Canadian Food Inspection Agency, Animal
Diseases Research Institute, Nepean, Ontario, Canada K2H
8P91;
International Atomic Energy
Agency, A1400, Vienna, Austria2;
Instituto Colombiano Agropecuaria, ICA-CEISA, Santafe de
Bogota DC, Colombia3;
Tropical Diseases
Research Program, School of Veterinary Medicine, National
University, Heredia, Costa Rica4;
Servicios Agricolas y Ganaderos, Laboratorio Regional
Osorno, Mackenna 674, Osorno, Chile5; and
Instituto de Patobiologia-DPTO Bacteriologia, INTA-CICV, CC
77, 1708 Moron, Buenos Aires, Argentina6
Received 23 February 1998/Returned for modification 14 April
1998/Accepted 27 May 1998
The results of a field trial conducted in Latin America with two
indirect enzyme-linked immunosorbent assays (ELISAs) and two
competitive ELISAs (CELISAs) for the detection of bovine antibody to
Brucella abortus are reported. One of the CELISA formats
performed most accurately. The percentage of positive reactions in the
CELISA relative to the selected positive rose bengal agglutination test (RBT) and complement fixation test (CFT) results was 97.47%, the percentage of negatives relative to the selected negative RBT and CFT
results for unexposed cattle was 98.32%, and the percentage of
negatives in cattle vaccinated with B. abortus 19 was
96.51%. The same assay format under Canadian conditions had an actual sensitivity of 100%, a specificity of 99.90% in nonvaccinates, and a
specificity of 97.7% in a strain 19-vaccinated population. Overall,
the CELISA performed as expected and the results were not dissimilar
from the results obtained in the Canadian study. This provided further
evidence that this CELISA can in many instances differentiate infected
cattle from those that are vaccinated or infected with a cross-reacting
organism while still giving very few false-positive or false-negative
results.
 |
INTRODUCTION |
The indirect enzyme-linked
immunosorbent assay (IELISA) for detection of antibody to
Brucella abortus was introduced in 1976 (1). The
reasons for using IELISAs were, firstly, to replace conventional
serological tests (3) that in many ways did not perform well
and frequently required a panel of tests for diagnosis and, secondly,
to introduce an assay which could be standardized, quality controlled,
and automated. A large number of IELISAs have been described in the
literature (13), but in spite of the numerous modifications,
the specificities of these assays were less than expected. The reason
for this is partly because antibody resulting from B. abortus 19 vaccination or from exposure to cross-reacting antigens
is detected by this procedure.
To increase specificity, competitive ELISAs (CELISAs) were developed
(4-6, 8). By selection of a suitable monoclonal antibody to
compete with antibody present in test serum, reactivity resulting from
the vaccine or cross-reacting antigens could be virtually eliminated.
Two of these assays were developed and validated largely in
circumstances where brucellosis had been eradicated (Canada) with sera
from animals in which B. abortus infection was confirmed by
culture as reference sera. It was therefore necessary to field test
these assays in areas with brucellosis and vaccination programs. For
these purposes, four laboratories in Latin America were selected. These
laboratories were selected based on the incidence of brucellosis in
each area. Chile had a relatively low incidence, while higher incidences were found in Costa Rica, Colombia, and Argentina.
This communication describes the results obtained with two IELISAs and
two CELISAs compared to those from the diagnostic serological tests in
use in each laboratory.
 |
MATERIALS AND METHODS |
Test samples.
Samples were defined on the basis of their
serological reactions on both the rose bengal agglutination test (RBT)
and the complement fixation test (CFT) by the official criteria for
positive results as determined by each country for the CFT.
Serologically negative samples were defined as those primarily from
regions that had no history or serological evidence of B. abortus infection and were negative on both the RBT and the CFT.
Some animals in the negative population were vaccinated with B. abortus 19.
Serologically positive samples were defined as those samples from
infected herds which were positive on both the RBT and the CFT. This
positive population was thought to include cattle with residual
vaccinal antibody or antibody resulting from exposure to cross-reacting
antigens.
Control sera.
