Department of Virology, EBV Reference Center, University Homburg/Saar, Homburg/Saar,1 HISS Diagnostics GmbH, Freiburg,2 Institute of Virology, University of Dresden, Dresden, Germany3
Received 30 May 2002/ Returned for modification 6 August 2002/ Accepted 20 September 2002
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
The diagnosis of a primary EBV infection builds upon an EBV-specific test for immunoglobulin G (IgG) and IgM antibodies to viral capsid antigens (VCA) and IgG antibodies to the EBV nuclear antigens (EBNA), especially EBNA-1, as the minimal requirement (10). Although the "gold standard" technique in EBV diagnostics is the indirect immunofluorescence assay (IFA), the enzyme immunoassay (EIA) technique is often used in routine diagnostics because of its reliability in high-throughput analyses.
EBV is reactivated frequently, resulting in intermittent excretion of the virus through saliva. Serological EBV reactivation has been studied in detail by using parameters such as antibodies to early antigens, VCA IgA, or the EBNA-1 IgG/EBNA-2 IgG ratio and different other parameters. However, since no clinically relevant disease is generally accepted as being linked to EBV reactivation in immunocompetent individuals, these parameters are only of limited value for the key concerns of routine EBV diagnostics. In contrast, in immunosuppressed patients, EBV reactivation plays a major role and is associated with disorders such as posttransplant lymphoproliferative disease or lymphoma in AIDS patients. However, serological diagnosis of EBV reactivation failed to correlate with the EBV viral load in immunosuppressed patients (8). Thus, antibody diagnostics is discouraged in these individuals.
The aim of the present study was to evaluate four commercially available EIAs, with IFA as the reference method, under routine conditions at three different locations. To address the key concerns of EBV, 264 samples from immunocompetent individuals, comprising EBV-seronegative and -seropositive subjects without EBV-related symptoms and patients with infectious mononucleosis, were tested. In addition, serum samples from patients with possible IgM cross-reactions to EBV due to acute CMV, herpes simplex virus (HSV), or varicella-zoster virus (VZV) disease (all IgM positive) were analyzed. A commercially available precharacterized EBV mixed-titer panel was assayed as well.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Performance of the assays. (i) Reference IFA. At locations A (EBV Reference Center, Department of Virology, University of Homburg/Saar, Homburg/Saar, Germany) and B (Virotest Prof. Enders, Stuttgart, Germany), the same in-house IFAs were performed as references. P3HR1 cells (ATCC HTB-62) were used for VCA IgG and IgM. Recombinant EBNA-1 was expressed in baculovirus-infected SF158 cells. Cells infected with empty virus served as controls. Briefly, cells were air dried and fixed with acetone. Slides were incubated with diluted serum samples for 30 min (VCA IgG and EBNA-1 IgG) or 2 h (VCA IgM) at 37°C and then incubated with fluorescein-conjugated goat anti-human antibody for 30 min. For details of the assays, see references 9, 16, and 17. At location C (Department of Virology, University of Dresden, Dresden, Germany), a commercially available IFA was used in accordance with the manufacturer's (Euroimmun, Groß Grönau, Germany) instructions. P3HR1 cells (ATTC HTB-62) were used for VCA IgG and IgM, and Raji cells (ATCC CCL-86) were used for EBNA IgG. At all locations, titers of at least 1:32 were considered positive.
(ii) EIAs. The four commercially available EIAs (DiaSorin [DiaSorin, Saluggia, Italy], Biotest [Biotest, Dreieich, Germany], Virotech [Genzyme Virotech, Rüsselsheim, Germany], and Novitec [Hiss Diagnostics, Freiburg, Germany]) were employed and interpreted in accordance with the manufacturers' instructions. None of the assays, except the Novitec assay, are approved for diagnostic use in the United States. The Novitec assay is identical to the ImmunoWell EIA (GenBio, San Diego, Calif.) and is approved for use in the United States. DiaSorin has an EBV EIA line approved for diagnostic use in the United States; however, this study was undertaken with the European assay line.
