Previous Article | Next Article 
Clinical and Diagnostic Laboratory Immunology, May 2000, p. 344-351, Vol. 7, No. 3
1071-412X/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Multisite Comparison of CD4 and CD8 T-Lymphocyte Counting by
Single- versus Multiple-Platform Methodologies: Evaluation of Beckman
Coulter Flow-Count Fluorospheres and the tetraONE System
Keith A.
Reimann,1,*
Maurice R. G.
O'Gorman,2
John
Spritzler,3
Cynthia L.
Wilkening,3
Daniel E.
Sabath,4
Karen
Helm,5
Donald E.
Campbell,6 and
The NIAID
DAIDS New Technologies Evaluation Group
Division of Viral Pathogenesis, Beth Israel
Deaconess Medical Center, Harvard Medical
School,1 and Harvard School of Public
Health,3 Boston, Massachusetts;
Children's Memorial Hospital, Northwestern University,
Chicago, Illinois2; University of
Washington, Seattle, Washington4;
University of Colorado Health Sciences Center, Denver,
Colorado5; and Children's Hospital of
Philadelphia, Philadelphia, Pennsylvania6
Received 22 February 1999/Returned for modification 23 April
1999/Accepted 18 August 1999
 |
ABSTRACT |
New analytic methods that permit absolute CD4 and CD8 T-cell
determinations to be performed entirely on the flow cytometer have the
potential for improving assay precision and accuracy. In a multisite
trial, we compared two different single-platform assay methods with a
predicate two-color assay in which the absolute lymphocyte count was
derived by conventional hematology. A two-color method employing
lymphocyte light scatter gating and Beckman Coulter Flow-Count
fluorospheres for absolute counting produced within-laboratory precision equivalent to that of the two-color predicate method, as
measured by coefficient of variation of replicate measurements. The
fully automated Beckman Coulter tetraONE System four-color assay
employing CD45 lymphocyte gating, automated analysis, and absolute
counting by fluorospheres resulted in a small but significant improvement in the within-laboratory precision of CD4 and CD8 cell
counts and percentages suggesting that the CD45 lymphocyte gating and
automated analysis might have contributed to the improved performance.
Both the two-color method employing Flow-Count fluorospheres and the
four-color tetraONE System provided significant and substantial improvements in between-laboratory precision of absolute counts. In
some laboratories, absolute counts obtained by the single-platform methods showed small but consistent differences relative to the predicate method. Comparison of each laboratory's absolute counts with
the five-laboratory median value suggested that these differences resulted from a bias in the absolute lymphocyte count obtained from the
hematology instrument in some laboratories. These results demonstrate
the potential for single-platform assay methods to improve
within-laboratory and between-laboratory precision of CD4 and CD8
T-cell determinations compared with conventional assay methods.
 |
INTRODUCTION |
The progressive loss of
CD4+ T lymphocytes (CD4 T cells) through virally mediated
cell destruction is the predominant pathophysiological manifestation of
human immunodeficiency virus type 1 (HIV-1) infection (15).
Enumeration of this cell subset provides an estimate of HIV disease
progression (17). A major component of the immune response
to HIV-1 infection is mediated by CD8+ T lymphocytes (CD8 T
cells) (6). CD8 T cells capable of suppressing HIV-1
replication adopt an activation phenotype and appear in increased
numbers in the blood and other body compartments following infection
(7, 19). Therefore, the CD4 and CD8 T cells in blood are
frequently quantified to assess immune competence and disease stage in
HIV-1-infected patients. Furthermore, changes in CD4 T-cell numbers are
an important estimate of the response to antiretroviral therapy. The
CD4 T-cell count remains the most important immunologic surrogate
marker of the efficacy of new antiretroviral regimens used in clinical
trial evaluations (9). The accurate quantitation of these
cells in blood is crucial for providing clinical care to HIV-1-infected
patients and for the systematic evaluation of new therapeutic
modalities. However, previous studies have identified substantial
variability in results between laboratories performing these assays
(1, 5).
The currently recommended method for CD4 T-cell determination (2,
3) utilizes three independently derived values from two different
instruments: a white blood cell (WBC) count and percent lymphocytes
derived from a hematology instrument, and percentage of CD4 or CD8 T
cells derived from a flow cytometer. A major disadvantage of this
multiple-platform assay method is that error in each independent
measurement is multiplied at each subsequent step in the calculation.
Recently, new analytic methods have emerged that permit absolute CD4
and CD8 T-cell determinations to be performed either using a
single-determination, non-flow cytometry-based assay (8, 10,
11) or entirely on the flow cytometer (14). In the
single-platform flow cytometry-based techniques, fluorospheres
are added to the blood at a known concentration. Absolute cell counts
in each specimen can be calculated ratiometrically by simultaneously
counting both fluorospheres and the cells of interest (18).
Other technological improvements, including automated instrument setup,
lymphocyte gating, and cursor placement, could also improve assay
performance. In the present study, we have assessed the
within-laboratory and between-laboratory precision of two alternative
assay methods: a two-color method that uses Beckman Coulter Flow-Count
fluorospheres to determine absolute counts, and a four-color method
that uses the fully automated tetraONE System. The precision obtained
with these alternative methods was compared with the precision obtained with a conventional multiple-platform assay method. We also assessed the accuracy of these two new assay methods by comparing the absolute counts obtained with the single-platform methods and with the conventional multiple-platform method.
 |
MATERIALS AND METHODS |
Sites.
