Clinical Virology Laboratory, University of Colorado Hospital,1 Departments of Pediatrics and Medicine, University of Colorado School of Medicine, Denver, Colorado2
Received 3 November 2004/ Returned for modification 9 December 2004/ Accepted 11 January 2005
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
Present methods of diagnosing viral respiratory infections include tissue culture isolation (TC); rapid shell vial detection (SV); direct antigen detection using enzyme, fluorescence, optical, or chromatographic immunoassays; and PCR (10, 14). The rapid influenza immunoassays (RIIA) provide a result in 15 to 30 min. They are frequently used at the point of care or as a screening method in full-menu virology laboratories because they allow the institution of therapeutic or prophylactic measures with minimal delays. The goal of this study was to compare three RIIA kits marketed for combined detection of influenza virus A and B.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Directigen uses a ColorPAC device with a flow controller shared by the influenza A and B wells. The specimen is mixed with extraction reagent E (proprietary) and filtered before being added to the A and B test wells. After the flow controller is removed, the wells are allowed to completely absorb four other proprietary reagents added sequentially, with 2- and 5-min incubations before the third and last reagents, respectively. A positive influenza A or B specimen generates a purple triangle against a light yellow to light purple background. A valid negative result is indicated by a purple control dot against the light yellow to light purple background.
The EZ test has separate influenza A and B devices. Specimens are mixed with an extraction reagent and filtered before being dispensed onto the A and B wells. After a 15-min incubation, influenza A- or B-containing specimens will result in a visible reddish-purple line against a white to light pink background in addition to the positive control line in the appropriate well. A valid negative result is indicated by the absence of the test line with a visible control line.
NOW has separate influenza A and B devices. Specimen is added directly to each apparatus. After a 15-min incubation, a positive influenza A or B specimen generates a pink to purple sample line in addition to the control line. A valid negative influenza A or B result is represented by the presence of the control line only.
Respiratory TC was performed by inoculating 0.3 ml of specimen into each of six tissue culture tubes: two rhesus monkey kidney tubes, each from a different vendor (Diagnostic Hybrids Inc., Athens, Ohio, and Viromed, Minnetonka, Minn.), and one Hep-2 tube (American Type Culture Collection), one A549 tube (American Type Culture Collection), and two primary human embryonic lung fibroblast tubes prepared in our laboratory. Tubes were incubated at 37°C for up to 28 days in maintenance medium consisting of Eagle's medium (Diagnostic Hybrids Inc.) with penicillin, streptomycin, amphotericin B, and 2% fetal bovine serum (Sigma, St. Louis, Mo.), except for rhesus monkey kidney tubes, which were maintained in fetal bovine serum-free medium. One human embryonic lung fibroblast tube was incubated at 33°C on a roller drum to enhance rhinovirus isolation. Medium was changed at 24 to 48 h after inoculation, after each of the weekly guinea pig red blood cell hemadsorptions, and as dictated by the appearance of the monolayer. Tubes were examined by light microscopy daily during the first week after inoculation and thrice weekly thereafter for up to 28 days. Monolayers that showed cytopathic effect or positive hemadsorption were scraped and stained with specific monoclonal antibodies. The procedure entailed spotting cells onto slides, followed by acetone fixation and staining with type-specific monoclonal antibodies (Dako, Carpenteria, Calif.). Slides were read with a fluorescence microscope. A positive result was defined as the presence of bright green fluorescence in the cytoplasm of two or more cells.
SV cultures. R-Mix SVs (Diagnostic Hybrids Inc.) were activated by incubation at 37°C for 2 to 4 h prior to inoculation. Growth medium was replaced by Refeed medium (Diagnostic Hybrids Inc.) immediately prior to infection, after which cultures were inoculated with 0.2 ml of the clinical specimen. Thereafter, R-Mix SVs were centrifuged at 700 x g for 60 min and incubated at 37°C. Coverslips harvested at 48 h were stained as described above.
PCR.
RNA, extracted from clinical specimens with a QIAGEN RNA extraction kit (QIAGEN, Valencia, Calif.), was reverse transcribed and amplified during 40 cycles using the one-step reverse transcription-PCR kit (QIAGEN) and the following biotinylated primer pairs: for influenza A, 5'-AAGGGCTTTCACCGAAGAGG and 5'-CCCATTCTCATACTGCTTC, and for influenza B, 5'-ATGGCCATCGGATCCTCAAC and 5'-TGTCAGCTATTATGGAGCTG (7). The amplicon was allowed to bind to capture probes 5'-GTCCTCGGAGGACTTGAATGGAATGAT and 5'-GTCAAGAGCACCGATTATCAC for influenza A and influenza B, respectively (7) (0.5 µg/ml in 1 M ammonium acetate), previously immobilized onto 96-well plates. Bound amplicon was detected with streptavidin-horseradish peroxidase (R&D Systems, Minneapolis, Minn.) and tetramethylbenzidine colorimetric substrate (R&D Systems). The optical density was measured at 450 nm using an enzyme immunoassay (EIA) reader. A sample with an optical density of
0.6 was considered positive for the presence of the nucleic acid of interest. This test has an analytic sensitivity of
1 50% tissue culture infective dose U/ml. Positive and negative controls, consisting of viral isolates and water, respectively, were included in each run. The runs were considered valid if the absorbance of the negative control was <0.3 and that of the positive control was >1.0.