Interpretation of assay results. The interpretation criteria used conformed to the manufacturers' guidelines for the EIAs and the standard interpretations for IFAs. They were identical for all combinations of antibodies except the simultaneous detection of VCA IgG, VCA IgM, and EBNA-1 IgG. This combination was interpreted as past infection in the Novitec and Biotest assays and as primary infection in the DiaSorin and Virotech assays.
Statistical analysis. Kappa statistics was used to measure agreement between the EIAs and the IFA. Kappa values were interpreted as follows: 0.0 to 0.20, poor; 0.21 to 0.40, fair; 0.41 to 0.60, moderate; 0.61 to 0.80, good; 0.81 to 1.00, very good (1).
| RESULTS |
|---|
|
|
|---|
Each of the five assay systems analyzed three parameters: VCA IgG, VCA IgM, and EBNA-1 IgG. The results were interpreted in accordance with the criteria of the respective manufacturers (see Materials and Methods) and allowed a diagnosis of EBV primary infection, past infection, seronegative status, or indeterminate status, respectively. With the IFA as the reference method, discordant EBV diagnoses are summarized in Table 1.
|
Not only the EBV diagnosis, based on the interpretation of all three parameters VCA IgG, VCA IgM, and EBNA-1 IgG, but also the performance of each individual parameter, is relevant when analyzing the quality of the four EIA systems. The serum samples were considered truly positive or truly negative according to clinical disease and IFA results; discrepant results are summarized in Table 2.
|
Simultaneous detection of VCA IgG, VCA IgM, and EBNA-1 IgG should be rare in EBV diagnostics and should be observed only after primary infection, when IgM persists for a while and EBNA-1 IgG is already produced, or during reactivations with increased IgM and EBNA-1 IgG not yet lost. Thus, the frequency of this marker combination is a good indicator of assay performance and the plausibility of the results obtained. The combination was not found by the IFA but was observed in 3 samples by the Novitec assay (1.1%), in 9 samples by the Virotech assay (3.4%), and in 19 serum samples by the DiaSorin and Biotest assays (7.2%)
The results obtained with the BBI panel are shown in Table 3. The samples were pretested by three assays for heterophilic antibodies, two EIAs, and two IFAs by the provider, BBI, but were not clinically characterized (see Materials and Methods). Results of the provider's pretesting, rather than those of the in-house IFA performed at location A (Homburg), were used as the reference. Some reference results were discrepant (indices in Table 3) and thus not comparable. However, 65 were concordant. Of these, two were discrepant with respect to the Novitec assay (3.1%), three were discrepant with respect to the IFA (4.6%), four were discrepant with respect to the Virotech assay (6.2%), six were discrepant with respect to the Biotest assay (9.2%), and eight were discrepant with respect to the DiaSorin assay (12.3%). Thus, the results of the BBI panel confirmed the findings that the Novitec and Virotech assays are more reliable for single-parameter analysis.
|
| DISCUSSION |
|---|
|
|
|---|
The accurate distinction of a primary infection from a past infection or seronegative status is the key concern of EBV diagnostics. Reactivation of EBV resulting in virus replication is a common phenomenon but is normally clinically silent in immunocompetent individuals. In contrast, EBV reactivation causes considerable problems in immunosuppressed patients. However, serological methods are discouraged and PCR is favored for these patients (8). For this reason, serum samples from immunosuppressed individuals were excluded and we did not differentiate between past and reactivated infections. Accordingly, there was no need to test further parameters, such as antibodies to early antigen or VCA IgA.
