Clinical and Diagnostic Laboratory Immunology, January 2000, p. 122-124, Vol. 7, No. 1
1071-412X/0/$04.00+0
Evaluation of the Quantiplex Human Immunodeficiency
Virus Type 1 RNA 3.0 Assay in a Tertiary-Care Center
Howard
Gale*
Infectious Diseases Section, Medical Service,
Veterans Affairs Medical Center, Washington, D.C.
Received 19 July 1999/Returned for modification 25 August
1999/Accepted 14 September 1999
 |
ABSTRACT |
The Quantiplex human immunodeficiency virus type 1 RNA 3.0 Assay
(bDNA) (Bayer Diagnostics, Walpole, Mass.) produced linear and
reproducible (intra-assay and interassay) results over its quantification range of 50 to 500,000 copies/ml of plasma with 96%
specificity. A threefold or 0.5-log10 change or greater was clinically significant for serial patient samples.
 |
TEXT |
Serial human immunodeficiency virus
type 1 (HIV-1) RNA testing is an essential component of management for
individual HIV-positive patients (1, 2). The Quantiplex
HIV-1 RNA 3.0 Assay (bDNA) (Bayer Diagnostics, Walpole, Mass.) is a
signal amplification nucleic acid probe assay for the direct
quantification of plasma HIV-1 RNA (3). In validating this
assay for use in our laboratory, we examined its (i) specificity, (ii)
linearity over its reportable range of 50 to 500,000 copies/ml of
plasma, (iii) intra-assay (well-to-well) variability, (iv) interassay
(run-to-run) variability, (v) variability between serial patient
samples and its clinical significance, and (vi) copy numbers, compared
to those of the (previous) 2.0 version. We also compared a plastic tube
to the standard glass tube used to collect the blood sample. Many
laboratories have switched to the plastic tube to avoid breakage.
(This work was presented in part at the Fifteenth Annual Clinical
Virology Symposium, Clearwater Beach, Fla., 9 to 12 May 1999, abstr. M19).
Bayer Diagnostics (formerly Chiron) provided frozen plasma for the
specificity and linearity testing. The other samples were obtained from
HIV-positive patients at the Veterans Affairs Medical Centers in
Washington, D.C., and Baltimore, Md. Peripheral blood was collected by
venipuncture, kept at room temperature, and centrifuged within 4 h
at 1,000 × g for 10 to 15 min. The plasma was frozen at
80°C. Each specimen was then tested according to the
manufacturer's directions (Quantiplex HIV-1 RNA 3.0 Assay [bDNA]
package insert; Bayer Diagnostics). A run of a complete plate contained
up to 80 samples; a half-plate run contained 35 samples. One person prepared and analyzed all of the plasma samples.
Seventy-four HIV antibody-negative plasmas were analyzed by using three
runs and three kits from two kit lots. For 71 samples, the HIV-1 RNA
level was below the detectable limit of 50 copies/ml. The three
remaining samples gave values of 85, 114, and 237 copies/ml of plasma.
These values were well below the range reported for acute HIV infection
of 27,200 to 1,600,000 copies/ml (4). These low quantities
were considered false positives, and the specificity of this assay was
96%. This is consistent with the manufacturer's statement of intended
use; this test is not to be employed for the diagnosis of HIV infection
(Quantiplex package insert). However, clinicians do order HIV-1 RNA
assays to screen for acute infection before seroconversion has
occurred. If this particular assay is being used, caution must be taken
to not interpret a low positive value as proof of a genuine HIV infection.
Seven samples from a serial dilution panel were analyzed eight times
each by using two runs and two kits from the same kit lot. As
illustrated in Fig. 1, the Quantiplex
HIV-1 RNA 3.0 Assay (bDNA) gave linear results over its quantification
range of 50 to 500,000 copies/ml. The least-squares linear regression
for the median log10 copies per milliliter was as follows:
y = 0.075 + 0.983x, with r2 = 0.996.

View larger version (25K):
[in this window]
[in a new window]
|
FIG. 1.
Linearity check of Quantiplex HIV-1 RNA 3.0 Assay
(bDNA). The expected values are the medians of the 88 quantifications
per sample performed by 11 laboratories and are expressed as
log10 copies per milliliter. The observed values are the
eight runs per sample performed by our laboratory and are expressed as
log10 copies per milliliter.
|
|
In the intra-assay study, matching plasma samples in adjacent wells
were assayed by using four runs and three kits of the same kit lot.
Table 1 shows little variation between
wells. There was also agreement on four samples being <50 copies/ml
and one sample being >500,000 copies/ml. In designing this study, it
was assumed that using two varieties of Vacutainer EDTA tubes would not
contribute significantly to the variability. The data supports this
assumption, and K3 EDTA liquid in glass tubes and
K2 EDTA sprayed in plastic tubes can be used
interchangeably with this assay.
View this table:
[in this window]
[in a new window]
|
TABLE 1.
Summary of intra-assay, interassay, and serial clinical
sample correlations for the Quantiplex HIV-1 RNA 3.0 assay (bDNA)a
|
|
In the interassay study, 35 patients were phlebotomized once, and their
plasmas were frozen, thawed, and assayed twice. Three or four samples
were compared at a time by using 11 runs and five kits of the same kit
lot. As shown in Table 1, the run-to-run variability was low. However,
its effect was still apparent: four of five runs of four samples and
two of five runs of three samples had values all higher or all lower
than those of the previous run. The median decrease of 15% from the
first to second run is likely the result of the extra freeze-thaw
cycle. Interassay variability was also examined with an in-house
control (Fig. 2). Each run had one
specific well dedicated to hold this control: mean, 17,166 copies/ml;
standard deviation (SD), 3,179; percent coefficient of variation, 18.5;
n, 48. A significant difference was found between kit lot
no. 3 (mean, 13,057 copies/ml; SD, 1,673; n, 7) and the
other six kit lots (mean, 17,868 copies/ml; SD, 2,828; n, 41 [P < 0.001, two-tailed t test]).