Statistical analysis. The sensitivity and specificity of each kit were calculated and compared by the chi-square test and the Fisher exact test using Instat3 software (GraphPad Software, San Diego, Calif.).
| RESULTS |
|---|
|
|
|---|
8 years old and 148 individuals who were
12 years old (and 2 individuals of unknown ages). The gender distribution showed 83 females, 92 males, and 3 cases in which the gender was not recorded. There were 30 influenza A and 7 influenza B specimens that met true-positive criteria (Table 1). Specificity varied from 94% for EZ (95% confidence interval [CI] of 89 to 98%) to 98% for Directigen (95% CI of 91 to 100%) and was not significantly different across tests. There were seven specimens that were tested by PCR, since they were positive by only one antigen detection method. All seven specimens had negative PCR results, confirming the TC results. Combined influenza A and B sensitivities of Directigen, EZ, and NOW were 56% (95% CI of 35 to 76%), 39% (95% CI of 20 to 61%), and 76% (95% CI of 50 to 93%), respectively. There was a marginal difference in sensitivity when all three tests were analyzed together (P = 0.06). Individual comparisons showed that NOW had significantly higher sensitivity than EZ (P = 0.03). The corresponding positive predictive values (PPV) and negative predictive values (NPV) were as follows: for Directigen, 93% (95% CI, 68 to 100%) and 85% (95% CI, 74 to 92%), respectively; for EZ, 56% (95% CI, 30 to 80%) and 89% (95% CI, 83 to 94%), respectively; and for NOW, 93% (95% CI, 66 to 100%) and 81% (95% CI, 58 to 95%), respectively. NPV were not significantly different among kits, but the PPV was lower for EZ than for either of the other two tests (P = 0.01).
|
|
Specificities of Directigen, EZ, and NOW in the younger age group were 100, 94, and 100%, respectively, whereas in the older age group they were 98, 94, and 94%, respectively. These data indicate that age did not significantly impact the specificities of the assays.
Technical characteristics of influenza A and B rapid tests. To determine differences in technical challenge and labor demands of each test, we compared them with respect to total duration of the test, technical complexity, hands-on time, ease of interpretation of results, specimen requirements, and cost (Table 3). Directigen was the most rapid (8 min) but had the most challenging readout and the highest cost of the three kits. EZ and NOW required similar lengths of time and technical skills. EZ was the most economical test when both labor and reagents were factored into the analysis.
|
| DISCUSSION |
|---|
|
|
|---|
The type of specimen influenced the sensitivity of each assay. We were particularly interested in the performance of these rapid tests on BAL fluids because they are frequently obtained from the sickest patients and immunocompromised hosts. Previous studies that included BAL fluids among specimens used to investigate the diagnostic accuracy of influenza rapid detection kits (6) did not provide a breakdown of sensitivity by specimen type. In this study, BAL fluids were associated with decreased sensitivity when tested by EZ. The number of influenza-containing BAL specimens was too small to draw definitive conclusions with respect to the other kits. The sensitivity of rapid antigen detection tests with BAL fluids deserves further investigation.
Age is a known determinant of influenza viral shedding and, therefore, the amount of antigen available for rapid detection. Children typically have less immunity to influenza than adults and tend to excrete large amounts of virus. A test with marginal sensitivity for influenza would be expected to show significant differences between children and adults. In this study, we stratified the subjects as younger or older than 9 years. This is the age when the recommendations for influenza vaccination change from two doses to a single dose based on the assumption that by this age most individuals would have already been infected by the wild-type virus (9). The data showed a trend toward higher sensitivity in the younger group across all tests. However, the number of specimens from children who were <9 years old was much smaller than the number from the older group (28 versus 148), which limited our ability to reach statistical significance. There were no significant differences among the three assays with respect to sensitivity in the younger age group, but NOW was superior to EZ in the older group. Due to the small sample size of this study, we could not study the interaction between age and specimen type on the sensitivity of each kit.
The rapid detection kits have streamlined execution formats conducive to point-of-care use. The Directigen assay, which was most cumbersome to perform and difficult to interpret, has the potential to provide results twice as fast as the other kits, although the 7.5- and 8-min differences for EZ and NOW assays, respectively, do not seem clinically relevant. Cost probably varies with discounts obtained by specific laboratories. At our institution, EZ was the most economical test.
A limitation of this study was the unequal number of specimens tested by the different methods. NOW, in particular, was tested on only 35 specimens. However, the number of positive samples tested by NOW, 17, was not significantly different from the number of positive specimens used in the evaluation of the Directigen and EZ tests (25 and 23, respectively).
Rapid tests constitute an important screening tool for management of influenza on individual and community levels. The high PPV of these tests allows early interventions that may limit the spread of the virus in high-risk populations (1). Conversely, the NPV of these tests is not always optimal, and negative results need to be backed up by culture or PCR whenever they play a critical role for clinical management.