Control sera were supplied by the Animal
Diseases Research Institute (ADRI) for one IELISA and both CELISAs from
ADRI. These consisted of a strong positive control serum from a cow
from which B. abortus had been isolated, a weakly positive
control for the IELISA that was from a cow inoculated with B. abortus 19 and negative on the CELISA, and a negative control from
a pool of cattle with no history of B. abortus infection.
Separate controls were supplied by the International Atomic Energy
Agency (IAEA) for the IAEA IELISA kit.
Test procedures.
The RBT antigen was prepared by
Rhone-Merieux, and the assay was performed as described in the National
Animal Diseases Laboratory diagnostic reagents manual (11).
The CFT reagents were prepared, and the assay was performed, as
described in Public Health monograph N74 (12).
The IELISA supplied by the Joint Food and Agriculture Organization
(FAO)-IAEA Division was performed as described in the FAO-IAEA kit. The
basic reagents and protocol have been adapted for this kit
(7). The IELISA (6, 8) supplied by the
Agriculture Canada ADRI was performed as described elsewhere. The
CELISA with smooth lipopolysaccharide (sLPS) as the antigen
(6) was performed as described elsewhere. The CELISA with O
polysaccharide of sLPS (CELISA-OC) as the antigen was performed as
described in 1994 (2). The procedures for each assay are
summarized in Table 1.
Data handling and statistical analysis.
The data for each
country was compiled in a database and divided into negative or
positive results according to serological reactions on both the RBT and
the CFT.
After the results were classified into serologically negative and
positive populations, initial optimal estimates of the criteria between
positive and negative reactions (the cutoff values) were determined by
receiver operating characteristics (ROC) analysis (10).
With the initial estimates of cutoff values, the percentage of samples
that were positive relative to the positive RBT and CFT reactions and
the percentage of samples that were negative relative to the negative
RBT and CFT reactions were calculated and the frequency distributions
were plotted to provide a visual confirmation that the cutoff value was
applicable.
Finally, assays were compared to each other for agreement, and a kappa
statistic was calculated (10).
 |
RESULTS |
The numbers of samples used in this study are presented in Table
2. The samples were divided into three
populations. The negative population was defined as those primarily
from regions that had no history of B. abortus infection and
having negative reactions on both the RBT and the CFT. The positive
population was defined as those samples from infected herds which had
positive reactions on both the RBT and the CFT. The vaccinated
population was defined as those animals that had been vaccinated with
B. abortus 19 according to the regulations in each country.
The exception to this was Argentina, where vaccination is routinely
practiced, and it was difficult to collect samples that were defined as
being from unexposed cattle. Consequently, the data for the negative category and that for the vaccinated category for Argentina were combined into the negative category.
View this table:
[in this window]
[in a new window]
|
TABLE 2.
Number of samples tested in each country in each group
(not exposed, serologically positive, or vaccinated with
B. abortus)
|
|
The data presented in Table 3 is defined
in two ways. The data for the positive population from Argentina,
Chile, Colombia, and Costa Rica and the combined data are percent
positives (%P), or the number of positives found for each ELISA
relative to the positive RBT and CFT reactions from cattle in infected
herds. The Canadian data is actual sensitivity, since the results were derived from animals from which B. abortus had been
isolated. The highest %P, of 100%, for both IELISAs from Colombia and
for the IELISA-ADRI from Costa Rica indicate that it is comparable to
the actual sensitivity achieved by the IELISA-ADRI in the Canadian study. Data for the IELISA-IAEA and the CELISA-OC for Canada was not
part of the original Canadian study and consequently is not available.
View this table:
[in this window]
[in a new window]
|
TABLE 3.
Comparison of percent positive reactions in each country
with actual sensitivity determined in the Canadian study
|
|
The data presented in Table 4 for the
negative population from Argentina, Chile, Colombia, and Costa Rica and
the combined data are similarily defined as percent negatives (%N), or
the number of negatives found for each ELISA relative to the negative RBT and CFT reactions primarily from regions with no history of B. abortus infection. The Canadian data is actual
specificity, since the results were derived from cattle in Canada.
Canada has been free of B. abortus infection in cattle since
1982.