The best concordance with the reference method for EBV diagnosis was obtained with the Novitec assay, followed by the Biotest assay. The Virotech and DiaSorin assays produced more than twice as many discrepant results as the Novitec assay (Table 1). In the single-parameter analysis (Table 2), the Novitec assay performed better than the Biotest assay, with almost half the number of discrepant results. Even the Virotech assay was better than the Biotest system. Again, the DiaSorin assay showed the lowest concordance. The data obtained with the DiaSorin assay are in accordance with data from a former study conducted at location A (Homburg) with the same reference assay. In that study, 168 serum samples with similar clinical characteristics were evaluated. The results obtained were almost identical, with 14.3% discrepant diagnoses and 7.7% discrepant results in the individual parameters (unpublished results).
The different performances of the individual parameters can be explained by the various antigens used. For an isolated EBNA-1 protein (p72), no native antigens are available; accordingly, the DiaSorin assay employs a synthetic peptide and the manufacturers of the other three use recombinant antigens. In contrast to the Novitec and Virotech assays, which make use of the full-length EBNA-1 antigen, the Biotest assay employs a protein lacking the glycine-alanine repeat sequence of EBNA-1 that is known to occasionally cross-react with cellular proteins (11, 13). Nevertheless, the number of false-positive EBNA-1 IgG results obtained with the Biotest assay was not lower than that obtained with the Novitec or Virotech assay. Thus, deletion of the glycine-alanine repeat sequence did not significantly affect assay performance with this panel of serum samples.
A greater variety of different proteins are employed in VCA assays. The DiaSorin assay uses a synthetic peptide (p18), whereas the Biotest assay makes use of recombinant VCAs (full-length p23 and the carboxy-terminal half of p18). The Novitec and Virotech assays, by contrast, employ an affinity-purified native antigen (p125), which performed best. The VCA of the Biotest assay produced poor results (Table 2). This is underscored by a comparable study in whichapart from a lower number of false-positive VCA IgM resultsalmost similar results were obtained (5). The lowest performance, however, was seen with the VCA protein p18 utilized by the DiaSorin assay. Antibodies against this antigen are produced late during primary infection (2). This is reflected by the fact that all seven of the false-negative VCA IgG serum samples tested with the DiaSorin assay were samples from primary infections.
Combined, the native antigens were obviously superior, in this study, to the recombinant antigens. Synthetic antigens, which lack natural conformational epitopes, showed the lowest concordance with the reference IFA.
Not only the quality of the individual parameters but also their interpretation is a critical factor in assay performance. This is illustrated by the fact that the Novitec and Virotech assays, which use similar antigens, produced roughly identical single-parameter analysis results (Table 2). However, differences in interpretation criteria led to substantially more misdiagnoses on the basis of the Virotech assay (11.7%) than on the basis of the Novitec assay (4.9%; Table 1). Again, the difference in the performance of the Biotest assay in EBV diagnosis versus the single-parameter analysis is explained by interpretation criteria, which are based mainly on the reasonably good Biotest assay EBNA-1 IgG and not on its problematic VCA IgM.
Furthermore, the interpretation schemes of the manufacturers allow diagnoses of EBV infection states that lack clinical equivalence. The only clinically relevant diagnoses in immunocompetent patients are primary infection as a cause of mononucleosis, past infection to exclude mononucleosis, and the absence of antibodies to also exclude EBV disease but show susceptibility. Interpretations such as "transient," "convalescent," and "recent" infections by the manufacturers are of little additional help. One may ask if a "recent," "transient," or "convalescent" infection is correlated with infectious mononucleosis or not. If so, these attributes should be taken to indicate a primary infection. If not, a past infection has to be considered. Taken together, these problems emphasize the urgent need for standardization of the interpretation criteria.
EIA technology offers the benefits of automated high-throughput analyses and objective results in routine diagnostics, whereas IFA technology is time consuming and takes experienced personnel for fluorescence image reading. Generally speaking, the IFA is less sensitive than the EIA. Nevertheless, the specificity of the IFA gold standard remains superior since unspecific reactions, such as anticellular reactivity, are unambiguously detected.
In summary, our results show that two of the four EIAs tested agree well with the reference IFA in the distinction of a primary infection from seronegativity and a past infection and may thus constitute convenient alternatives to the gold standard IFA.