Clinical and Vaccine Immunology, July 2006, p. 764-767, Vol. 13, No. 7
1071-412X/06/$08.00+0 doi:10.1128/CVI.00199-05
Copyright © 2006, American Society for Microbiology. All Rights Reserved.
Centre for Infectious Diseases and Microbiology, Institute of Clinical Pathology and Medical Research, Westmead Hospital, Westmead,1 National Centre for Immunisation Research and Surveillance of Vaccine Preventable Diseases, The Children's Hospital at Westmead and the University of Sydney, New South Wales 2145, Australia2
Received 23 June 2005/ Returned for modification 4 August 2005/ Accepted 21 April 2006
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
The neutralizing antibody test (NT) is considered the most specific indicator of protective mumps antibodies, but it is labor-intensive and difficult to perform (4, 10). Enzyme-linked immunosorbent assays (ELISAs) are simple, rapid, and suitable for automation and so ideally suited to large-scale mumps serosurveys (17). Generally they are reported to be more sensitive than the NT (2, 5, 9, 13, 15, 16). However, the NT can detect functional antibody of any class and has been shown to detect low levels of specific measles antibody below the level of detection of immunoglobulin G (IgG) binding antibody assays, such as ELISAs (3). It is plausible that the same could apply to mumps antibody.
In this study we compared the performance of two commercial ELISAs (Enzygnost Anti-Parotitis-Virus/IgG [Dade Behring, Marburg, Germany] and Mumps IgG Screen ELISA [Microimmune Ltd., Brentford, Middlesex, United Kingdom]) for the detection of serum IgG antibody to the mumps virus and for suitability for a national serosurvey.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Antibody assays. Both ELISAs detect IgG (antigen binding antibody) to the mumps virus by the indirect ELISA technique. The assays were performed on all sera according to the manufacturers' specifications with reagents supplied with the kits. Sera giving equivocal results were retested by the same method and reclassified as positive or negative if appropriate.
All sera with sufficient remaining volume that gave persistently equivocal results and/or discrepant results between the two ELISAs were tested by the NT. A representative sample of 7.5% (90/1,197) of sera that were positive and 70% (102/146) of sera that were negative in both ELISAs were also tested by NT, and the results were used to predict the distribution of NT results for the remaining sera.
(i) Enzygnost Anti-Parotitis-Virus/IgG.
The materials supplied with the test kit included Reference P/N, containing human IgG specific for mumps virus, and microtitration plates containing six paired strips of eight wells. The first well of each pair was coated with antigen derived from simian kidney cells infected with the Enders strain of the mumps virus, and the second (control) well was coated with uninfected simian kidney cells. Each specimen and the P/N were tested in the paired wells at a dilution of 1:231. The P/N was run at the beginning and end of each plate. On completion of the assay, the difference in absorbance between the antigen-positive and control antigen wells,
A, was calculated for each specimen and the Reference P/N. A plate run was validated when
A for each Reference P/N was
0.5. The specimens were reported as negative (
A < 0.100), positive (
A > 0.200), or retest/equivocal (0.100
A
0.200).
(ii) Microimmune Mumps IgG Screen ELISA.
The Microimmune assay detects IgG antibodies specifically directed against the mumps nucleoprotein. Reagents supplied with the kit included positive-control and negative-control (NC) sera and a microtitration plate with each well coated with a recombinant mumps nucleoprotein antigen. The recombinant nucleoprotein was made in yeast cells and derived from a wild-type United Kingdom mumps isolate designated the Gloucester strain (18). The controls were tested as supplied, without further dilution, in the first four wells of each plate; the NC was tested in triplicate. The specimens were tested at a dilution of 1:201, one well per specimen. A plate run was validated when the optical density (OD) of the positive control was >0.4 and the OD of each of the three NC wells was
0.2. The cutoff (CO) was determined by adding 0.4 to the mean of the three NCs. The OD of each individual NC had to be within 20% of the mean of the NCs; if one of the three OD values differed by more than 20%, it was omitted and the mean value recalculated. Specimens were reported as positive [OD
(CO x 1.1)], negative [OD
(CO x 0.9)], or retest/equivocal [(CO x 0.9) < OD < (CO x 1.1)].