View this table:
[in this window]
[in a new window]
|
TABLE 4.
Comparison of percent negative reactions in each country
with specificity determined in the Canadian study
|
|
The highest %N, of 99.82%, for both IELISAs from Colombia is
comparable to the actual specificity of 99.40% achieved by the IELISA-ADRI in the Canadian study. Data for the IELISA-IAEA and the
CELISA-OC for Canada was not part of the original Canadian study and
consequently is not available.
The %N of the IELISAs and CELISAs relative to the RBT and CFT for the
vaccinated population are presented in Table
5. The %N for each country and that for
the combined data of Chile, Colombia, and Costa Rica are compared to
each other and to the Canadian data. The largest difference in
percentage was between the IELISA-ADRI and the CELISA-sLPS in the
Canadian study. This was 41.4%. The difference between the IELISA-ADRI
and the CELISA-sLPS for the data from Chile was 21.2%. In all cases,
the %N for the vaccinated population was greater for the CELISA-sLPS
than for the IELISA-ADRI or the IELISA-IAEA, although for the
IELISA-IAEA the differences were smaller. Similarly, the %N of
the CELISA-OC was greater than that of the IELISA-ADRI in all cases.
However, the %N of the IELISA-IAEA was greater than that of the
CELISA-OC for Chile and for the data from Costa Rica with calf
vaccination. The maximum difference was 2.4%.
Cutoff values for each ELISA by country are presented in Table
6. The IELISA data is expressed as %P.
The CELISA data is expressed as percent inhibition (%I). For example,
the cutoff value for the IELISA-ADRI for Argentina is 67%P. Samples
greater than or equal to 67%P are positive and samples less than 67%P are negative on this IELISA. The lowest cutoff value for the
IELISA-ADRI was 16%P. The highest cutoff value for the IELISA-ADRI was
70%P, a difference of 54%. Similarly, the lowest cutoff value for the IELISA-IAEA was 14%P, and the highest cutoff value was 73%P, a difference of 59%. The difference for the CELISA-OC and the
CELISA-sLPS was 17 and 26%, respectively, which indicated that the
CELISAs were more specific for the negative population.
Agreements between assays are compared in Table
7. The kappa statistic for each ELISA by
country is presented. For example, the kappa indices of Argentina,
Chile, Colombia, and Costa Rica for the IELISA-ADRI and the IELISA-IAEA
are 0.824, 0.963, 0.994, and 0.850, respectively, indicating good
agreement between the IELISA-ADRI and the IELISA-IAEA despite the
differences in the cutoff values. Except for Costa Rica, the kappa
statistic for all assays indicated good agreement between assays. It is
generally accepted that a kappa statistic greater than or equal to 0.8 indicates good agreement between assays. The kappa results for Costa
Rica were not much lower than 0.8 and were all greater than 0.5, indicating agreement beyond chance.
The cutoff values of each ELISA by country were determined by a
combination ROC analysis and by frequency distributions. The ROC
analyses are presented in Fig. 1 to
5,
along with the respective areas under the curve (AUC). For example, in
Fig. 1a, the optimal cutoff value for the IELISA-ADRI is 67%P. In Fig.
1b, the optimal cutoff value for the IELISA-IAEA is 40%P. In Fig. 1c,
the optimal cutoff value for the CELISA-sLPS is 44%I, while an optimal
cutoff value for the CELISA-OC in Fig. 1d is 35%I. The frequency
distributions are presented in Fig. 6 to
10.
The frequency distribution for the IELISA-ADRI in Fig. 6a shows
considerable overlap between the selected negative and positive
populations. With the cutoff as determined by ROC analysis, it is much
easier to identify the false negatives. The same is true of the
IELISA-IAEA, CELISA-OC, and CELISA-sLPS presented in Fig. 6b, c, and d.
The other frequency distributions for the other countries can be
interpreted in a similar fashion.

View larger version (30K):
[in this window]
[in a new window]
|
FIG. 1.