All five sites that participated in this study were
clinical flow cytometry laboratories that performed lymphocyte
immunophenotyping for the AIDS Clinical Trials Group and were certified
by the National Institute of Allergy and Infectious Disease (NIAID)
Division of AIDS Flow Cytometry Certification Program (2).
All laboratories had extensive experience in performing clinical
immunophenotyping for CD4+ and CD8+ T
lymphocytes by conventional methods but had no prior experience with
the new, single-platform methods evaluated in this study. Before
participating in this study, workers in each laboratory underwent a
2-day on-site training in the new methods by a representative from
Beckman Coulter. At each site, assays were performed by one laboratory
technologist except for lab D, where study data were generated by three
different technologists.
Instrumentation.
The hematology instruments used for
determining absolute lymphocyte counts varied at the five sites and
included a Coulter STK-S, a Technicon H2, a Sysmex HST430, and an
Abbott CellDyn 3500 at labs A and D, B, C, and E, respectively. All
laboratories performed flow cytometric analyses with the Beckman
Coulter EPICS XL flow cytometer (Beckman Coulter, Miami, Fla.) using
System II software and tetraONE System (version 1.0; Beckman Coulter). Specimen preparation was performed on the Multi-Q-Prep workstation (Beckman Coulter). All pipetting was performed with a
positive-displacement Eppendorf Repeater pipetter (Brinkmann
Instruments, Westbury, N.Y.).
Reagents.
All reagents were provided by Beckman Coulter and
used as recommended by the manufacturer.
Specimens.
To measure within-laboratory precision, specimens
were obtained locally at each site from HIV-1-infected donors after
informed consent was obtained (hereafter referred to as local
specimens). Laboratories were requested to obtain specimens from seven
donors with CD4 T cell counts of <200 cells/µl and seven with CD4 T
cell counts of >200 cells/µl. Blood was drawn directly into
evacuated blood collection tubes containing EDTA and delivered to the
laboratory within 1 h. To measure between-laboratory precision,
specimens were obtained from HIV-1-infected donors by a central
contractor (FAST Systems, Inc., Gaithersburg, Md.). We attempted to
obtain an equivalent number of specimens representing the following
three CD4 T-cell count strata: <200 cells, 200 to 500 cells, and >500 cells. Blood was drawn directly into evacuated blood collection tubes
containing EDTA and then shipped at ambient temperature by overnight
courier service to each site (hereafter referred to as central
specimens). Confirmation of HIV-1 infection status was made by routine
clinical serology. The number of specimens analyzed in the comparison
of within-laboratory precision and between-laboratory precision studies
was chosen to provide an 80% power to detect shifts in percent
coefficient of variation (%CV) or in median CD4 or CD8 T-cell counts.
These calculations were based on variability estimates from a previous
study (11).
Flow cytometric immunophenotyping for CD4+ and
CD8+ T lymphocytes.
All local and central blood
specimens were analyzed by three different methods (summarized in Table
1). Each site performed these assays in
one of three different orders for the entire study.
(i) Predicate method.
Absolute lymphocyte counts were
determined as the product of the WBC count and percent lymphocyte
differential as measured by the hematology instruments. Flow cytometer
alignment, calibration, and spectral compensation were performed
according to each individual laboratory's operating procedure. For
specimen preparation, 100 µl of EDTA-anticoagulated blood was
incubated with premixed antibodies for 20 min and erythrocytes were
lysed with the ImmunoPrep reagent system (Beckman Coulter) on a
Multi-Q-Prep work station and analyzed within 6 h of lysis. The
percentage of lymphocytes expressing CD3 and CD4 or CD3 and CD8 was
determined according to the published guidelines for Flow Cytometric
Immunophenotyping (2, 3). A lymphocyte light scatter gate
(scattergate) was manually drawn on a dot plot of side versus forward
light scatter parameters. The lymphocyte scattergate was adjusted to
meet or exceed the guidelines' target percent purity (>85%) and
percent recovery (>90%) using CD45 and CD14 expression (2,
3). The percentages of CD3+ CD4+ and
CD3+ CD8+ lymphocytes within the total
lymphocyte light scattergate were determined by using the two-color
antibody combinations described in Table 1 and adjusted for purity by
dividing the measured subset percentage by the percent purity. Absolute
CD3+ CD4+ and CD3+ CD8+
lymphocyte counts (herein referred to as CD4 T-cell counts and CD8
T-cell counts, respectively) were calculated by multiplying the
specific subset percentage by the absolute lymphocyte count.
(ii) Flow-count method.
The percentages of CD3+
CD4+ and CD3+ CD8+ lymphocytes were
determined by using a manually drawn lymphocyte scattergate and
two-color immunofluorescence as described for the predicate method
above. Immediately prior to analysis, 100 µl of Flow-Count
fluorospheres (Beckman Coulter) were added to each lysed specimen.
Absolute CD4 and CD8 T-cell counts were automatically determined by the System II software using the ratio of CD3+ CD4+
or CD3+ CD8+ lymphocytes to fluorospheres
counted using the following formula: cells per microliter = [(cells counted)/(fluorospheres counted)] × fluorospheres/microliter.
(iii) tetraONE method.