Neutralization test. Sera inactivated at 56°C for 30 min were prediluted to 1:2.5 in dilution buffer (minimal essential medium containing 4% fetal calf serum). Twenty-five microliters of the 1:2.5 dilution was then transferred to each of the first three wells of a tissue culture plate (Falcon; Becton Dickinson, Franklin Lakes, NJ). Commencing with the third well, doubling dilutions were made in 25-µl volumes of dilution buffer to a final dilution of 1:80. The Jeryl Lynn strain of mumps virus (MUMPSVAX*; Merck & Co., Inc., Whitehouse Station, NJ), grown and harvested after two passages in Vero cells, was diluted to give 100 50% tissue culture infective doses, and 25 µl was added to all wells with the exception of the serum controls (25 µl diluent). Following incubation at 4°C for 90 min, 200 µl of a Vero cell suspension (5 x 105 cells/ml) was added to all wells. The plates were incubated at 37°C under 5% CO2 and read microscopically on day 6 for evidence of cytopathogenic effect. The end point was read at total inhibition of all cytopathogenic effect. A neutralizing antibody titer of <2.5 (i.e., no well showed total inhibition) was considered negative. A positive serum control was included in each plate, and the viability of the virus and cell culture was confirmed in the last plate of each run.
Statistical analysis. The percentages of sera with positive, negative, and equivocal results were determined for each age group and assay. The results for each assay, both overall and for each age group, were compared using the chi-square test. P values of <0.05 were considered statistically significant.
Because different types of assay measure different types of antibody and there is no accepted "gold standard" for both sensitivity and specificity, we used two different "gold standards." First, sensitivity and specificity were estimated by comparing the results of each ELISA with those of the NT. The NT results were extrapolated for all sera (see above), on the assumption that the sera tested in each category (equivocal, discrepant, or consistent between the two ELISAs) were representative. The criteria for the second "gold standard" were as follows: a result was regarded as positive when either the NT or both ELISAs were positive and as negative when the NT and one or both ELISAs were negative (on the assumption that false-positive results were more likely with ELISA).
Ethics approval. The study was approved by the Human Research Ethics committee of the Western Sydney Area Health Service and the State-wide Health Confidentiality and Ethics Committee of the New South Wales Health Department.
| RESULTS |
|---|
|
|
|---|
12 years), the Microimmune assay gave consistently higher seropositive rates (
93%) than the Enzygnost assay, with its lower and more variable rates (54 to 82%).
|
20 years than in younger age groups, but this difference was statistically significant only for the Microimmune assay (P = 0.02). The overall agreement between the Enzygnost and Microimmune ELISAs was 70.7% (1,353/1,915) (Table 2).
|
10, with 71% (23/32) giving a titer of 2.5. Most (91%) were from children, who would have received only a single dose of the measles-mumps-rubella (MMR) vaccine and would have had minimal exposure to mumps infection: 68% (22/32) were aged 1 to 6 years, and 23% (7/32) were aged 7 to 16 years. Only 3% (3/90) of sera positive by both ELISAs were negative by the NT. By extrapolation of the NT results from the sample of sera tested, it was estimated that approximately 85% of all sera would have been positive by the NT. Using either "gold standard," sensitivity was higher, but specificity lower, for the Microimmune than for the Enzygnost assay (Table 4).
|
|
| DISCUSSION |
|---|
|
|
|---|
The pattern by age for both assays reflects the history of vaccination in Australia. A monovalent mumps vaccine was licensed for use on infants aged 12 to 15 months in 1980, and routine mumps vaccination was introduced in 1982 as a combined measles-mumps vaccine (7). In 1989, this was replaced by the MMR vaccine for 12-month-old infants, and in 1994, a second dose of MMR was introduced for children aged 10 to 16 years. Therefore, most individuals aged up to 20 years in 1997 to 1998, when sera were collected for this study, would have been eligible for at least one dose of mumps vaccine. Hence, the higher seroprevalence and low proportion of equivocal results in the over-20 age groups in both assays are probably due to stronger immune responses induced by exposure to the wild-type mumps virus, compared with vaccine-induced immunity in younger subjects.
Despite the similar trends by age group, the assays provide significantly different estimates of seroprevalence. Both assays are indirect ELISAs, but they use different antigens to coat the wells of the microtitration plates: purified Enders strain mumps virus in the Enzygnost test and recombinant mumps nucleoprotein in the Microimmune assay. Exposure to mumps vaccine elicits an immune response to both the S (soluble nucleoprotein) and V (viral envelope) antigens; therefore, detection of antibody to nucleoprotein is an indicator of a specific antibody response to exposure. The Microimmune assay specifically measures IgG antibody to nucleoprotein. However, core proteins, such as nucleoprotein, are also prominent antigens in the whole-virus lysate (14). It is therefore not clear why there is such an apparent difference between the assays.