ROC curves representing each ELISA for data from
Argentina. The cutoff value for each assay is indicated in the upper
right corner of each panel. The AUC is indicative of how well the test
performed. A value of 1.0 is perfect, and a value below the diagonal
line represents reactivity due to chance. (a) IELISA-ADRI, AUC = 0.983. (b) IELISA-IAEA, AUC = 0.983. (c) CELISA-sLPS, AUC = 0.995. (d) CELISA-OC, AUC = 0.991.
|
|

View larger version (30K):
[in this window]
[in a new window]
|
FIG. 2.
ROC curves representing each ELISA for data from Chile.
(a) IELISA-ADRI, AUC = 1.000. (b) IELISA-IAEA, AUC = 0.996. (c) CELISA-sLPS, AUC = 1.000. (d) CELISA-OC, AUC = 1.000. See
the Fig. 1 legend for additional explanation of the data.
|
|

View larger version (30K):
[in this window]
[in a new window]
|
FIG. 3.
ROC curves representing each ELISA for data from
Colombia. (a) IELISA-ADRI, AUC = 1.000. (b) IELISA-IAEA, AUC = 1.000. (c) CELISA-sLPS, AUC = 0.994. (d) CELISA-OC, AUC = 0.999. See the Fig. 1 legend for additional explanation of the data.
|
|

View larger version (31K):
[in this window]
[in a new window]
|
FIG. 4.
ROC curves representing each ELISA for data from Costa
Rica. (a) IELISA-ADRI, AUC = 0.992. (b) IELISA-IAEA, AUC = 0.991. (c) CELISA-sLPS, AUC = 0.974. (d) CELISA-OC, AUC = 0.969. See the Fig. 1 legend for additional explanation of the data.
|
|

View larger version (30K):
[in this window]
[in a new window]
|
FIG. 5.
ROC curves representing each combined ELISA for data
from Argentina, Chile, Colombia, and Costa Rica. (a) IELISA-ADRI,
AUC = 0.985. (b) IELISA-IAEA, AUC = 0.989. (c) CELISA-sLPS,
AUC = 0.995. (d) CELISA-OC, AUC = 0.995. See the Fig. 1
legend for additional explanation of the data.
|
|

View larger version (29K):
[in this window]
[in a new window]
|
FIG. 6.
Frequency distribution of ELISA data from Argentina.
Open bars, serologically negative samples; closed bars, serologically
positive samples. The numbers in each class limit are indicated on top
of the bars. (a) IELISA-ADRI. (b) IELISA-IAEA. (c) CELISA-OC. (d)
CELISA-sLPS.
|
|

View larger version (25K):
[in this window]
[in a new window]
|
FIG. 7.
Frequency distribution of ELISA data from Chile. (a)
IELISA-ADRI. (b) IELISA-IAEA. (c) CELISA-OC. (d) CELISA-sLPS. See the
Fig. 6 legend for additional explanation of the data.
|
|

View larger version (22K):
[in this window]
[in a new window]
|
FIG. 8.
Frequency distribution of ELISA data from Colombia. (a)
IELISA-ADRI. (b) IELISA-IAEA. (c) CELISA-OC. (d) CELISA-sLPS. See the
Fig. 6 legend for additional explanation of the data.
|
|

View larger version (23K):
[in this window]
[in a new window]
|
FIG. 9.
Frequency distribution of ELISA data from Costa Rica.
(a) IELISA-ADRI. (b) IELISA-IAEA. (c) CELISA-OC. (d) CELISA-sLPS. See
the Fig. 6 legend for additional explanation of the data.
|
|

View larger version (29K):
[in this window]
[in a new window]
|
FIG. 10.
Frequency distribution of combined ELISA data from
Argentina, Chile, Colombia, and Costa Rica. (a) IELISA-ADRI. (b)
IELISA-IAEA. (c) CELISA-OC. (d) CELISA-sLPS. See the Fig. 6 legend for
additional explanation of the data.
|
|
 |
DISCUSSION |
ELISAs have a distinct advantage over conventional serological
tests in that they are primary binding assays that do not rely on
secondary properties of antibodies such as their ability to agglutinate
or to fix complement. Secondly, ELISAs can be tailored to be more
specific by using highly purified reagents such as antigens and
monoclonal antibodies.