The tetraONE System (Beckman Coulter)
consists of a fully automated software-reagent combination that
performs CD4 and CD8 T-cell counts by four-color analysis on the EPICS
XL flow cytometer. In this integrated software-reagent system,
instrument standardization and spectral compensation are performed
automatically by analysis of four tubes of bead standards or
fluorochrome-stained reference cells prior to specimen analysis. For
each specimen, 100 µl of EDTA-anticoagulated blood was incubated with
10 µl of tetraCHROME reagent containing the four
antibody-fluorochrome combinations described in Table 1. Specimens were
then lysed as described above on a Multi-Q-Prep work station.
Immediately prior to analysis, 100 µl of Flow-Count fluorospheres was
added to each tube. Tubes were placed into an autoloading carousel, and
sample acquisition and analysis were performed automatically by the
tetraONE System software. A lymphocyte gate was determined
automatically using CD45 fluorescence and side scatter parameters. The
percentages of CD3+ CD4+ and CD3+
CD8+ lymphocytes within the lymphocyte gate were reported.
Absolute CD4 and CD8 T-cell counts were automatically calculated using the ratio of cells to fluorospheres as described above for the Flow-Count method.
Data review and exclusion.
All histograms from the
hematology analyzers and flow cytometers were reviewed by operators for
consistent light scatter and fluorescence patterns prior to reporting
data. Data were considered unusable for the following reasons: (i)
inability to generate an absolute lymphocyte count on the hematology
instrument, (ii) absence of detectable fluorescence (e.g., no antibody
added), and (iii) incomplete or abnormal lysis of erythrocytes, as
determined by observing abnormal light scatter patterns or differences
in CD3% of >5 percentage points between tubes. Analyses that failed to yield usable results for these reasons accounted for less than 1%
of the total data collected. Data for central specimens that were
deemed unusable at one site were excluded from all sites. Of 75 central
specimens that arrived on time, 8 were excluded from analysis at all
sites for the reasons described above.
Within-laboratory precision.
To compare the
within-laboratory precision of the three assay methods, each site
attempted to obtain local specimens from 14 HIV-positive donors in
which half of the specimens had a CD4 T-cell count of >200 cells/µl
and half had a CD4 T-cell count of <200 cells/µl. A total of eight
tubes (5 ml each) of blood were drawn sequentially from each donor.
Four tubes were used for "fresh" analysis, where eight replicate
analyses (two analyses for each of four tubes) were performed by all
three methods within 6 h of blood collection. The remaining four
tubes were held at ambient temperature overnight, and the identical
eight replicate analyses were performed by all three methods. Analysis
of "aged" specimens was performed 22 to 33 h after blood collection.
Between-laboratory precision.
Each site received identical
amounts of 103 blood specimens obtained by a central contractor.
Equivalent representation of donors in the following CD4 T-cell count
strata was attempted: <200 cells/µl, 200 to 500 cells/µl, and
>500 cells/µl. Each central specimen was analyzed at each site by
all three methods within 36 h of blood collection.
Accuracy of absolute counts.
We calculated the differences
in absolute counts between each single-platform method and the
predicate method to determine the extent of agreement between the new
and predicate methods using the central specimens that had been assayed
at each site. Differences between new and predicate methods were
plotted for each laboratory. We also attempted to assess whether
differences between new and predicate methods in individual
laboratories were due to assay biases in the predicate or the
single-platform methods. For each specimen, we calculated the
difference between the individual laboratory's absolute CD4 or CD8
T-cell count and the median absolute count from all five laboratories.
Deviations from the median value were plotted for each laboratory by method.
Statistical analyses.
Variability was determined by using
the %CV. Determination of significant differences between the
variability of the Flow-Count method and the predicate method and
between the tetraONE method and the predicate method were based on the
Wilcoxon rank sum test. This test was applied to the %CVs of CD4 and
CD8 T-cell counts and the %CVs of CD4 and CD8 T-cell percentages for
each specimen (between laboratories for central specimens and among
replicates for local specimens). Comparisons of the differences in
absolute counts (and percentages) between the Flow-Count method and the predicate method and between the tetraONE method and the predicate method were based on the Wilcoxon rank sum test applied to the CD4 and
CD8 T-cell counts (and percentages) for each specimen at each
laboratory, stratifying by donor in the case of local specimens. The
same test was used to compare the differences in absolute counts (and
percentages) between specimens analyzed fresh and the same specimens
analyzed after 22 to 33 h (using the same analytical method). A
Bonferroni correction for multiple testing (k = 2, statistical significance defined as P < 0.025) was
used to adjust for comparing both the Flow-Count and tetraONE methods with the same predicate method results.
 |
RESULTS |
Description of data. (i) Local specimens.
All laboratories
analyzed specimens from
14 local donors, and a description of these
specimens is shown in Table 2. Despite a
protocol requesting that each laboratory recruit an equivalent number
of donors with CD4 T-cell counts of >200 cells/µl and <200 cells/µl, two laboratories (A and D) obtained fewer than seven specimens from donors having <200 CD4 T cells/µl.
(ii) Central specimens.
Of 103 central specimens shipped by
the central contractor, 87 were received by all sites within 33 h
of collection. Of the 87 specimens received on time, usable data were
generated by all sites on 67 specimens comparing the predicate and
Flow-Count methods and on 71 specimens comparing the predicate and
tetraONE methods (Table 3). All three CD4
T-cell strata were represented in these specimens. However, the CD4
>500 cells/µl stratum was somewhat underrepresented (<20% of all
specimens).
Within-laboratory precision. (i) Flow-Count method versus predicate
method.
