Performance of Two Commercial Enzyme-Linked Immunosorbent Assay Kits Using Recombinant Glycoprotein G2 Antigen for Detection of Herpes Simplex Virus Type 2 Specific Antibodies
Sharmila M. Reddy,1
P. Balakrishnan,1*
S. Uma,1
S. P. Thyagarajan,2 and
Suniti Solomon1
YRG CARE, Centre for AIDS Research and Education,1
University of Madras, Chennai, India2
Received 22 July 2004/
Returned for modification 15 September 2004/
Accepted 17 November 2004
For 93 stored serum samples tested by HerpeSelect2 and the Euroimmun enzyme-linked immunosorbent assay for detection of herpes simplex virus type 2-specific immunoglobulin G antibodies, the concordance of positive and negative results was 100%. Moreover, all the results that were equivocal by HerpeSelect2 (negative by Euroimmun) were confirmed as being negative by a Western blot assay.