In Canada, which is free of brucellosis in domestic animals, both the
IELISA and the CELISA were recently validated (9). Approximately 8,000 samples from cattle with no evidence of B. abortus infection were collected and tested in both the IELISA and
the CELISA. Similarly, 692 samples from cattle from which B. abortus was isolated from milk or tissues were also tested. Another 261 samples from cattle that were vaccinated with B. abortus 19 and that contained residual antibodies were tested as
well.
Unlike in Canada, conditions in Latin America for validation of assays
are different. It is more difficult to define negative and positive
sera because diagnosis is based on serological evidence or the
isolation of B. abortus from herds rather than from
individual cattle. In most countries, areas overlap between regions
free of B. abortus and regions that contain infected herds,
and strain 19 vaccination is widely practiced. For these reasons and
for consistency, the negative population and positive population were defined based on the RBT and the CFT reactions in each country under
study. As well, determining the B. abortus 19 vaccination status of cattle is sometimes difficult due to insufficient data being
available, including the time of vaccination, the number of times that
cattle were vaccinated, and identification of cattle that were
vaccinated. The numbers of samples defined as positive, negative, and
vaccinated are tabulated in Table 2.
Comparison of %P is summarized in Table 3. The results are not
dissimilar from the results obtained in the Canadian study (9). Both the IELISA and the CELISA achieved a sensitivity estimate of 100% in Canada. The results obtained in Latin America were
comparable. Percent positive values obtained ranged from 92.10% for
the CELISA-OC in Costa Rica to 100% for the CELISA-sLPS in Chile, the
IELISA-ADRI in Colombia, the IELISA-IAEA in Colombia, and the
IELISA-ADRI in Costa Rica. When the data was combined for all countries
(except Canada), the performance of both CELISAs was marginally better
than that of the IELISAs (presented in Table 3). The maximum difference
between the CELISAs and the IELISAs for the combined data is 1.19%.
The CELISA-sLPS at 97.47% detects 11.9 more positive reactions per
1,000 animals than does the IELISA-IAEA at 96.28%.
Comparison of %N is presented in Table 4. The specificity for the
IELISA in Canada was 99.40%, while the specificity for the CELISA was
99.90% (9). The results obtained in Latin America were
similar. The lowest %N achieved was 93.35% for the CELISA-OC in Costa
Rica. The highest %N achieved was 99.82% for the IELISA-ADRI, IELISA-IAEA, and CELISA-OC in Colombia. When the data was combined for
all the countries (except Canada), it is obvious that the overall
performance of both CELISAs is better than that of the IELISAs
presented in Table 4. The maximum difference between the CELISAs and
the IELISAs for the combined data is 4.75%. The CELISA-sLPS at 98.32%
is more specific than the IELISA-ADRI at 93.57%. Thus, the IELISA-ADRI
detected 47.5 more animals per 1,000 animals than did the CELISA-sLPS.
Comparison of the %N for vaccinated cattle is tabulated in Table 5.
The results of the Canadian study indicated that the CELISA-sLPS was
capable of distinguishing animals that were vaccinated or negative from
those that were infected, in the majority of the cases. In the Canadian
study, the specificity of the IELISA-ADRI was 56.30% while the
specificity for the CELISA-sLPS was 97.70%. Similar results were
achieved in Latin America. In Chile, the %N for the IELISA-ADRI was
78.82% while the %N for both CELISAs was 94.44 and 100%. In
Colombia, the %N for both IELISAs was 86.76 and 87.57%, respectively.
The %N for both CELISAs was 95.50 and 92.25%. The combined data
clearly indicates that the %N of the CELISAs as presented in Table 5
is better than that of the IELISAs for distinguishing vaccinal
antibody. The maximum difference between the CELISAs and the IELISAs
for the combined data is 5.98%. The CELISA-sLPS for the combined data
at 96.51% is more specific than the IELISA-ADRI at 90.53%. The
CELISA-sLPS misinterprets as positives 59.8 fewer vaccinated animals
per 1,000 animals than does the IELISA-ADRI.