Eight replicates of each local specimen were analyzed by
both the predicate and Flow-Count methods within 6 h of specimen collection. The precision of absolute CD4 and CD8 T-cell counts and
percentages was determined for each method by calculating the %CV from
the replicate measurements of each specimen. As shown in Table
4, the median %CV for absolute CD4 and
CD8 T-cell counts obtained by the Flow-Count method were usually
similar or lower than the %CV for results obtained by the predicate
method but did not differ significantly. As expected, the median %CV
for percentage of CD4 and CD8 T cells obtained by the Flow-Count and predicate methods (Table 5) were nearly
identical, since both of these methods calculated CD4 and CD8 T-cell
percentages using a two-color assay with a manually drawn lymphocyte
light scatter gate. These results suggest that the use of Flow-Count
fluorospheres for absolute cell counting resulted in overall
within-laboratory precision that was equivalent to that obtained using
the predicate method.
View this table:
[in this window]
[in a new window]
|
TABLE 4.
Comparison of within-laboratory precision of absolute CD4
and CD8 T-cell counts for predicate method versus Flow-Count and
tetraONE methodsa
|
|
View this table:
[in this window]
[in a new window]
|
TABLE 5.
Comparison of within-laboratory precision of CD4 and CD8
T-cell percentages for predicate method versus Flow-Count and
tetraONE methodsa
|
|
There was considerable variation between laboratories in the relative
precision achieved by these two methods. In three laboratories, the
median %CV was lower for the Flow-Count method than for the predicate
method. In the other two laboratories, the median %CV was lower for
the predicate method (Fig. 1A). However,
differences in precision between methods were never statistically
significant. Of note, the two laboratories whose CD4 and CD8 T-cell
counts were less precise by the Flow-Count than by the predicate method also had the lowest %CV by the predicate method (Fig. 1A).

View larger version (27K):
[in this window]
[in a new window]
|
FIG. 1.
Comparison of within-laboratory precision in absolute
CD4 and CD8 T-cell counts between predicate method and single-platform
methods by laboratory. Median %CVs are illustrated for each method
from eight replicates of 14 specimens. (A) Flow-Count method versus
predicate method; (B) tetraONE method versus predicate method. *,
significantly different from predicate method, P < 0.025.
|
|
(ii) tetraONE method versus predicate method.
The precision in
CD4 and CD8 T-cell absolute counts and percentages obtained by the
four-color tetraONE method was compared with that of the predicate
method in the manner described above. As shown in Table 4, the %CV for
both absolute CD4 and CD8 T-cell counts was significantly lower for
values obtained by the tetraONE method than for those obtained by the
predicate method. This improved precision for both CD4 and CD8 T-cell
subsets occurred predominantly in specimens with CD4 T cell counts of
<200 cells/µl. Within this stratum, %CVs decreased overall by 1 to
2%. While statistically significant, the improvement in precision was
relatively small. The tetraONE method, which employed an automated CD45
lymphocyte gating strategy, also resulted in significantly lower %CV
for CD4 and CD8 T-cell percentages (predominantly in the <200 CD4 T-cell count stratum) compared with the predicate method (Table 5).
Thus, use of the tetraONE method resulted in a small but significant
improvement in the within-laboratory precision of CD4 and CD8 T-cell
counts and percentages over the predicate method.
Striking differences were observed in the relative precision that
individual laboratories obtained by the tetraONE and predicate methods.
Of the five participating laboratories, three showed a lower median
%CV with the tetraONE method than with the predicate method for
absolute CD4 or CD8 T-cell counts (Fig. 1B). However, in two
laboratories, the tetraONE method resulted in a higher median %CV than
the predicate method. The two laboratories that obtained poorer
performance with the tetraONE method were the same laboratories in
which poorer performance occurred with the Flow-Count method.
(iii) Fresh versus aged specimens.
The effect of overnight
aging of specimens on assay precision and on absolute CD4 and CD8
T-cell counts was also assessed with the local specimens. To measure
the effect on precision, we determined the %CV of the absolute CD4 and
CD8 T-cell counts for specimens analyzed within 6 h of drawing and
compared that with the %CV observed for the same specimens analyzed
between 24 and 33 h after drawing (all CD4 strata combined). There
were no significant differences in assay precision (i.e., %CV) when fresh and aged specimens were analyzed by either the predicate or the
Flow-Count method. Aging also had no effect on the precision of CD8
T-cell counts obtained by the tetraONE method. The %CV of absolute CD4
T-cell counts obtained using the tetraONE method was significantly
greater for aged specimens than for fresh specimens. However, the
median increase in %CV was small (<1%), and despite this increase,
the median %CV for CD4 T-cell counts on aged specimens analyzed by the
tetraONE method was still smaller than the median %CV for fresh
specimens analyzed by the predicate or Flow-Count method (data not shown).
The overnight aging of specimens resulted in very small but
statistically significant changes in absolute CD4 and CD8 T-cell counts, as measured by both the Flow-Count and tetraONE methods (median
change in CD4 and CD8 T-cell counts for the Flow-Count method was
7
and
18 cells/µl, respectively, and for the tetraONE method it was
4 and +9 cells/µl, respectively). Aging had no effect on absolute
CD4 T-cell counts determined by the predicate method. However, by the
predicate method, aged specimens yielded significantly higher CD8
T-cell counts (median change, +18 cells/µl) (data not shown).
Between-laboratory precision. (i) Flow-Count method versus
predicate method.
