CVI
Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Candal, D.
Right arrow Articles by Parekh, B. S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Candal, D.
Right arrow Articles by Parekh, B. S.

 Previous Article  |  Next Article 

Clinical and Diagnostic Laboratory Immunology, November 2001, p. 1282-1285, Vol. 8, No. 6
1071-412X/01/$04.00+0   DOI: 10.1128/CDLI.8.6.1282-1285.2001
Copyright © 2001, American Society for Microbiology. All rights reserved.

Efficacy of a Less-Sensitive Enzyme Immunoassay (3A11-LS) for Early Diagnosis of Human Immunodeficiency Virus Type 1 Infection in Infants

Debra Candal,1 Marc Bulterys,1 Elaine J. Abrams,2 Richard W. Steketee,1 and Bharat S. Parekh1,*

Centers for Disease Control and Prevention, Atlanta, Georgia 30333,1 and Harlem Hospital Center, New York, New York 100372

Received 30 January 2001/Returned for modification 16 May 2001/Accepted 8 August 2001

We evaluated a less-sensitive enzyme immunoassay (3A11-LS) for its possible use for early diagnosis of human immunodeficiency virus type 1 (HIV-1) infection in infants. The results were compared with those from the immunoglobulin G-capture enzyme immunoassay. A total of 239 sera from 77 infants were tested. All 25 sera from the 10 infants born to seronegative mothers were found to be negative by both assays. Forty-one seroreverting infants showed a complete decay of maternal antibodies by 4 months by the 3A11-LS assay. However, the assay detected HIV antibodies in only 9 (36%) of 25 sera collected from infected infants between 4 and 6 months and in 27 (63%) of 43 sera collected after 6 months of age. Further analysis with alternative cutoff values indicated that the 3A11-LS had a sensitivity of 12 to 44% and a specificity of 90 to 100% for infants between 4-6 months of age. This data suggest that a diagnosis of HIV infection in some of the infants could be made after 4 months of age by the 3A11-LS assay, although a negative 3A11-LS test result may not rule out infection and may require a further followup.


* Corresponding author. Mailing address: HIV Immunology and Diagnostics Branch, CDC/NCID, Mailstop D12, 1600 Clifton Rd., Atlanta, GA 30333. Phone: (404) 639-3647. Fax: (404) 639-2660. E-mail: bparekh{at}cdc.gov.


Clinical and Diagnostic Laboratory Immunology, November 2001, p. 1282-1285, Vol. 8, No. 6
1071-412X/01/$04.00+0   DOI: 10.1128/CDLI.8.6.1282-1285.2001