Ideally, harmonization of cutoff values should be the same in each
country for the IELISAs or for the CELISAs. However, analysis of data
indicated that this was not possible. The cutoff values for each
country and for the combined data were determined by ROC analysis as
presented in Fig. 1 to 5 and tabulated in Table 6. From Table 6, the
only assay that had cutoff values approximating the 30% chosen for
Canada was the CELISA-sLPS, except for Costa Rica. The frequency
distributions presented in Fig. 6 to 10 show the difficulty in choosing
an optimal cutoff value for each assay. For instance, most of the
frequency distributions for the IELISA have some overlap between the
negative and the positive populations. The exceptions to this were the
frequency distributions from Colombia. The reason for the binomial
distribution is better separation of the negative and positive sera.
The sera were from defined areas free from B. abortus
infection and from areas with a relatively high prevalence of
infection. Despite the differences in how the IELISA-ADRI and the
IELISA-IAEA were performed, the distribution patterns were very
similar. This became quite evident when the frequency distributions of
the combined data for the IELISAs presented in Fig. 10 were examined.
The distribution patterns of the CELISAs, although different from those
of the IELISAs, were similar to each other, and again the similarity
was quite evident from the frequency distribution of the combined data
presented in Fig. 10. Choosing a cutoff value solely on the basis of
frequency distribution could give erroneous results. The frequency
distributions of the CELISAs were marginally better than those of the
IELISAs due to less overlap between the selected negative and positive
populations. However, obtaining the optimal percentage of positives and
percentage of negatives for each assay in each country was best
determined by ROC analysis and frequency distributions together to get
a clearer picture in each instance.
The ROC curves presented in Fig. 1 to 5 all had AUC greater than 0.95. An AUC of 0.95 indicates that a randomly selected individual animal
from a positive population will have a test value greater than that of
a randomly selected individual animal from the negative population 95%
of the time. The lowest AUC was 0.969 for the CELISA-OC in Costa Rica,
while the highest AUC was 1.000 for the IELISA-ADRI, the CELISA-sLPS,
the CELISA-OC in Chile, and the IELISA-ADRI and the IELISA-IAEA in
Colombia. Both CELISAs for the combined data had an AUC of 0.995, which
was approximately 1% better than that of the IELISAs.
Finally, a comparison of agreement between assays was calculated and
presented in Table 7. A kappa statistic of 1 indicates perfect
agreement between assays. A kappa of 0.5 indicates agreement beyond
chance. It is generally accepted that kappa indices greater than or
equal to 0.8 indicate good agreement between tests. The best agreement
was 0.994 between the IELISAs in Colombia. Again, this is probably due
to better separation of the negative and positive populations. The
lowest kappa statistic was 0.720 between the CELISAs from Costa Rica,
where separation of negative and positive populations was more
difficult. The highest kappa for both CELISAs was 0.981 from Chile.
Overall, the kappa statistics for all the assays were good, indicating
good agreement among all assays.
Generally, the technical performance of the assays was good and the
results were similar to results obtained in the Canadian study.
However, there are some reasons why the results could be improved.
Firstly, a bias was introduced in the study. The selected negative and
positive populations were defined according to the RBT and CFT
reactions. The RBT can produce false-positive results, which when used
to define sera can affect the sensitivity of the assay being validated.
Secondly, a better separation of the negative and positive populations
would have produced better results. For example, if individual animals
with proven infection based on isolation of the organism had been
selected, instead of positive animals from infected herds, the
sensitivity values should have been higher. Thirdly, the RBT and the
CFT both detect antibody resulting from B. abortus 19 vaccination or from exposure to cross-reacting antigens. Therefore, the
results are biased against the CELISAs, which eliminate many such
reactions.
Based on the combined data, the CELISA-sLPS was the best-performing
ELISA. It detected 1.19% more positives in the selected positive
population, 4.75% fewer positives in the selected negative population,
and 5.98% fewer positives in the selected vaccinated population. The
implication of this is important. For example, in a population of
15,000,000 animals with a high incidence of brucellosis the CELISA-sLPS
would detect 712,500 fewer false positives and, if vaccination were
part of the control program, 897,000 fewer false positives. By using
the CELISA-sLPS as the primary screening assay in an eradication and
control program, significant savings in repeat testing and elimination
of other conventional assays can be realized. In addition, the
CELISA-sLPS is less costly in reagents than is conventional assays and
has excellent quality control, leading to additional savings.