The between-laboratory precision in absolute CD4
and CD8 T-cell counts was compared between the Flow-Count and predicate
methods using central specimens that were analyzed at all five
laboratories. The %CV was calculated from the values obtained in the
five laboratories for each specimen by the different assay methods.
Analysis of absolute CD4 and CD8 T-cell counts by the Flow-Count method
resulted in significantly improved between-laboratory precision (Table 6). Compared with the predicate method,
the Flow-Count method reduced between-laboratory %CV for absolute
CD4 and CD8 T-cell counts by approximately 7%. There was a tendency
toward greater decreases in %CV among specimens with lower CD4 counts.
Significant reductions in between-laboratory %CV were observed for CD4
and CD8 T-cell counts on specimens in the <200 and 200 to 500 strata (Table 6). The %CV of the CD4 and CD8 T-cell percentages did not vary
significantly between these two assay methods (data not shown).
View this table:
[in this window]
[in a new window]
|
TABLE 6.
Comparison of between-laboratory precision in absolute
T-cell counts for predicate method versus
Flow-Count methoda
|
|
(ii) tetraONE method versus predicate method.
Similar analyses
were performed to compare the between-laboratory precision in absolute
CD4 and CD8 T-cell counts by the tetraONE method and the predicate
method. Compared with the predicate method, the tetraONE method also
resulted in a significant reduction of between-laboratory %CV (Table
7). The magnitude of this reduction was
similar to that observed in the Flow-Count method. The reduction in
%CV occurred across all CD4 strata (Table 7). The %CVs of CD4 T-cell
percentages did not differ significantly between the two methods.
However, the %CVs of CD8 T-cell percentages were slightly but
significantly higher for tetraONE results than for predicate results
(median difference in %CV, +1.9%) (data not shown).
View this table:
[in this window]
[in a new window]
|
TABLE 7.
Comparison of between-laboratory precision in absolute
T-cell counts for predicate method versus
tetraONE methoda
|
|
Accuracy of single-platform methods.
The accuracy of the
single-platform methods was determined by comparing results from the
new methods with those obtained using the predicate method. Using
results from the central specimens, subset percentages or absolute
counts from the predicate method were subtracted from results derived
by either the Flow-Count or tetraONE method to determine the direction
and magnitude of the differences. Overall, there was good agreement
between CD4 and CD8 T-cell percentages derived by the predicate method
and those obtained using either the Flow-Count or tetraONE method. Median CD4 and CD8 T-cell percentages varied by less than 2 percentage points between predicate and Flow-Count or tetraONE except for laboratory D, in which the tetraONE method yielded a median CD8 T-cell
percentage that was 4 percentage points lower than the median CD8
T-cell percentage from the predicate method (data not shown).
The differences in absolute counts between the predicate and
single-platform methods by laboratory are shown in Fig.
2. Four laboratories (A, B, D, and E)
produced absolute counts using the Flow-Count method that were slightly
but consistently lower that the absolute counts derived by their
predicate method (median absolute difference: CD4,
18 to
47
cells/µl [Fig. 2A]; CD8,
48 to
112 cells/µl [Fig. 2B]). The
remaining laboratory (C) produced results using the Flow-Count method
that were consistently higher than those derived by the predicate
method (median absolute difference: CD4, +7; CD8, +42) (Fig. 2A and B).
Interestingly, the identical pattern of differences was observed in
each laboratory when absolute counts derived by their predicate method
and tetraONE method were compared (Fig. 2C and D). The same
laboratories (A, B, D, and E) produced results using the tetraONE
method that were lower than their predicate method results (median
absolute difference: CD4,
14 to
36 cells/µl; CD8,
30 to
196
cells/µl). The same remaining laboratory (C) produced higher absolute
counts by the tetraONE method than with the predicate method (median
absolute difference: CD4, +19; CD8, +22).

View larger version (25K):
[in this window]
[in a new window]
|
FIG. 2.
Absolute differences in CD4 and CD8 T-cell counts
between single-platform and predicate methods. (A and B) Flow-Count
method minus predicate for CD4 (A) and CD8 (B); (C and D) tetraONE
method minus predicate for CD4 (C) and CD8 (D). Box indicates median
and interquartile range (25th and 75th percentiles); upper and lower
lines indicate 90th and 10th percentiles, respectively.
|
|
Source of difference in absolute counts between predicate and
single-platform methods.
The magnitude and direction of the
differences we observed in absolute counts between either the
Flow-Count or tetraONE method and the predicate method were consistent
within each laboratory. However, it was unclear whether discrepancies
in absolute counts between the predicate and single-platform methods
occurred because one of the methods consistently yielded higher or
lower results in an individual laboratory. In an attempt to determine
the source of this bias, we compared the absolute count generated
by each laboratory with the median absolute counts from all five
laboratories for each method. This analysis assumed that the
five-laboratory median absolute count would be closer to the actual
absolute count than any individual measurement.