View larger version (18K):
[in this window]
[in a new window]
|
FIG. 2.
Interassay variability of Quantiplex HIV-1 RNA 3.0 Assay
(bDNA). The in-house control was pooled plasma frozen in 1-ml aliquots
at 80°C; it was analyzed by using 48 runs, seven kit lots, and one
operator from September 1998 to July 1999. , kit lot no. 3; , all
other kit lots.
|
|
During a 3-month period, 64 pairs of serial samples had values within
the quantification range (Fig. 3). A
retrospective chart review determined that 25 patients had decreases
(0.381 to 3.103 log10; median, 1.519 log10) in
their HIV-1 RNA following changes in their antiretroviral therapy.
Eight patients had increases (0.645 to 2.240 log10; median,
1.214 log10) because they were nonadherent. The remaining
31 patients (Fig. 3; Table 1) had stable clinical and therapeutic
histories. Their samples, drawn 6 to 84 days apart (median of 42 days),
had less than a threefold or 0.5-log10 95th percentile
change. There was a 24% median decrease from the first to second
sample. Despite the lower control values obtained with kit lot no. 3 (Fig. 2), its use in analyzing 18 of the second samples was not a
significant factor: the median decrease for those pairs was also 24%.

View larger version (24K):
[in this window]
[in a new window]
|
FIG. 3.
Serial clinical sample variability with Quantiplex HIV-1
RNA 3.0 Assay (bDNA). , antiretroviral therapy (ART) started or
changed; , ART interrupted; , no clinical or ART changes.
|
|
Eighty-six patient samples with version 2.0 viral loads of 600 to
416,000 copies/ml (median, 8,000 copies/ml) were reanalyzed with
version 3.0 by using two runs and two kits from the same kit lot. The
linear regression correlation (r2) was 0.965 for
the log10 copies per milliliter. The 3.0 version values
were consistently higher, with a mean assay ratio (version 3.0 to
version 2.0) of 2.5 ± 0.9 (1 SD). For 78 specimens with version
2.0 values of <500 copies/ml, 41% had values of <50 copies/ml, 38%
had values of 50 to 500 copies/ml, and 21% had values of 500 to 2,600 copies/ml with the 3.0 version. These findings are consistent with data
from other laboratories (C. Major, T. Degazio, R. Galli, A. Rachlis, C. Kovacs, S. Read, and M. Fearon, 12th World AIDS Congr., poster
presentation 1998; J. L. Perez, P. Perez, J. M. Escriba, D. Podzamczer, and R. Martin, 6th Conf. Retrovir. Opportun. Infect.,
abstr. 146, 1999; L. Sawyer, S. Barr, G. Gorrin, D. Hendricks, J. Kolberg, J. Robertson, L. Shen, and B. Irvine, 14th Annu. Clin. Virol.
Symp., abstr. T34, 1998.
In conclusion, the Quantiplex HIV-1 RNA 3.0 Assay (bDNA) was 96%
specific and gave linear and reproducible (intra-assay, interassay, and
serial clinical samples) results over its quantification range of 50 to
500,000 copies/ml. It correlated well with the 2.0 assay, generally
giving 1.5- to 4.5-fold-higher values. Either glass or plastic EDTA
Vacutainer tubes can be used with this assay.
 |
ACKNOWLEDGMENTS |
I thank V. Kan for reviewing the manuscript and making helpful suggestions.
 |
FOOTNOTES |
*
Mailing address: I. D. Section (151B), Veterans
Affairs Medical Center, 50 Irving St. NW, Washington, DC 20422. Phone:
(202) 745-8301. Fax: (202) 745-8432. E-mail:
gale.howard{at}washington.va.gov.
 |
REFERENCES |
| 1.
|
Carpenter, C. C.,
A. Fischl,
S. M. Hammer,
D. M. Jacobson,
D. A. Katzenstein,
J. S. G. Montaner,
D. D. Richman,
M. S. Sang,
R. T. Schooley,
M. A. Thompson,
S. Vella,
P. G. Yeni, and P. A. Volberding.
1998.
Antiretroviral therapy for HIV infection in 1998: updated recommendations of the International AIDS Society-USA panel.
JAMA
280:78-86[Abstract/Free Full Text].
|
| 2.
|
Centers for Disease Control and Prevention.
1998.
Report of the NIH panel to define the principles of therapy of HIV infection and guidelines for the use of antiretroviral agents in HIV-infected adults and adolescents.
Morbid. Mortal. Weekly Rep.
47:1-82[Medline].
|
| 3.
|
Collins, M. L.,
B. Irvine,
D. Tyner,
E. Fine,
C. Zayati,
C. Chang,
T. Horn,
D. Ahle,
J. Detmer,
L. Shen,
J. Kolberg,
S. Bushnell,
M. S. Urdea, and D. D. Ho.
1997.
A branched DNA signal amplification assay for quantification of nucleic acid targets below 100 molecules/mL.
Nucleic Acids Res.
25:2979-2984[Abstract/Free Full Text].
|
| 4.
|
Schacker, T. W.,
J. P. Hughes,
T. Shea,
R. W. Coombs, and L. Corey.
1998.
Biological and virological characteristics of primary HIV infection.
Ann. Intern. Med.
128:613-620[Abstract/Free Full Text].
|
Clinical and Diagnostic Laboratory Immunology, January 2000, p. 122-124, Vol. 7, No. 1
1071-412X/0/$04.00+0