Comparison of ELISA (IgG antigen binding assay) and NT (a whole-antibody functional assay) results also demonstrates differences between them, as described above. The NT apparently can detect functional antibody below the level of detection by ELISA, as shown by the relatively high proportion of sera that were negative by both ELISAs but positive (at low titers) by the NT. The significantly higher starting dilutions for both ELISAs than for the NT (
1:200 compared with 1:2.5) could also contribute to these differences, which are similar to those recently reported in a comparison between two measles IgG antibody ELISAs and a plaque reduction NT (3). We used a second "gold standard" in part to balance these differences, although both are still weighted in favor of the NT, because of its high specificity. Given that there is a difference between the two ELISAs and that no "gold standard" is ideal, we believe that the Microimmune assay provides a more valid estimate of population seroprevalence than the Enzygnost assay for the following reasons.
First, by both of the "gold standards" used in this study, the Microimmune ELISA is more sensitive than the Enzygnost ELISA. The higher sensitivity of the Microimmune assay produces a seroprevalence closer to that expected on the basis of known MMR immunization rates. For instance, mumps seroprevalence measured using the Microimmune assay is much closer to that of measles (8) than that estimated by the Enzygnost assay. In addition, the expected proportion of seropositive 2- to 6-year-olds, based on vaccination coverage and effectiveness estimates, is closer to levels demonstrated by the Microimmune assay; MMR uptake for 2- to 6-year-olds in 1998 was estimated to be
90% (11). Given that vaccine effectiveness is between 80 and 100% (19), we would expect 72 to 90% of 2- to 6-year-olds to be seropositive. This range encompasses the result from the Microimmune but not the Enzygnost assay.
Second, we believe that the Microimmune assay is more specific than suggested by comparison with both "gold standards." Theoretically, low specificity could be due to cross-reactivity with the parainfluenza virus (9). Sera containing parainfluenza virus antibody were not available to confirm this hypothesis, but Frankova et al. (6) found that cross-reactivity was eliminated when purified mumps nucleoprotein antigen was used for ELISA, as it is in the Microimmune assay. In addition, although it is (probably) highly specific, the NT has been shown to be generally less sensitive (2, 5, 9, 13, 15, 16) than ELISA, notwithstanding its ability to detect low levels of functional antibody not detectable by ELISAs. Moreover, the strain of mumps virus used in the NT is known to affect its performance (14). Thus, the apparently low specificity of the Microimmune assay appears to be due to a lack of sensitivity of the "gold standards" used in this study.
The higher sensitivity of the Microimmune assay may also explain why seroprevalence in the 17- to 19-year age group was lower than in surrounding age groups only in the Enzygnost assay. This probably reflects a combination of waning vaccine-induced immunity and low rates of natural immunity in this first cohort of children eligible for immunization in the 1980s. Presumably the more sensitive Microimmune assay can detect lower antibody levels in this age group.
Although there are no internationally agreed serological correlates of protective immunity, detection of mumps IgG antibody is an indication of past infection or immunization (1). Thus, seroprevalence of IgG antibodies is central to the determination of herd immunity in a population. The level of immunity required to achieve herd immunity is 85% to 90% (12). In this study seroprevalence, as determined by the Microimmune assay, was at an acceptable level in all age groups from the 7- to 11-year group up. However, as determined by the Enzygnost assay, seroprevalence would not be considered sufficient for herd immunity.
In conclusion, these results exemplify how assay selection can have a significant impact on age-specific and thus population-based estimates of immunity to mumps. In this study, the Microimmune assay produced a higher proportion of positive results and fewer equivocal results than the Enzygnost assay and, based on expected levels of immunity and comparison with the results of two "gold standards," probably provides a more realistic estimate of seroprevalence in the community.
| ACKNOWLEDGMENTS |
|---|
We thank the staff of the 45 laboratories (http://immunise.health.gov.au/metadata/measeval.htm, p. 8-9) that provided the sera, the nurses at the Children's Hospital at Westmead Centre for Immunization Research, and laboratory staff at the Centre for Infectious Diseases and Microbiology, in particular Beverley Bowcock, for help in processing and testing the sera.