Figure 3 illustrates the differences
observed when the five-laboratory median value was subtracted from the
predicate value (x axis), and the tetraONE value
(y axis) for absolute CD4 T-cell counts measured in each
laboratory. Laboratory B, which showed good agreement between absolute
counts derived by the tetraONE and predicate methods (Fig. 2C), also
showed good agreement between CD4 T-cell counts and the five-laboratory
median CD4 T-cell count obtained by both assay methods (Fig. 3B). The
median difference between this laboratory's values and the
five-laboratory median values for the predicate and Flow-Count methods
was 0 and 0, respectively. Laboratory C, which produced higher absolute
CD4 T-cell counts by tetraONE than by their predicate method (Fig. 2C),
showed a bias toward underestimation of CD4 T-cell count by their
predicate method (median difference,
41 cells) compared with the
five-laboratory median. However, this laboratory's results by tetraONE
were in good agreement with the five-laboratory tetraONE median values (median difference, 0 cells) (Fig. 3C). This suggested that the bias in
absolute counts from laboratory C was largely due to underestimation by
the predicate method. In contrast, laboratory E produced lower CD4
T-cell counts by the tetraONE method than by their predicate method
(Fig. 2C). A comparison of this laboratory's values with the
five-laboratory median values suggested that the difference in this
laboratory was due to an overestimation of CD4 T-cell count by the
predicate method. Laboratory E's results from the tetraONE method
agreed well with the five-laboratory median values (predicate median
difference, +5 cells; tetraONE difference, 0 cells) (Fig. 3E). The
results from laboratories A and D were less clear cut. It appeared that
bias in both the tetraONE and predicate methods might have contributed
to the overall difference in CD4 T-cell counts generated by these two
methods. Although only data from CD4 T-cell counts by the tetraONE
method are illustrated, the same trend was observed for absolute CD8
T-cell counts generated by the tetraONE method and for all absolute
counts generated by the Flow-Count method.

View larger version (33K):
[in this window]
[in a new window]
|
FIG. 3.
Source of absolute CD4 T-cell count bias in individual
laboratories. The absolute CD4 T-cell count generated in each
laboratory by the predicate or tetraONE method was subtracted from the
five-laboratory median value by that method to determine whether method
biases occurred in individual laboratories. A through E are laboratory
designations. The shaded region indicates the interquartile range (25th
and 75th percentiles) of these differences.
|
|
 |
DISCUSSION |
The single-platform flow cytometric methodologies for CD4 and CD8
T-cell determinations represent simple and efficient modifications in
clinical lymphocyte subset enumeration. The results of this study
confirm that use of Flow-Count fluorospheres for absolute counting in a
two-color, light scatter-gated assay can provide within-laboratory
precision that is equivalent to a predicate, two-color, scatter-gated
assay in which absolute lymphocyte counts are derived by conventional
hematology. The tetraONE System has automated CD45 lymphocyte gating
and analysis and uses fluorosphere-derived absolute counting. This
method provided a small but significant improvement in
within-laboratory precision for both percentages and absolute counts
compared with the two-color predicate method. The evaluation of these
two single-platform assays suggests that two features of the tetraONE
System, automated CD45 lymphocyte gating and automated analysis,
contribute more to improving within-laboratory precision than the use
of fluorospheres for absolute counting.
More strikingly, use of Flow-Count fluorospheres in either a two-color,
light scatter-gated assay or the four-color, automated tetraONE System
resulted in significant and substantial improvement in
between-laboratory precision. Both single-platform assay methods decreased between-laboratory %CV by nearly half compared with the
predicate method. The within-laboratory and between-laboratory precision that we observed using a conventional multiple-platform method was nearly identical to the precision observed in other multisite studies that analyzed similar specimens (1, 11, 14).
Despite an overall improvement in assay precision, the performance of
individual laboratories varied considerably. Only three of the five
participating laboratories, when examined individually, showed a
significant improvement in within-laboratory precision using the
tetraONE method. The same trend among laboratories, although not
statistically significant, was observed for the Flow-Count method. The
two laboratories that failed to improve their precision were the same
laboratories that enrolled fewer local donors in the <200 CD4 cells
stratum. The single-platform methods resulted in greater improvement in
precision in specimens with lower CD4 T-cell counts. Therefore, this
different distribution of CD4 T-cell counts in the specimens analyzed
by these two laboratories may have contributed to their apparent
difference in performance using the single-platform methods. The
single-platform methodologies also required greater pipetting precision
than that needed to perform the predicate methodologies because
accurate results rely on a precise 1:1 volume ratio of blood and
fluorosphere suspension. Furthermore, the single-platform assay methods
represented new procedures at all participating sites, while these
laboratories were all very experienced in the predicate assay. These
facts may have also served to favor the predicate method at some sites.
We also analyzed the differences in absolute counts (and percentages)
obtained by the new and predicate assays as an estimate of accuracy of
the single-platform methods and to understand how the single-platform
assay methods would impact the absolute CD4 and CD8 T-cell counts (and
percentages) currently reported in individual laboratories. Overall,
there was good agreement between the new and predicate methods in CD4
and CD8 T-cell percentage values. However, individual laboratories
observed small but consistent differences in absolute counts between
the predicate and single-platform assays. These differences could have
resulted from assay bias in either the predicate or single-platform
methods. To address this question, we used the results from central
specimens analyzed by all laboratories. We concluded that the absolute
count bias in at least two laboratories occurred in the predicate
method. The bias in these laboratories was probably introduced by the lymphocyte counts from the automated hematology analyzer because the
CD4 and CD8 T-cell percentages did not vary by assay method. However,
the absolute counts derived by the predicate method showed more
deviation from the median than the absolute counts derived by the
single-platform methods. Consistent with our observation, others have
identified the lymphocyte count derived from the hematology instrument
as the predominant source of CD4 count bias (13; R. Gelman and C. Wilkening, submitted for publication). In two other
laboratories, both the predicate and single-platform method results
differed from the median results of all laboratories. Of note, these
were the same laboratories that were unable to show improved
within-laboratory precision using the new methods.