This project was funded by the Joint FAO-IAEA Commission on
Animal Health and the Canadian Food Inspection Agency.
| 1.
|
Carlsson, H. E.,
B. Hurvell, and A. A. Lindberg.
1976.
Enzyme linked immunosorbent assay (ELISA) for titration of antibodies against Brucella abortus and Yersinia enterocolitica.
Acta Pathol. Microbiol. Scand.
84:168-176.
|
| 2.
|
Gall, D., and K. Nielsen.
1994.
Improvements to the competitive ELISA for the detection of antibody to Brucella abortus in cattle.
J. Immunoass.
15:277.
|
| 3.
|
MacMillian, A.
1990.
Conventional serological tests, p. 153.
In
K. Nielsen, and J. R. Duncan (ed.), Animal brucellosis. CRC Press, Inc., Boca Raton, Fla.
|
| 4.
|
MacMillian, A.,
I. Greiser-Wilke,
V. Moennig, and L. Mathias.
1990.
A competitive enzyme immunoassay for brucellosis diagnosis.
Dtsch. Tieraerztl. Wochenschr.
97:83.
|
| 5.
|
Nielsen, K.,
J. Cherwonogrodzky,
J. R. Duncan, and D. R. Bundle.
1989.
Enzyme immunoassay for the differentiation of the antibody response of Brucella abortus infected or strain 19 vaccinated cattle.
Am. J. Vet. Res.
50:5[Medline].
|
| 6.
|
Nielsen, K.,
D. Gall,
W. Kelly,
A. Vigliocco,
D. Henning, and M. Garcia.
1996.
Immunoassay development: application to enzyme immunoassay for diagnosis of brucellosis. Monograph.
Agriculture and Agri-Food Canada, Animal Diseases Research Institute, Nepean, Ontario, Canada.
|
| 7.
|
Nielsen, K. H.,
P. F. Wright,
W. A. Kelly, and J. H. Cherwonogrodzky.
1988.
A review of enzyme immunoassay for the detection of antibody to Brucella abortus in cattle.
Vet. Immunol. Immunopathol.
18:331-347[Medline].
|
| 8.
|
Nielsen, K. H.,
D. Gall,
W. Kelly,
D. Henning, and M. M. Garcia.
1992.
Enzyme immunoassay: application to diagnosis of bovine brucellosis. Monograph.
Agriculture Canada, Nepean, Ontario, Canada.
|
| 9.
|
Nielsen, K. H.,
L. Kelly,
D. Gall,
S. Balsevicius,
J. Bossé,
P. Nicoletti, and W. Kelly.
1996.
Comparison of enzyme immunoassays for the diagnosis of bovine brucellosis.
Prev. Vet. Med.
26:17-32.
|
| 10.
|
Schoojans, F.,
A. Zalata,
C. E. Depuydt, and F. H. Comhaire.
1995.
MedCalc: a new computer program for medical statistics.
Comput. Methods Programs Biomed.
48:257-262[Medline].
|
| 11.
|
U.S. Department of Agriculture.
1965.
NADL diagnostic reagents manual, p. 65C-65E.
Diagnostic Reagents Division, National Animal Diseases Laboratory, U.S. Department of Agriculture, Ames, Iowa.
|
| 12.
|
U.S. Department of Health, Education, and Welfare.
1965.
Standardized diagnostic complement fixation and adaptation to the microtest. Public health monograph N74. U.S.
Department of Health, Education, and Welfare, Washington, D.C.
|
| 13.
|
Wright, P. F.,
K. Nielsen, and W. Kelly.
1990.
Primary binding assay techniques for the serodiagnosis of bovine brucellosis enzyme immunoassay, p. 199.
In
K. Nielsen, and J. R. Duncan (ed.), Animal brucellosis. CRC Press, Inc., Boca Raton, Fla.
|