of the plasma was assayed in the HM-CAP EIA as if it were a diluted serum sample (i.e., 100 µl of the wash buffer containing the plasma was
added to each test well of the microwell plate, and the EIA was
performed in accordance with the manufacturer's instructions). All
samples were assayed within 30 min to 1 h of rehydration and extraction. Serum and card samples were evaluated on separate assays on
separate days. For the HM-CAP EIA, quantitative enzyme-linked immunosorbent assay (ELISA) values (EV) were extrapolated and interpreted for each sample, for either serum or dried plasma, according to the manufacturer's instructions.
Reproducibility.
Six plasma card samples were obtained from
each of three patients. These patients had previously been demonstrated
to have H. pylori antibody titers which were negative, low
positive, and mid-range positive as judged by HM-CAP analysis of serum
samples obtained by venipuncture. All six samples were obtained on the same day. The dried plasma samples were stored desiccated at 4 and
6°C until rehydration, extraction, and analysis on the HM-CAP EIA.
The plasma samples were assayed as follows: samples 1 and 2 were
evaluated in tandem on the same assay at 2 weeks, samples 3 and 4 were
evaluated in tandem on the same assay at 1 month, and samples 5 and 6 were analyzed in tandem on the same assay at 3 months.
Effect of freeze-thawing.
Three EDTA whole-blood samples
giving plasma card results in the range of high-negative-indeterminate,
low-positive, and high-positive EV on EIA were used. For each sample, a
total of 10 plasma cards were spotted. The dried plasma samples thus
obtained were stored in a desiccated zip-lock pouch. One sample at each
EV was stored at 2 to 8°C for a total of 7 days. The remaining seven
were frozen at
20°C overnight. The following day, all seven cards
for each were removed to 2 to 8°C for a minimum of 4 h to thaw.
One card was left at 2 to 8°C for the remainder of the experiment.
The remaining six were refrozen at
20°C overnight. This pattern was continued until all the plasma cards were at 2 to 8°C, each having undergone 1, 2, 3, 4, 5, 6, or 7 freeze-thaw cycles. On the final day,
all 10 pouched plasma cards for each EV level were removed to room
temperature, rehydrated, and assayed in tandem on the HM-CAP EIA assay.
The reproducibility of results for the three which were never frozen
and thawed versus the remaining seven which had undergone freeze-thaw
cycles 1 through 7 was evaluated.
 |
RESULTS |
H. pylori infection was defined as a positive HM-CAP
EIA from the frozen serum sample. Values of <1.8 were scored as
negative, and those that were >2.2 were scored as positive. Values of
1.8 to 2.2 were considered indeterminate and repeated as directed by
the manufacturer. If the repeat result was positive, the sample was
scored as positive, and if negative, the sample was scored as negative.
If the repeat result was indeterminate, the sample was scored as indeterminate.
A total of 289 simultaneously obtained plasma and serum samples were
collected. Of these, 204 simultaneously obtained pairs of plasma and
serum samples were positive and 46 were negative on the HM-CAP EIA.
Five serum samples were positive, whereas the corresponding plasma
samples were negative. One serum sample was negative, whereas the
corresponding plasma sample was positive. Insufficient material
remained to retest the plasma samples that yielded indeterminate
results, and so the 25 indeterminate plasma results (8.6% of the
samples) were excluded from the comparative analysis. The relative
sensitivity and relative specificity of the dried plasma samples
compared to the serum samples were 97.6 and 97.9%, respectively,
excluding the indeterminate samples.
Of the 25 indeterminate plasma sample results, 17 had corresponding
serum EV results in the low-positive range of <3.0. One of the
twenty-five had a corresponding serum result in the high-negative range
of 1.4 EV, and two had corresponding serum results which were also
indeterminate. Of eight sera yielding indeterminate results, two had
corresponding plasma card results in the high-negative range of
1.0 EV.
Reproducibility.
Results obtained using the dried plasma card
were very reproducible, as shown in Table
1. Duplicate samples assayed in the HM-CAP EIA in tandem were reproducible within 0.1 EV or 6.0%. All six
samples gave a standard deviation of 0.2 EV or less. These results are
consistent with the HM-CAP assay reproducibility, as stated in the
manufacturer's package insert.
Results of repeated freeze-thawing.
The EV of each sample
having undergone freeze-thaw cycles 1 through 7 were comparable to the
mean of the three samples that were never frozen (Table
2). Results were reproducible within a
standard deviation of 0.2 EV and/or coefficient of variation of <6%
for each EV level.
 |
DISCUSSION |
The results of the two methods of blood collection were
comparable, with a specificity and sensitivity of >97%. The
comparative analysis showed that the two collection methods were highly
reliable and reproducible, with 250 of 256 determinations having
essentially identical HM-CAP EIA results. There were only six instances
(2.3%) in which the results differed between the collection methods. The only disadvantage to the Chemcard (compared to serum) is that samples yielding indeterminate results cannot be retested. This can be
overcome by collection of two or more samples per patient, and this is
recommended for future studies. In this study we collected only one
sample and those with indeterminate values could not be retested. The
results of this comparative study in Kazakhstan are similar to those
obtained in the carefully controlled conditions in the United States in
which 84 patients were tested and the results were compared to the
HM-CAP (2), despite the potential problems associated with
collection and storage in an underdeveloped country.