The comparison of absolute counts (and percentages) between predicate
and new methods was performed using shipped, central specimens.
Consensus guidelines recommend that the automated hematology for
obtaining the absolute lymphocyte count be performed within 6 h of
drawing (2, 3). With some automated hematology analyzers, substantial changes in cell counts can occur after EDTA-anticoagulated blood is aged for periods longer than 12 to 24 h (12,
16). Therefore, the use of 1-day-old central specimens might have
created a slight disadvantage for the predicate method.
This multisite evaluation of Flow-Count fluorospheres and the tetraONE
System in a clinical setting provides assurance that these assay
methods are precise and reliable alternatives to multiple-platform CD4
and CD8 T-cell determinations. These methods provided not only modest
improvement in within-laboratory precision but substantial improvement
in between-laboratory agreement. Therefore, these single-platform
methods have considerable potential to improve data consistency within
multicenter trials. However, the impact of implementing a
single-platform method on assay precision and accuracy within
individual laboratories will depend on biases inherent in current
methods and the skill of testing personnel.
 |
ACKNOWLEDGMENTS |
This work was supported by the NIAID Immunophenotyping Quality
Assessment Program, contract NO-AI-45175, and by funds from Beckman Coulter.
 |
ADDENDUM |
After this study was completed, Beckman Coulter modified the
algorithm used by the tetraONE System to automatically gate
lymphocytes. In the current version (version 2.0 and subsequent
releases), the tetraONE System utilizes forward light scatter, CD45
fluorescence, and side light scatter parameters for lymphocyte gating.
In the present study, only CD45 fluorescence and side light scatter
parameters were used for this gating.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Beth Israel
Deaconess Medical Center, RE-113, P.O. Box 15732, Boston, MA 02215. Phone: (617) 667-4583. Fax: (617) 667-8210. E-mail:
kreimann{at}caregroup.harvard.edu.
M. O'Gorman, Chairman, Northwestern University, Chicago, Ill.;
E. Bessent, University of California, San Diego; S. Plaeger, University
of California, Los Angeles; A. Donnenberg and S. Douglas, Children's
Hospital of Philadelphia, Philadelphia, Pa.; F. Mandy, Bureau for
HIV/AIDS and STD, Health Canada, Ottawa, Canada; J. Nicholson, Centers
for Disease Control and Prevention, Atlanta, Ga.; K. Reimann, Beth
Israel Deaconess Medical Center, Boston, Mass.; J. Schmitz,
University of North Carolina; Chapel Hill; C. Schnizlein-Bick; Indiana
University, Indianapolis; J. Kagan and D. Livnat, DAIDS/NIAID/National
Institutes of Health, Bethesda, Md.
 |
REFERENCES |
| 1.
|
Aboulker, J. P.,
B. Autran,
K. Beldjord,
F. Touraine, and P. Debre.
1992.
Consistency of routine measurements of CD4+, CD8+ peripheral blood lymphocytes.
J. Immunol. Methods
154:155-161[CrossRef][Medline].
|
| 2.
|
Calvelli, T.,
T. N. Denny,
H. Paxton,
R. Gelman, and J. Kagan.
1993.
Guideline for flow cytometric immunophenotyping: a report from the National Institute of Allergy and Infectious Diseases, Division of AIDS.
Cytometry
14:702-715[CrossRef][Medline].
|
| 3.
|
Centers for Disease Control and Prevention.
1997.
1997 revised guidelines for performing CD4+ T-cell determinations in persons infected with human immunodeficiency virus (HIV).
Morbid. Mortal. Weekly Rep.
46(RR-2):1-29[Medline].
|
| 4.
|
Connelly, M. C.,
M. Knight,
J. V. Giorgi,
J. Kagan,
A. L. Landay,
J. W. Parker,
E. Page,
C. Spino,
C. Wilkening, and T. J. Mercolino.
1995.
Standardization of absolute CD4+ lymphocyte counts across laboratories: an evaluation of the Ortho CytoronAbsolute Flow Cytometry System on normal donors.
Cytometry
22:200-210[CrossRef][Medline].
|
| 5.
|
Giorgi, J. V.,
H.-L. Cheng,
J. B. Margolick,
K. D. Bauer,
J. Ferbas,
M. Waxdal,
I. Schmid,
L. E. Hultin,
A. L. Jackson,
L. Park,
J. M. G. Taylor, and the Multicenter AIDS Cohort Study Group.
1990.
Quality control in the flow cytometric measurement of T-lymphocyte subsets: the Multicenter AIDS Cohort Study experience.
Clin. Immunol. Immunopathol.
55:173-186[CrossRef][Medline].
|
| 6.
|
Giorgi, J. V.
1993.
Characterization of T lymphocyte subset alterations by flow cytometry in HIV disease.
Ann. N. Y. Acad. Sci.
677:126-137[Medline].
|
| 7.
|
Ho, H. N.,
L. E. Hultin,
R. T. Mitsuyasu,
J. L. Matud,
M. A. Hausner,
D. Bockstoce,
C. C. Chou,
S. O'Rourke,
J. M. Taylor, and J. V. Giorgi.
1993.
Circulating HIV-specific CD8+ cytotoxic T cells express CD38 and HLA-DR antigens.
J. Immunol.
150:3070-3079[Abstract].
|
| 8.
|
Johnson, D.,
D. Hirschkorn,
M. P. Busch, and the National Heart, Lung, and Blood Institute Retrovirus Epidemiology Donor Study.
1995.
Evaluation of four alternative methodologies for determination of absolute CD4+ lymphocyte counts.
J. Acquir. Immune Defic. Syndr. Hum. Retrovirol.
10:522-530[Medline].
|
| 9.
|
Mildvan, D.,
A. Landay,
V. De Gruttola,
S. G. Machado, and J. Kagan.
1997.
An approach to the validation of markers for use in AIDS clinical trials.
Clin. Infect. Dis.
24:764-774[Medline].
|
| 10.
|
Nicholson, J. K. A.,
W. M. Velleca,
S. Jubert,
T. A. Green, and L. Bryan.
1994.
Evaluation of alternative CD4 technologies for the enumeration of CD4 lymphocytes.
J. Immunol. Methods
177:43-54[CrossRef][Medline].
|
| 11.
|
O'Gorman, M. R. G.,
R. Gelman Site Investigators, and the NIAID New CD4 Technologies Focus Group.
1997.
Inter- and intrainstitutional evaluation of automated volumetric capillary cytometry for the quantitation of CD4- and CD8-positive T lymphocytes in the peripheral blood of persons infected with human immunodeficiency virus.
Clin. Diagn. Lab. Immunol.
4:173-179[Abstract].
|
| 12.
|
Reardon, D. M.,
B. Warner, and E. A. Trowbridge.
1991.
EDTA, the traditional anticoagulant of haematology: with increased automation is it time for a review?
Med. Lab. Sci.
48:72-75[Medline].
|
| 13.
|
Robinson, G.,
L. Morgan,
M. Evans,
S. McDermott,
S. Pereira,
M. Wansbrough-Jones, and G. Griffin.
1992.
Effect of type of haematology analyser on CD4 count.
Lancet
340:485[CrossRef][Medline].
|
| 14.
|
Schnizlein-Bick, C.,
J. Spritzler,
C. Wilkening,
J. K. A. Nicholson,
M. R. G. O'Gorman,
Site Investigators, and the NIAID New Technologies Evaluation Group.
2000.
Evaluation of TruCount absolute-count tubes for determining CD4 and CD8 cell numbers in human immunodeficiency virus-positive adults.
Clin. Diagn. Lab. Immunol.
7:336-343[Abstract/Free Full Text].
|
| 15.
|
Smith, R. D.
1990.
The pathobiology of HIV infection.
Arch. Pathol. Lab. Med.
114:235-239[Medline].
|
| 16.
|
Spence, N.
1993.
Differential leucocyte analysis of samples up to 48 hours old: improved results with citrate pyridoxal phosphate anticoagulant.
Br. J. Biomed. Sci.
50:313-320[Medline].
|
| 17.
|
Stein, D. S.,
J. A. Korvick, and S. H. Vermund.
1992.
CD4+ lymphocyte cell enumeration for prediction of clinical course of human immunodeficiency virus disease: a review.
J. Infect. Dis.
165:352-363[Medline].
|
| 18.
|
Stewart, C. C., and J. A. Steinkamp.
1982.
Quantitation of cell concentration using the flow cytometer.
Cytometry
2:238-243[Medline].
|
| 19.
|
Walker, B. D.,
S. Chakrabarti,
B. Moss,
T. J. Paradis,
T. Flynn,
A. G. Durno,
R. S. Blumburg,
J. C. Kaplan,
M. S. Hirsch, and R. T. Schooley.
1987.
HIV-specific cytotoxic T lymphocytes in seropositive individuals.
Nature (London)
328:345-348[CrossRef][Medline].
|
Clinical and Diagnostic Laboratory Immunology, May 2000, p. 344-351, Vol. 7, No. 3
1071-412X/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Imade, G. E., Badung, B., Pam, S., Agbaji, O., Egah, D., Sagay, A. S., Sankale, J.-L., Kapiga, S., Idoko, J., Kanki, P.
(2005). Comparison of a New, Affordable Flow Cytometric Method and the Manual Magnetic Bead Technique for CD4 T-Lymphocyte Counting in a Northern Nigerian Setting. CVI
12: 224-227
[Abstract]
[Full Text]
-
Kuss, I., Hathaway, B., Ferris, R. L., Gooding, W., Whiteside, T. L.
(2004). Decreased Absolute Counts of T Lymphocyte Subsets and Their Relation to Disease in Squamous Cell Carcinoma of the Head and Neck. Clin. Cancer Res.
10: 3755-3762
[Abstract]
[Full Text]
-
Janossy, G., Jani, I. V., Bradley, N. J., Bikoue, A., Pitfield, T., Glencross, D. K.
(2002). Affordable CD4+-T-Cell Counting by Flow Cytometry: CD45 Gating for Volumetric Analysis. CVI
9: 1085-1094
[Abstract]
[Full Text]
-
Al Qouzi, A., Al Salamah, A., Al Rasheed, R., Al Musalam, A., Al Khairy, K., Kheir, O., Al Ajaji, S., Hajeer, A. H.
(2002). Immunophenotyping of Peripheral Blood Lymphocytes in Saudi Men. CVI
9: 279-281
[Abstract]
[Full Text]
-
O'Gorman, M. R. G., Nicholson, J. K. A.
(2000). Adoption of Single-Platform Technologies for Enumeration of Absolute T-Lymphocyte Subsets in Peripheral Blood. CVI
7: 333-335
[Full Text]