Problems related to blood collection, storage, and shipping are the
major impediments to conducting seroepidemiologic studies in areas
where modern facilities are lacking. In rural areas, storage and
transportation are often the critical elements responsible for the
failure or success of a study, and the failure of any support system
(refrigerator, freezer) or a delay in transit may result in the loss of
valuable specimens and necessitate the recollection of samples (Broutet
et al., Letter).
Dried-blood collection has been previously recognized as a rapid,
convenient, and inexpensive means of collecting samples for later
quantitative laboratory analysis and is especially useful where
collection of a blood sample is more difficult and/or less desirable,
for example, in infants (4-8; G. F. Batstone, E. J. Coombes,
and T. R. Gamlen, Letter, Lancet 2:99, 1984; J. V. Kenny, Letter, Trop. Doct. 23:128, 1993; S. O'Shea, J. Mullen, K. Corbett, I. Chrystie, M. L. Newell, and J. E. Banatvala, Letter, AIDS 13:630-631, 1999; G. T. Werner, G. G. Frosner, and C. Epp, Letter, Trans. R. Soc. Trop.
Med. Hyg. 79:135-136, 1985). Typically, a standardized area
of the dried-blood sample is punched out and rehydrated in an attempt
to obtain a relatively standardized sample. The Chemcard collects a
standardized amount of plasma without contaminating red blood. Plasma,
unlike serum, is stable at ambient temperature and thus may allow for
the archiving of samples for longitudinal studies and possibly for
comparison of acute and convalescent or quantitative laboratory
analysis, which is not possible in rapid test devices.
The Chemcard allowed collection of plasma samples with little
extraneous equipment (e.g., a fingerstick apparatus such as disposable
lancets) and could be performed with minimal training or discomfort.
The dried sample was also simple to handle, store, and transport, and
shipping costs were minimal because the samples were compact and could
be shipped at ambient temperature, thus producing a saving both in
convenience and in shipping costs. In addition, the results of the
freeze-thaw experiment showed that this method is very robust.
Together, these features of the Chemcard make it highly desirable for
seroepidemiologic studies in both developed and developing countries.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Veterans Affairs
Medical Center, 2002 Holcombe Blvd. (111-D), Houston, TX 77030. Phone: (713) 791-1414. Fax: (713) 790-1040. E-mail:
dgrahamm{at}bcm.tmc.edu.
 |
REFERENCES |
| 1.
|
Breuer, T.,
H. M. Malaty, and D. Y. Graham.
1997.
The epidemiology of H. pylori-associated gastroduodenal diseases, p. 1-14.
In
P. Ernst, P. Michetti, and P. D. Smith (ed.), The immunobiology of Helicobacter pylori from pathogenesis to prevention. Lippincott-Raven, Philadelphia, Pa.
|
| 2.
|
Kearney, D. J.,
L. Boes, and J. S. Peacock.
1999.
Use of a dried plasma collection card for simplified diagnosis of Helicobacter pylori infection.
Aliment. Pharmacol. Ther.
13:1531-1534[Medline].
|
| 3.
|
Malaty, H. M., and D. Y. Graham.
1999.
Epidemiology of Helicobacter pylori infection: disease management implications for peptic ulcer disease.
Dis. Manage. Health Outcomes
6:9-18.
|
| 4.
|
McCarron, B.,
R. Fox,
K. Wilson,
S. Cameron,
J. McMenamin,
G. McGregor,
A. Pithie, and D. Goldberg.
1999.
Hepatitis C antibody detection in dried blood spots.
J. Viral. Hepat.
6:453-456[CrossRef][Medline].
|
| 5.
|
O'Broin, S. D., and E. W. Gunter.
1999.
Screening of folate status with use of dried blood spots on filter paper.
Am. J. Clin. Nutr.
70:359-367[Abstract/Free Full Text].
|
| 6.
|
Parker, S. P., and W. D. Cubitt.
1999.
The use of the dried blood spot sample in epidemiological studies.
J. Clin. Pathol.
52:633-639[Medline].
|
| 7.
|
Parker, S. P.,
H. I. Khan, and W. D. Cubitt.
1999.
Detection of antibodies to hepatitis C virus in dried blood spot samples from mothers and their offspring in Lahore, Pakistan.
J. Clin. Microbiol.
37:2061-2063[Abstract/Free Full Text].
|
| 8.
|
Perez Guevara, M. T.,
F. M. Rolo Gomez,
C. Nibot Sanchez,
O. Cruz Sui, and O. Rodriguez.
1998.
The determination of antibodies to the human immunodeficiency virus type 1 (HIV-1) in samples of dried blood on filter paper.
Rev. Cubana Med. Trop.
50:93-95[Medline].
|
Clinical and Diagnostic Laboratory Immunology, November 2000, p. 882-884, Vol. 7, No. 6
1071-412X/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.